30
Auto print modes A and B
When the displayed value is stable and the conditions of “Auto print polarity”, “Auto print difference”
and reference value are met, the balance outputs the weight data.
When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the
data and the display blinks one time.
Mode A: Example
“For outputting the weight value each time a sample is added (or
removed), with “
ar-d
” set to “
1
” (to adjust zero after the data is
output).”
Required
setting
dout prt
1
Auto print mode A (reference = zero)
dout ap-p
Auto print polarity
dout ap-b
Auto print difference
dout ar-d
1
Adjusts zero automatically.
Mode B
:
Example
“For outputting the weight value while a sample is added.”
Required
setting
dout prt
2
Auto print mode B (reference = last stable value)
dout ap-p
Auto print polarity
dout ap-b
Auto print difference
Stream mode
The balance outputs the weight data at the specified display refresh rate, regardless of the display
condition. The display does not blink in this mode.
Example
“For monitoring data on the AD-8920A remote display”
Required setting
dout prt
3
Stream
mode
ba5fnc 5pd
Display refresh rate
5if bp5
Baud
rate
Caution
The balance may not transmit the data completely at the specified display refresh rate and
baud rate. Set the baud rate higher.
Key mode B
When the PRINT key is pressed, the balance outputs the weight data, regardless of the display
condition. The display does not blink in this mode.
Key mode C
When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the
weight data. When the stabilization indicator is not on, the balance waits for the indicator to turn on
and outputs the data. The balance blinks one time in this mode.
Interval output mode
The weight data is periodically output.
When the PRINT key is pressed, the balance starts to output the weight data at a preset interval
time. When the PRINT key is pressed again, the balance stops outputting the weight data.
Example
“For outputting the weight data periodically."
Required setting
dout prt
6
Interval output mode
dout int
Interval
time
Caution
The balance may not transmit the data completely at the specified interval times and baud
rate. Set the baud rate higher.
Содержание FZ-1200CT
Страница 35: ...33 9 7 Data Format Examples LF...
Страница 36: ...34...
Страница 74: ...72 20 EXTERNAL DIMENSIONS FZ 1200CT 700CT 500CT FX 1200CT 700CT 500CT Unit mm...
Страница 80: ...78 MEMO...
Страница 81: ...79 MEMO...
Страница 82: ...80 MEMO...
Страница 83: ......