FG-D-CWP
Weighing Indicator
For Waterproof Digital Platform Scale
1WMPD4004854
Страница 1: ...FG D CWP Weighing Indicator For Waterproof Digital Platform Scale 1WMPD4004854 ...
Страница 2: ...ribed or translated into any language in any form by any means without the written permission of A D Company Ltd The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice Windows Word and Excel are registered trademarks of the Microsoft Corporation WARNING A potentially hazardous situation which if not avoided c...
Страница 3: ...TTINGS 11 6 PART NAMES 12 6 1 DISPLAY AND SYMBOLS 12 6 2 OPERATION KEYS 13 7 SENSITIVITY ADJUSTMENT 14 7 1 SENSITIVITY ADJUSTMENT ITEMS 14 7 2 SETTING OF THE WEIGHING CAPACITY MINIMUM DIVISION DECIMAL POINT POSITION 16 7 3 SENSITIVITY ADJUSTMENT USING A WEIGHT 17 7 4 CORRECTING FOR GRAVITATIONAL ACCELERATION 18 8 SETTING FUNCTIONS OR UNITS FOR USE 19 8 1 SETTING OF THE CF FUNCTION 19 8 2 SETTING U...
Страница 4: ...nvironment If this unit is operated in a residential area it might cause some interference and under these circumstances the user would be required to take at his own expense whatever measures are necessary to eliminate the interference FCC Federal Communications Commission in the U S A 1 2 Classification of Protection Provided by Enclosures This equipment complies with the IP Code IEC 60529 The m...
Страница 5: ...from the A D website https www aandd jp 3 Outline and Features The FG D CWP is an electronic indicator that is appropriate for composing an IP67 waterproof digital platform scale by connecting to a weighing platform load cell and it has the following features The indicator made of all stainless steel and has waterproof dustproof performance compliant with IP67 Intrusion of water into the device is...
Страница 6: ...tc Attach here FG D CWP This document is included in the packing contents for the balance Quick Start Guide For Water proof platform scale Engli sh EN Guide de démarrage rapide Pour balance plate forme étanche Français FR Schnellstartanleitung Für wasserdichte Plattformwaage Deut sch DE Guía de inicio rápido Para báscula de plataforma a prueba de agua Español ES Guida introduttiva Per bilancia a p...
Страница 7: ...Scale in Instruction Manual FG CWP Series on the A D website https www aandd jp for other precautions cautions as an assembled scale storage and cleaning 5 2 Load Cell Connection FG D CWP has two connection configurations the 6 wire configuration and the 4 wire configuration We recommend connecting using the 6 wire configuration for optimum accuracy and stability The applicable diameter of load ce...
Страница 8: ...B series 1 White Excitation Input terminal 2 Red Excitation Input terminal 3 4 Blue Signal Output terminal 5 6 Green Signal Output terminal 7 Yellow Shield Method Advantage Disadvantage Remark 6 wire configuration recommended Less error even if the length of load cell cable is long or a thin load cell cable is used Less error even if multiple load cells are used Slightly complicated to wire When u...
Страница 9: ...mp does not need to be replaced 4 Refer to 5 2 Load Cell Connection and connect the cable core to the terminal block of the load cell Hold down the white button on each terminal in the terminal block of the load cell when connecting or disconnecting the wire Additional tool is not required 5 Once the load cell is connected make sure that wiring has been done correctly 6 Fasten the lock of the cabl...
Страница 10: ...as possible Calculating formula 0 1 E x B x d A Rated capacity of load cell B Rated output mV V d Minimum division E Excitation voltage mV 5000 mV for FG D CWP n Number of load cells A x n Example of calculation In the case of designing a scale with a capacity of 60 kg and minimum division of 0 005 kg using one load cell with a rated capacity of 100 kg and rated output of 1 mV V Rated capacity of ...
Страница 11: ... The USB cable serves as both a power and a communication If using the scale in a location where static electricity is easily generated or where it may get wet secure the grounding wire with the protective grounding terminal screw 5 6 Sensitivity Adjustment and Parameter Settings When connecting to a load cell weighing platform sensitivity adjustment needs to be performed For details refer to 7 Se...
Страница 12: ...alues NET This is lit when weight of the container tare is subtracted ZERO This is lit when the scale is at the zero point reference point for weighing Weighing Units and is lit This is lit when the display is held Load Cell Panel Comparator LED Weighing Units Weighing Condition Conditions for Each Function Display Unit SUS304 Display unit knob Serial Number USB Plug Load Cell Connector USB Cable ...
Страница 13: ...ows or hides the display SAMPLE Key Selecting as the unit activates individual weight setting mode Pressing and holding for about 2 seconds activates the comparator threshold setting mode PRINT Key Outputs the weighing value as data Increases the value of the digit flashing when setting MODE Key Switches the weighing unit Shifts to the right the digit flashing when making settings ZERO Key Based o...
Страница 14: ...rement unit 2 Press and hold the TARE key for 4 sec or longer Release the key when is displayed Then go to 7 2 Setting of the Weighing Capacity Minimum Division Decimal Point Position 7 3 Sensitivity Adjustment Using a Weight or 7 4 Correcting for Gravitational Acceleration Method 2 using the CAL switch 1 Confirm that the scale indicator is in weighing mode display shows measurement unit 2 Remove ...
