15 

6.4.

 

Setting Threshold Values 

 

This explanation uses   

          (5-stage comparison) and specifies   

       

(Threshold value using numerical input) or   

          (Threshold value using sample 

weight) to   

    (input method) in the function table and inputs the threshold value. 

 

Selecting an input method 

(If   

          or     ) 

Step 1  Press and hold the    SAMPLE    key to enter the function table.   

    is displayed. 

Step 2  Press the   SAMPLE   key several times to display   

   (Comparator threshold value). 

Step 3  Press the    PRINT    key to enter   

    (Comparator threshold value). 

Step 4  Press the    SAMPLE    key several times to input threshold value. Refer to the following 

"Threshold value using numerical input" or "Threshold value using sample weight".  

Step 5  If the    PRINT    key is pressed when threshold value is displayed, the current parameters 

are displayed (all digits blinks). 

 
 

Numerical input of threshold values 

(If   

          ) 

Step 6  If changing the parameters, press the    RE-ZERO    key to store the following keys. 

  SAMPLE    key 

Moves blinking digit. 

  RE-ZERO    key 

Changes value of blinking digit. 

  MODE    key 

Changes polarity. 

  PRINT    key 

Stores and proceeds to step 7. 

  CAL    key 

Cancels and proceeds to step 7. 

If the current parameters will be maintained, press the    PRINT    or    CAL    key to proceed 
to step 7. 

Step 7  Repeat step 4, 5 and 6 for other threshold values. 
 
 

Storing threshold values using sample weight 

(If   

          ) 

Step 8  Press the    RE-ZERO    key to display    0.00 g . 

Put the sample on the weighing pan and press the    PRINT    key to store weighing value as 
the threshold value. 

  RE-ZERO    key 

Resets display of the balance to zero. 

  PRINT    key 

Stores weighing value and proceeds to step 9. 

Step 9  Repeat step 4, 5 and 8 for other threshold values. 

Содержание GX-L Series

Страница 1: ...1WMPD4003912B GXM 04 GXL 04 GXM 06 GXL 06 COMPARATOR OUTPUT ANALOG VOLTAGE OUTPUT...

Страница 2: ...duced transmitted transcribed or translated into any language in any form by any means without the written permission ofA D Company Ltd The contents of this manual and the specifications of the instru...

Страница 3: ...9 6 GXM 04 GXL 04 Comparator Output 12 6 1 Specifications of Comparator Output 12 6 2 Using the Comparator Output 13 6 3 ComparisonMethodSelection Tolerance Range and Center Value or Threshold Values...

Страница 4: ...equipped with six contact outputs HH HI OK LO LL of the comparison output and READY output to indicate the status of the balance 3 stage comparison or 5 stage comparison can be selected The GXM 04 GXL...

Страница 5: ...llowing procedure The installation procedure of the GXM 04 and GXM 06 is the same Step1 Remove the two screws from the panel and remove the RS 232C USB board from the balance Step2 Pull gently and rem...

Страница 6: ...e balance Step2 Pull gently and remove the two cables from theRS 232C USB board Step3 Insert the two cables into the sockets of the GXL 06 board Step4 Insert the rubber packing and plug the GXL 06 int...

Страница 7: ...d terminal Step3 Cover the terminal with the top metal case and bottom metal case Step4 Fasten wires with the top metal case Step5 Insert the assembled parts into the rubber sleeve Rubber sleeve Top m...

Страница 8: ...to Center value using sample weight input of Input method in Comparator Caution The balance may not function properly depending on the settings and the operating environment Be sure to set parameters...

Страница 9: ...ts from zero Comparison excluding 20 digits counts from zero Comparison excluding 50 digits counts from zero Comparison excluding 100 digits counts from zero Polarity Comparison of positive values onl...

Страница 10: ...yed when 5 stage comparison is selected Upper limit Lower limit Second lower limit Center value Refer to explanation of 6 GXM 04 GXL 04 Comparator Output Referencevaluewhenusingthedatainputmethod or i...

Страница 11: ...interface Analog output Analog output mode 2 digit output mode Displayed only when the GXM 06 GXL 06 are connected 3 digit output mode Net full scale output mode Gross full scale output mode Output di...

Страница 12: ...male Accessory E8 200J 100 or equivalent Externalcontact input connector Compatibleplug 3 5mm3pole ministereojack MP 013LC or equivalent parts Panel External View Example The GXM 04 Thepanel externalv...

Страница 13: ...One is the method using second upper limit upper limit lower limit and second lower limit Another is the method using center value and tolerance range There two types of input methods for second uppe...

Страница 14: ...tting Centervalue using numerical input or Centervalue using sample weight Procedure Step1 Press and hold the SAMPLE key to enter the function table is displayed Step2 Press the SAMPLE key to display...

Страница 15: ...old value is displayed the current parameters are displayed all digits blinks Numerical input of threshold values If Step6 If changing the parameters press the RE ZERO key to store the following keys...

Страница 16: ...table is displayed Step2 Press the SAMPLE key several times to display comparator threshold value Step3 Press the PRINT key to display comparator threshold value Step4 Pressthe SAMPLE keyseveraltimest...

Страница 17: ...e PRINT key while displaying The current parameter is displayed If the tolerance range is to be changed change it using the following keys Input the tolerance range with center value assumed to be 100...

Страница 18: ...to 0V 0 2 V to 1 V With the slide switch set to 0 2V Input impedance of 10 k or greater the device connected Output update To be updated depending on the cycle of display rewriting The cycle is change...

Страница 19: ...t using the RE ZERO key Gross full scale output mode Outputs 0 000 V when the gross weight is zero Outputs 1 000 V when the gross weight is full scale Tare operation using the RE ZERO key will not aff...

Страница 20: ...V 0 456 V 0 345 V 0 234 V 0 023 V 0 002 V If or Model GX M GF M series Full scale GX 8202M GX 8202MD GF 8202M GF 8202MD 8 kg GX 10202M GF 10202M 10 kg GX 12001M GF 12001M 12 kg GX 22001M GF 22001M 22...

Страница 21: ...100 V 10 kg Caution Full scale of the full scale output mode indicates the full scale values shown in the tables above The output voltage may exceed 1 000 V depending on the weighing value Example Whe...

Страница 22: ...stment and a voltmeter voltage output can be finely adjusted Procedure Step1 Turn off the display using the ON OFF key Voltage output becomes zero point status Adjust the ZERO volume so that the voltm...

Страница 23: ...rload is displayed 0 V is output or 0 2 V if the slide switch is set to 0 2V Step4 When E weighing value is positive overload is displayed the following voltages are output Example Voltage output when...

Страница 24: ...arton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D KOREA Limited 6 33 817 07331 817 Manhattan Bldg 33 Gukjegeumyung ro 6 gil Yeongdeungpo gu Seoul 07331 Korea 82 2 78...

Отзывы: