19
10. Counting accuracy is improved when the ACAI indicator turns off. Each time the above
operation is performed, a more accurate unit weight will be obtained. There is no
definite upper limit of ACAI range for the number of samples exceeding 100. Try to add
a similar number of samples as that displayed.
5.9.
Percent Mode (%)
The balance displays the weight value in percentage compared to a reference (100%)
weight.
Selecting the Percent Mode
1. Press the UNITS key to select
%
. (%:percent)
Storing the Reference (100%) Weight
2. Press the SAMPLE key to enter the reference
weight storing mode.
3. Press the RE-ZERO key to display
100 0
%.
4. Place the sample to be set as the reference weight
on the pan.
5. Press the PRINT key to store the reference
weight. Remove the sample.
When the reference weight is too light to store, the
display shows for the moment and returns
to the former display.
Reading the Percentage
6. Place the object to be compared to the reference
weight on the pan. The displayed percentage is
based on 100% of the reference weight.
To store
0.00
%
100 -
%
100.00
%
75.80
%
100 0
%
ZERO
Weighing pan
Sample corresponding
to the 100% weight
Object to be compared
Содержание EJ-1202
Страница 50: ...48 EJ 54D2 190 209 133 Min 114 Unit mm 50 ...
Страница 53: ...51 World Map ...
Страница 54: ...52 MEMO ...
Страница 55: ...53 MEMO ...