22
Item
Parameter
Description
Used internally
Not available for setting change
Saving zero value
and tare weight
Do not save
Saves zero value
and tare weight
even after the
power turns off.
Save
Display limit of “ ”
Weighing ca 8d
Weighing ca 8d + (capacity/30)
Weighing ca 8d + (capacity/15)
Weighing ca 8d + (capacity/6)
Weighing ca 8d + (capacity/3)
Zero tracking
width/ time
0.5d / 0.5 seconds
Zero tracking:
The function to
track the zero point
and keep the display
zero
0.5d / 1.0 seconds
0.5d / 1.5 seconds
0.75d / 0.5 seconds
0.75d / 1.0 seconds
0.75d / 1.5 seconds
1.0d / 0.5 seconds
1.0d / 1.0 seconds
1.0d / 1.5 seconds
:
Factory settings (which differ depending on the destination market) d = minimum division
The settings are not changeable for CF1, 2, 8, 12, or 13.
Note: About minus and printing limit setting (
)
When setting
to “ ”, the indicator has the following limit.
When the total weight is -19d or less, the indicator displays
-
.
When the total weight is minus, the indicator does not perform the printing.
9. Options
9.1. Option List
Refer to the optional instruction manual for details.
Option name
Contents
FG-27CWP
Bluetooth communication interface
AD-8541-PC
Bluetooth dongle for PC connection
AD-8931
Bluetooth External Display
AX-TB301
AC adapter