background image

 

2

9-6 Description of the Item “Data Format”......................................................................................31 
9-7 Data Format Examples ............................................................................................................33 
9-8 Units ........................................................................................................................................35 
9-9 Storing Units ............................................................................................................................37 
9-10 Counting Mode (PCS)............................................................................................................38 
9-11 Percent Mode (%) ..................................................................................................................40 
9-12 Description of the Item “Application Function” .......................................................................41 
9-13 Statistical Calculation Mode...................................................................................................42 
9-14 Comparator Function .............................................................................................................46 
9-15 Clock and Calendar Function (Only for the FZ-CT series).....................................................47 

10. ID NUMBER AND GLP REPORT ................................................................................................49 

10-1 Setting the ID Number ...........................................................................................................49 
10-2 GLP Report............................................................................................................................49 

11. UNDERHOOK .............................................................................................................................53 
12. PROGRAMMABLE-UNIT ............................................................................................................54

 

 

RS-232C SERIAL INTERFACE 

13. RS-232C SERIAL INTERFACE ...................................................................................................55 
14. CONNECTION TO PERIPHERAL EQUIPMENT.........................................................................56 

14-1 Connection to the AD-8121B Printer......................................................................................56 
14-2 Connection to a Computer.....................................................................................................57 
14-3 Using Windows Communication Tools (WinCT).....................................................................57 

15. COMMANDS ...............................................................................................................................59 

15-1 Command List .......................................................................................................................59 
15-2 Acknowledge Code and Error Codes.....................................................................................60 
15-3 Settings Related to RS-232C.................................................................................................61

 

 

MAINTENANCE 

16. MAINTENANCE ..........................................................................................................................62 
17. TROUBLESHOOTING ................................................................................................................63 

17-1 Checking the Balance Performance and Environment ..........................................................63 
17-2 Error Codes ...........................................................................................................................64 
17-3 Asking For Repair ..................................................................................................................66 

18. OPTIONS ....................................................................................................................................67 
19. SPECIFICATIONS.......................................................................................................................70 
20. EXTERNAL DIMENSIONS ..........................................................................................................72 
21. TERMS/INDEX ............................................................................................................................73 
 
 

 

 

 

Содержание FZ-1200CT

Страница 1: ...FZ CT SERIES FZ 1200CT FZ 700CT FZ 500CT FX CT SERIES FX 1200CT FX 700CT FX 500CT Carat Balance INSTRUCTION MANUAL 1WMPD4002564...

Страница 2: ...ed into any language in any form by any means without the written permission of A D Company Limited The contents of this manual and the specifications of the instrument covered by this manual are subj...

Страница 3: ...ation only for the FZ CT series 14 7 3 Calibration Test Using the Internal Mass Only for the FZ CT series 15 7 4 Calibration Using an External Weight 16 7 5 Calibration Test Using an External Weight 1...

Страница 4: ...49 11 UNDERHOOK 53 12 PROGRAMMABLE UNIT 54 RS 232C SERIAL INTERFACE 13 RS 232C SERIAL INTERFACE 55 14 CONNECTION TO PERIPHERAL EQUIPMENT 56 14 1 Connection to the AD 8121B Printer 56 14 2 Connection t...

Страница 5: ...ult settings of the unit of weight are the metric carat ct and grams g The unit can be switched by pressing the MODE key Standard RS 232C serial interface to communicate with a computer and to output...

Страница 6: ...nterference and under these circumstances the user would be required to take at his own expense whatever measures are necessary to eliminate the interference FCC Federal Communications Commission in t...

Страница 7: ...lance in the future See the illustration to confirm that everything is contained For accurate weighing install the breeze break before use The AC adapter plug is protected against dust and may be diff...

Страница 8: ...illustration on page 5 3 Adjust the leveling feet to level the balance Confirm it using the bubble spirit level 4 Confirm that the AC adapter type is correct for the local voltage and power receptacl...

Страница 9: ...hich produces magnetic fields Install the balance in a stable place avoiding vibration and shock Corners of rooms on the first floor are best as they are less prone to vibration The weighing table sho...

Страница 10: ...the sample Make each weighing gently and quickly to avoid errors caused by the evaporation of moisture from the sample or the absorption of moisture by the sample Use a pair of long tweezers to place...

Страница 11: ...d detergent to clean The edge of the weighing pan is sharp Use caution when cleaning the pan Breeze break standard accessory An antistatic treatment has been applied to the breeze break components Use...

Страница 12: ...ighing mode turns the minimum weighing value on or off In the counting or percent mode enters the sample storing mode Enters the function table mode Refer to 9 FUNCTION TABLE Switches the weighing uni...

