18
5.8.
Counting Mode (pcs)
The balance weighs the sample pieces and calculates the unit weight.
Using the sample unit weight, the balance counts the number of items in the sample.
As for the minimum unit weight acceptable, refer to the item “
Umin
” of “
8.5. Function
List
”.
Selecting the Counting Mode
1. Press the UNITS key to select
. ( :pieces)
Storing the Sample Unit Weight
2. Press the SAMPLE key to enter the sample unit
weight storing mode.
3. To select the number of samples, press the
1
SAMPLE key. It may be set to 5, 10, 25, 50, or 100.
4. Place a tare container on the weighing pan, and
press the RE-ZERO key. Confirm that the
right side of the number of samples shows zero.
5. Place the number of samples specified on the pan.
In this example, 25 pieces.
6. Press the PRINT key to calculate and store the
unit weight. Remove the sample. The balance is set
to count objects with the unit weight.
When the unit weight is too light to store, the display
shows
lo
for the moment and returns to the
former display. When the sample weight is light and
the counting error could become large, the balance
will request you to use a larger number of samples.
Place the displayed number of samples on the pan
and press the PRINT key to calculate and store
the unit weight. Refer to “Sample quantity error” and
“Unit weight error” in “
11.2.Error Codes
”.
Counting the Objects
7. Place the objects to be counted on the pan.
Counting Mode Using the ACAI Function
ACAI
TM
(Automatic Counting Accuracy Improvement)
is a function that improves the accuracy of the unit
weight by increasing the number of samples as the
counting process proceeds.
8. After storing the unit weight at step 6, add a few
more samples on the pan. Then, the ACAI indicator
turns on. (Add more than three pieces. The ACAI
indicator will not turn on for the number of samples
beyond the ACAI range.)
9. The ACAI indicator will flash and the balance re-calculates the unit weight. Do not touch
the balance or samples on the pan until the ACAI indicator turns off.
Weighing pan
0
10 -
25 -
Container
25
ACAI indicator
ACAI
25
Confirm the display
25 0
ZERO
Switches the number of samples
Samples
(25 pcs)
To store