14
7. Sensitivity Adjustment
The description of sensitivity adjustment in this manual is different from that in the instruction
manual for the FG-CWP series. Please read the description below.
7.1. Sensitivity Adjustment Items
Sensitivity adjustment mode has the following three functions.
Setting of the weighing capacity, minimum division, and decimal point position.
Sensitivity adjustment using a weight
Correcting for gravitational acceleration
There are two ways to enter sensitivity adjustment mode as shown below.
Method 1 (using operation keys)
1. Confirm that the scale(indicator) is in weighing mode (display shows measurement unit).
2. Press and hold the TARE key for 4 sec. or longer. Release the key when
is
displayed. (Then go to “
7.2. Setting of the Weighing Capacity / Minimum Division / Decimal
7.3. Sensitivity Adjustment Using a Weight
”.)
Method 2 (using the CAL switch)
1. Confirm that the scale(indicator) is in weighing mode
(display shows measurement unit).
2. Remove the protective cover from the four screws
of the load cell panel on the rear of the display unit,
and then loosen the four screws and open the load
cell panel.
3. Press the CAL switch to display
.
(Then go to
7.2. Setting of the Weighing Capacity /
Minimum Division / Decimal Point Position
Sensitivity Adjustment Using a Weight
Correcting for Gravitational Acceleration
”)
Load cell panel
Rear of display unit
CAL switch