17
Storing center value using sample weight
(If
)
Step 7 Press the PRINT key while displaying
. The current parameter is displayed (all
digits blinks). Press the RE-ZERO key to enter to the sample weight input mode. Press
the RE-ZERO key to display 0.00 g . Put sample of center value on the pan and press
the PRINT key. The current parameter can be confirmed (all digits blinking). Press the
PRINT key to store the center value.
Step 8 Next,
is displayed.
Step 9 Proceed to "
Storing tolerance range from center value using numerical input
".
Storing tolerance range from center value using numerical input
Step 10 Press the PRINT key while displaying
. The current parameter is displayed.
If the tolerance range is to be changed, change it using the following keys.
Input the tolerance range with center value assumed to be 100%.
SAMPLE key
Moves blinking digit.
RE-ZERO key
Changes value of blinking digit.
PRINT key
Stores and proceeds to step 11.
CAL key
Cancels and proceeds to step 11.
Step 11 Press the PRINT key while displaying
. The current parameter is displayed.
If the second tolerance range is to be changed, change it using the following keys.
Input the second tolerance range with center value assumed to be 100%.
SAMPLE key
Moves blinking digit.
RE-ZERO key
Changes value of blinking digit.
PRINT key
Stores and proceeds to step 12.
CAL key
Cancels and proceeds to step 12.
Step 12 Press the CAL key to return to weighing mode.