Страница 13: ...e container 3 Wait for the stabilization indicator to be displayed Read the value While the stabilization indicator is on pressing the PRINT key will output the weight value using the RS 232C serial i...

Страница 14: ...n release the key 2 Press the MODE key to select a rate of the response adjustment Either FAST MID or SLOW can be selected 3 After a few seconds of inactivity or when the PRINT key is pressed the bala...

Страница 15: ...the first time The balance has been moved The ambient environment has changed Do not allow vibration or drafts to affect the balance during calibration To output the GLP GMP compliant report using the...

Страница 16: ...nce displays end after calibration If the GLP output info parameter of the function table is set to 1 or 2 the balance displays glp and outputs Calibration Report using the RS 232C interface For detai...

Страница 17: ...t info to 1 or 2 3 Press and hold the CAL key until CC in is displayed then release the key 4 The balance measures the zero point Prevent vibration and drafts to affect the balance 5 The measured zero...

Страница 18: ...SAMPLE key and proceed to step 4 If you use the calibration weight value stored in the balance proceed to step 5 4 Specify the calibration weight value as follows SAMPLE key To switch the display con...

Страница 19: ...on or drafts to affect the balance 7 The balance displays end Remove the weight from the pan 8 When the GLP output is set the balance displays 1 glp and outputs Calibration Report 9 The balance automa...

Страница 20: ...want to change the target weight a list of usable weights is shown on page 13 press the SAMPLE key and proceed to step 4 If you use the target weight value stored in the balance proceed to step 5 4 S...

Страница 21: ...rget weight value 6 Place the displayed target weight on the pan and press the PRINT key The balance measures the target weight and displays the measured value Do not allow vibration or drafts to affe...

Страница 22: ...8 FUNCTION SWITCH AND INITIALIZATION and set the internal mass value correction switch to 1 3 Press the MODE key to select the gram mode 4 Press and hold the SAMPLE key to display ba5fnc 5 Press the S...

Страница 23: ...1 2 Press the MODE key to select the gram mode 3 Press and hold the SAMPLE key to display ba5fnc 4 Press the SAMPLE key several times until C5 in 21 is displayed If C5 in 2 is not displayed perform s...

Страница 24: ...ng these parameters Each switch can select either permit or inhibit The inhibit protects parameters against unintentional operations FZ CT series FX CT series Function switch Function table Calibratio...

Страница 25: ...ey once again 8 2 Initializing the Balance This function returns the following parameters to factory settings Calibration data Function table The sample unit mass value counting mode 100 reference mas...

Страница 26: ...e A for Data output mode and 10 digits for Auto print difference 9 2 Display and Keys Display Key Description The symbol indicates that the parameter displayed is in effect When pressed and held in th...

Страница 27: ...urns on the weighing mode display when the AC adapter is connected 0 OFF poff Auto display OFF 1 ON 10 minutes Turns off the display after 10 minutes of inactivity 0 Displays rng Display at start 1 Do...

Страница 28: ...print polarity 2 Both Regardless of displayed value 0 10 digits 1 100 digits ap b Auto print difference 2 1000 digits Difference between reference value and displayed value 0 Every display refresh 1...

Страница 29: ...ID number Refer to 10 ID NUMBER AND GLP REPORT 0 Normal weighing mode 1 Capacity indicator apf Application function 2 Statistical calculation mode 0 Number of data sum 1 Number of data sum maximum mi...

Страница 30: ...s slight fluctuation of a weight value Used to prevent a weight value from drifting due to vibration or drafts Note With Hold function Hold set to ON 1 this item is used to set the stabilization range...

Страница 31: ...pnt The decimal point format can be selected Auto display ON p on When the AC adapter is connected the display is automatically turned on in the weighing mode without the ON OFF key operation Used whe...

Страница 32: ...xample For monitoring data on the AD 8920A remote display Required setting dout prt 3 Stream mode ba5fnc 5pd Display refresh rate 5if bp5 Baud rate Caution The balance may not transmit the data comple...

Страница 33: ...data Output example Print example DP Dump print format 5if type 1 This format is suitable for the peripheral equipment that prints the received data as is When an AD 8121B printer is used set the prin...

Страница 34: ...balance condition The polarity sign is used only for negative data The weight data uses spaces in place of the leading zeros The character length of this format changes dependent upon the unit NU num...

Страница 35: ...33 9 7 Data Format Examples LF...

Страница 36: ...34...

