11
6-2. Function list
Item
Function number
Description
C01-
0
CAL switch enabled,
0
key enabled
C01-
1
CAL switch disabled,
0
key disabled
Calibration
C01-
2
CAL switch enabled,
0
key disabled
Do not set
“
C01-1
”
C02-
0
6 kg (15 lb)
C02-
1
15 kg (35 lb)
C02-
2
30 kg (70 lb)
C02-
3
60 kg (150 lb)
C02-
4
150 kg (350 lb)
C02-
5
300 kg (700 lb)
C02-
6
600 kg (1500 lb)
Capacity
C02-
7
1500 kg (3500 lb)
C03-
0
kg, g
Weighing display
unit
C03-
1
kg, g, lb, oz, lb-oz
See “
11-1.
Capacity ...
”.
C04-
0
±
10% of the capacity
OIML R76 / NTEP
Power-on-zero
range
C04-
1
±
50% of the capacity
Internal use
C05-
0
Do not change.
C06-
0
.
(point)
Decimal
separator
C06-
1
,
(comma)
C07-
0
ZERO does not clear tare weight
Tare clear by
ZERO operation
C07-
1
ZERO clears tare weight
OIML R76
C08-
0
Tare weight is rounded to the weight
display count
NTEP
Tare operation
(Net weight
calculation)
C08-
1
Internal count is used for TARE weight.
OIML R76
C09-
0
Function disabled
Preset tare
function
C09-
1
Function enabled
C10-
0
lb-oz disabled
lb-oz unit
C10-
1
lb-oz enabled
C11-
0
±
2% of the capacity
ZERO operation
range
C11-
1
Up to the capacity
C12-
0
“
f02
” setting specifies the resolution.
C12-
1
Fixed to “
f02-0
” (1/3,000~1/3,500)
C12-
2
Fixed to “
f02-1
” (1/6,000~1/7,500)
Display resolution
C12-
3
Fixed to “
f02-2
” (1/12,000~1/15,000)
C13-
0
Function disabled
Auto-tare function
C13-
1
Function enabled
C13-0
deletes
“
f21
” ~ “
f23
”.
Factory setting
Содержание FS-D