15
6.4.
Setting Threshold Values
This explanation uses
(5-stage comparison) and specifies
(Threshold value using numerical input) or
(Threshold value using sample
weight) to
(input method) in the function table and inputs the threshold value.
Selecting an input method
(If
or )
Step 1 Press and hold the SAMPLE key to enter the function table.
is displayed.
Step 2 Press the SAMPLE key several times to display
(Comparator threshold value).
Step 3 Press the PRINT key to enter
(Comparator threshold value).
Step 4 Press the SAMPLE key several times to input threshold value. Refer to the following
"Threshold value using numerical input" or "Threshold value using sample weight".
Step 5 If the PRINT key is pressed when threshold value is displayed, the current parameters
are displayed (all digits blinks).
Numerical input of threshold values
(If
)
Step 6 If changing the parameters, press the RE-ZERO key to store the following keys.
SAMPLE key
Moves blinking digit.
RE-ZERO key
Changes value of blinking digit.
MODE key
Changes polarity.
PRINT key
Stores and proceeds to step 7.
CAL key
Cancels and proceeds to step 7.
If the current parameters will be maintained, press the PRINT or CAL key to proceed
to step 7.
Step 7 Repeat step 4, 5 and 6 for other threshold values.
Storing threshold values using sample weight
(If
)
Step 8 Press the RE-ZERO key to display 0.00 g .
Put the sample on the weighing pan and press the PRINT key to store weighing value as
the threshold value.
RE-ZERO key
Resets display of the balance to zero.
PRINT key
Stores weighing value and proceeds to step 9.
Step 9 Repeat step 4, 5 and 8 for other threshold values.