Страница 37: ...d and installed at the factory For details about the units and modes see the table below Name unit mode Display Function table Storing mode Conversion factor 1 g Gram g g 1 g Counting mode Percent mod...

Страница 38: ...ound 0 555 0 335 0 224 0 000005 Pound Ounce 0L 8 88 oz 0L 5 36 oz 0L 3 60 oz 0L 0 01 oz Troy Ounce 8 10 4 88 3 27 0 00005 Metric Carat 1260 760 510 0 001 Momme 67 2 40 5 27 2 0 001 Pennyweight 162 0 9...

Страница 39: ...pecify a unit or mode in the order to be displayed using the following keys SAMPLE key To sequentially display the units RE ZERO key To specify a unit or mode The stabilization indicator appears when...

Страница 40: ...displayed store the counting mode as described in 9 9 Storing Units Storing a sample unit mass 2 Press the SAMPLE key to enter the sample unit mass storing mode Even in the storing mode pressing the...

Страница 41: ...on the pan While the stabilization indicator is on pressing the PRINT key will output the weight value number of objects using the RS 232C serial interface Note Peripheral equipment that is sold sepa...

Страница 42: ...the sample to be set as the 100 reference mass on the pan or in the container 5 Press the PRINT key to store the reference mass The balance displays 100 00 The decimal point position depends on the r...

Страница 43: ...ode the indicator displays the weight data relative to the weighing capacity in percentage Zero 0 maximum capacity 100 Statistical calculation mode apf 2 The mode statistically calculates the weight d...

Страница 44: ...efficient of variation and relative error are obtained by the equation below Standard deviation Standard deviation Coefficient of variation CV Average Maximum value Average Relative error of maximum v...

Страница 45: ...er the data is entered In this case clear the data as described on page 45 and select the unit using the MODE key When the unit used for the statistical calculation mode is to be enabled upon power on...

Страница 46: ...in Statistical function mode output items 5taf and Notes When the number of data is 1 the coefficient of variation and the relative error are displayed as When the average is 0 the coefficient of vari...

Страница 47: ...ete the latest data and exclude it from statistical calculation The number of data decreases by 1 when the balance returns to the weighing mode Clearing the statistical data All the statistical data w...

Страница 48: ...ode Entering the upper limit value 6 With displayed press the PRINT key The current setting of the upper limit value is displayed with all the digits blinking When the current setting is not to be cha...

Страница 49: ...h all the digits blinking When the date is correct and the operation is to be finished press the CAL key and go to step 8 When the date is correct and the time is to be confirmed press the SAMPLE key...

Страница 50: ...the value by one MODE key To decrease the value by one PRINT key To store the new setting display end and go to step 8 CAL key To cancel the new setting and go back to step 4 Quitting the operation 8...

Страница 51: ...test using an external weight Title block and End block for the weight data 10 1 Setting the ID Number 1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed 2 Press the SAMP...

Страница 52: ...for the FZ CT series When the setting is info 1 When the setting is info 2 Calibration test report using the internal mass Only for the FZ CT series Calibration test does not perform calibration When...

Страница 53: ...ting is info 2 Calibration test report using an external weight Calibration test does not perform calibration When the setting is info 1 When the setting is info 2 With FZ CT series the time and date...

Страница 54: ...use MODE 3 of the AD 8121B printer Operation 1 With the weight data displayed press and hold the PRINT key until S5tart is displayed The Title block is output 2 The weight data is output according to...

Страница 55: ...ilt in underhook is revealed by removing the plastic cap on the bottom of the balance Use the underhook as shown below Caution Do not apply excessive force to the underhook When not in use attach the...

Страница 56: ...coefficient 4 The current coefficient is displayed with the first digit blinking When it is not to be changed press the CAL key and proceed to step 6 When it is to be changed press the RE ZERO key and...

Страница 57: ...00 1200 2400 4800 9600 19200 bps Data bits 7 or 8 bits Parity Even Odd Data bits 7 bits None Data bits 8 bits Stop bit 1 bit Code ASCII D Sub 9 pin assignments Pin No Signal name Direction Description...

Страница 58: ...t pause 0 0 0 0 1 2 bp5 Baud rate 2 2 2 2 btpr Data bit parity bit 0 0 0 0 Crlf Terminator 0 0 0 0 5if Serial interface type Data format 0 0 0 1 1 Set appropriate parameters for ap p Auto print polari...

Страница 59: ...smit the weight data to the personal computer The WinCT has three communication methods RsCom RsKey and RsWeight RsCom Can transmit commands to control the balance Can make bi directional communicatio...