Страница 15: ...justment mode press the ZERO key or CAL switch The display turns off You can also turn it off using the ON OFF switches In sensitivity adjustment mode transition method Weighing mode Method 1 Press and hold Method 2 Press Setting of the weighing capacity minimum division and decimal point Display OFF Setting of the CF function Sensitivity adjustment using a weight Press or Press Press Press Press ...
Страница 16: ...s the TARE key to store the currently set minimum division and calibration unit and proceed to the next step 4 After displaying the indicator displays the currently set weighing capacity and decimal point position Display example kg Set the weighing capacity and decimal point position using the following keys MODE Shift the blinking digit to the right PRINT Increases the number 1 at the blinking d...
Страница 17: ...m or place something on the weighing pan to keep the automatic power off function from activating 2 Refer to 7 1 Sensitivity Adjustment Items is displayed when you enter sensitivity adjustment mode 3 Press the MODE key to display 4 Press the TARE key to display Make sure there is nothing on the weighing pan and wait for the stability indicator to light up 5 Press the TARE key to save the zero poin...
Страница 18: ...d change the gravitational acceleration value recorded in the scale indicator to the one for appropriate location Note If sensitivity adjustment is done using a weight at the place where the scale indicator is used it does not need to be corrected for gravitational acceleration 1 Refer to 7 1 Sensitivity Adjustment Items is displayed when you enter sensitivity adjustment mode 2 Press the MODE key ...
Страница 19: ...er using the following keys MODE Proceed to next setting item PRINT Increases the number 1 at the blinking digit Change the parameter When the currently stored set is changed the STABLE indicator turns off Changing the parameter using the PRINT key at CF1 2 8 12 or 13 is not available 5 Press the TARE key to store the setting After displaying the indicator displays and proceeds to the unit setting...
Страница 20: ...g mode Each of the display digits are assigned to a unit setting The setting at the first and fourth digit is not available Display assignment 7 6 5 4 3 2 1 Unit pcs g kg Set disabled oz lb t The unit during selection is indicated With each digit if setting to 0 its unit is not available and if setting to 1 its unit is available Only when set to 1 is a unit available Even if a unit is stored avail...
Страница 21: ...ion Do not illuminate the zero point mark during tare subtraction Illuminate the zero point mark during tare subtraction only when the net weight is zero Illuminate the zero point mark during tare subtraction Tare clear by zero operation Tare is not cleared by zero operation Tare is cleared by zero operation Used internally Not available for setting change Minus and printing limit Non limit Refer ...
Страница 22: ... 75d 0 5 seconds 0 75d 1 0 seconds 0 75d 1 5 seconds 1 0d 0 5 seconds 1 0d 1 0 seconds 1 0d 1 5 seconds Factory settings which differ depending on the destination market d minimum division The settings are not changeable for CF1 2 8 12 or 13 Note About minus and printing limit setting When setting to the indicator has the following limit When the total weight is 19d or less the indicator displays ...
Страница 23: ...nnected The scale does not display zero at first Check around the weighing pan Is there anything on the weighing pan Refer to 7 Sensitivity Adjustment and perform zero point sensitivity adjustment is displayed Weighing error that meaning Overloaded is displayed Indicates that the sample weight is too light to set the unit weight in the counting mode is displayed Sensitivity adjustment error that m...
Страница 24: ...rrors are displayed try turning the display off and on again Display Content is displayed The temperature sensor has failed is displayed The memory circuit has failed is displayed The internal circuitry has failed is displayed The mass sensor has failed ...
Страница 25: ...typ 0 1 μV C max Temperature coefficient Span 3 ppm C typ 15 ppm C max Linearity 0 01 of full scale Maximum display resolution 100 000 Display 7 segment LCD with backlight Character height 30 mm Display update Approx 10 times second Operating Consitions 10 C to 40 C 85 R H or less non condensing Power Supply AC adapter or mobile battery or USB port USB Cable Length Approx 3 m USB cable is also use...
Страница 26: ...26 11 2 Dimensions FG D CWP Unit mm ...
Страница 27: ...03 m s2 Ottawa Paris 9 807 m s2 9 810 m s2 Copenhagen 9 816 m s2 Rio de Janeiro 9 788 m s2 Cyprus 9 797 m s2 Rome 9 803 m s2 Frankfurt 9 811 m s2 San Francisco 9 800 m s2 Glasgow Havana 9 816 m s2 9 788 m s2 Singapore Stockholm 9 780 m s2 9 819 m s2 Helsinki 9 819 m s2 Sydney 9 796 m s2 Jakarta 9 781 m s2 Taichung 9 789 m s2 Kolkata Calcutta 9 788 m s2 Taipei 9 790 m s2 Kuwait Lisbon London Greenw...
Страница 28: ...g 33 Gukjegeumyung ro 6 gil Yeongdeungpo gu Seoul 07331 Korea 전화 82 2 780 4101 팩스 82 2 782 4264 OOO A D RUS OOO ЭЙ энд ДИ РУС Почтовый адрес 121357 Российская Федерация г Москва ул Верейская дом 17 Юридический адрес 117545 Российская Федерация г Москва ул Дорожная д 3 корп 6 комн 8б 121357 Russian Federation Moscow Vereyskaya Street 17 тел 7 495 937 33 44 факс 7 495 937 55 66 A D Instruments India...