89
External calibration test
Display settings: MENU key
Calibration test/check button
External calibration test button
[External calibration test] screen
3
4
1
5
2
6
Name
Description
1
External weight value input
Input the external weight value. Input range: Input range
*1
This is common to the external weight value input field in the
[Sensitivity adjustment setting] screen.
2
Back button
Returns to the previous screen.
3
Calibration test result display
Displays the measurement result.
4
Confirm button
Confirms the current input setting and proceeds to the next
instruction.
5
PRINT button
Outputs the measurement result to the device connected to the
balance.
The result is automatically output if [GLP output] in [Data output]
is set to ON.
6
Exit button
Exits the screen that shows the measurement result.
Settings in the red box are default values (factory settings).
*1
Input range
BA-6TE/BA-6DTE
: 0.9 g ~
5 g ~
5.1 g
BA-225TE/BA-225DTE
: 9.9 g ~ 200 g ~ 200.1 g
BA-125DTE
: 9.9 g ~ 100 g ~ 100.1 g
Your external weight is used to perform the calibration test.
Be sure to warm up the balance with nothing on the weighing pan for at least an hour, or at least
four hours for BA-6TE/BA-6DTE, with the AC adapter connected to the power supply.
Do not apply vibration and the like to the balance during the calibration test.
When the test is completed, the [Calibration test result] screen is displayed.
Caution
No sensitivity adjustment is performed.
Содержание BA-125DTE
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315 ...
Страница 74: ...74 Output example for daily check output results Daily check output results Daily check PDF results ...
Страница 84: ...84 Output example for periodic check output results Periodic check output results ...
Страница 85: ...85 Periodic check results PDF ...
Страница 166: ...166 Unit mm Display unit 138 182 73 Ionizer AX ION 25 162 68 129 ...
Страница 170: ...MEMO ...
Страница 171: ......