3
3.
Calibration
Calibration and sealing must be performed by weights and measures official, or an authorized service
agent.
(1)
Remove the sealing screws, sealing plate, blank panel and rubber packing.
(See “Fig.1” for reference)
(2)
Enter the weighing mode.
(3)
Press and hold the calibration switch S201 (Refer to “Fig.2”), and press
the CAL key on the front of the balance for a few seconds.
The balance will display “Cal 0”.
(4)
Perform the calibration per the instruction manual.
(5) After performing calibration, replace the rubber packing,
blank panel, and sealing plate, then seal the balance.
(Refer to “Fig.1”)
4.
Sealing of FX-iN/FX-iWPN legal for trade approved models
Once calibration following the instruction manual is completed, replace
the blank plate (See “Fig.3”).
Once the sealing plate is installed, seal by threading wire through the
eyelets of the round sealing knobs and through the eyelet at the base
or bottom of the balance. (See”Fig.4”)
Important note:
The options, FXi-02,08 and 09 cannot be used at the same time.
The selected option should be installed into the balance before sealing.
5.
Function switch
The settings for the LFT modes are shown below for reference.
FX-iN/FX-iWPN: A-0000
1
Function table: 1; To permit change
Calibration using the internal mass: 0; No function
Calibration using the external weight: 0; To inhibit (No change to 1)
Automatic self-calibration: 0; No function
Internal mass correction:0; No function
Fig 4
Fig 3