Страница 60: ...2 Controlling the balance using commands from a personal computer By using RsCom the personal computer sends commands such as re zero or send weight data to the balance and controls the balance 3 Pri...

Страница 61: ...mmands the S and E SCP commands behave the same Commands to control the balance CAL Same as the CAL key EXC Calibration using an external weight OFF Turns the display off ON Turns the display on P Sam...

Страница 62: ...ommand to control the balance and can not process it the balance transmits an error code EC Exx When the balance receives a command to control the balance and can process it the balance transmits the...

Страница 63: ...noise or a parity error has occurred due to transmission error the balance transmits an error code In this case send the command again 15 3 Settings Related to RS 232C Concerning the RS 232C the bala...

Страница 64: ...se caution when cleaning the pan Breeze break standard accessory An antistatic treatment has been applied to the breeze break components Use a soft dry lint free cloth to clean Cleaning with a cloth t...

Страница 65: ...value Checking that the operating environment or weighing method is proper Operating environment Is the weighing table solid enough Is the balance level Refer to 3 1 Before Use Is the operating enviro...

Страница 66: ...t error The calibration weight is too heavy Confirm that the weighing pan is properly installed Confirm the calibration weight value Press the CAL key to return to the weighing mode EC E21 Calibration...

Страница 67: ...irm the format baud rate and parity EC E01 Undefined command error An undefined command was received Confirm the command EC E02 Not ready A received command can not be processed e g The balance receiv...

Страница 68: ...ice or repair contact the local A D dealer The balance is a precision instrument Use caution when handling the balance and observe the following when transporting the balance Use the original packing...

Страница 69: ...t to a personal computer FXi 08 Ethernet interface Used to connect the balance to a LAN The WinCT Plus data communication software is provided as an accessory and can perform the following Acquire dat...

Страница 70: ...if the personal computer has no COM ports serial communication software such as WinCT can be used via USB connection AD 8920A Remote display Connected to the FZ CT and FX CT series balances using the...

Страница 71: ...e static charge on the sample tare or peripheral equipment and displays the result If those are found to be charged discharge them using the AD 1683 DC static eliminator AD 1687 Weighing environment l...

Страница 72: ...it mass 0 001 g Counting mode 3 Number of samples 5 10 25 50 or 100 pieces Minimum 100 reference mass 0 100 g Percent mode 3 Minimum 100 display 0 01 0 1 1 Depends on the reference mass stored Externa...

Страница 73: ...0 001 g Counting mode 3 Number of samples 5 10 25 50 or 100 pieces Minimum 100 reference mass 0 100 g Percent mode 3 Minimum 100 display 0 01 0 1 1 Depends on the reference mass stored External calib...

Страница 74: ...72 20 EXTERNAL DIMENSIONS FZ 1200CT 700CT 500CT FX 1200CT 700CT 500CT Unit mm...

Страница 75: ...ncluded in the weight data Mode Balance operational function Re zero To set the display to zero GLP Good Laboratory Practice GMP Good Manufacturing Practice Repeatability Variation in measured values...

Страница 76: ...difference 26 apf Application function 27 42 ap p Auto print polarity 26 56 Application 27 42 Application function 27 42 ar d Zero after output 26 at f Auto feed 26 Auto display OFF 25 29 Auto displa...

Страница 77: ...6 Crlf Terminator 27 56 CSV format 27 32 CvV Coefficient of variation 44 Data bits 55 Data format 27 Data logger 69 Data output mode 26 Data output pause 26 DC static eliminator 69 DCE 55 57 Decimal p...

Страница 78: ...42 44 Minimum 100 reference mass 70 71 Minimum unit mass 70 71 Minimum weighing value 70 71 MODE key 3 10 37 MT format 27 32 NU format 27 32 Number of data 27 42 One touch calibration 14 ON OFF key 10...

Страница 79: ...41 43 Statistical function mode output items 27 5t b Stability band width 25 28 Stop bit 55 Stream mode 26 30 5um Sum 44 Sum 27 42 44 Tabletop breeze break 69 Tare 11 38 40 Terminator 27 Timeout 27 Ti...

Страница 80: ...78 MEMO...

Страница 81: ...79 MEMO...

Страница 82: ...80 MEMO...

Страница 83: ......

Страница 84: ...NTS LIMITED German Sales Office Gro e Stra e 13 b 22926 Ahrensburg Deutschland Telefon 49 0 4102 459230 Telefax 49 0 4102 459231 A D Australasia Pty Ltd 32 Dew Street Thebarton South Australia 5031 AU...

Отзывы: