![A&D AD-1641 Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/aandd/ad-1641/ad-1641_instruction-manual_3071505022.webp)
20
RS-232C
cable
USB
extension
cable
8.3.5 Transmitting Weighing Data And Environmental Data
Step 1
Connect the AD-1641 unit to the weighing instrument with the RS-232c
cable while the AD-1641 is powered off. After that, connect the AD-1641
unit to the PC with the USB cable. After connecting, the display will change
the "8.3.1" display.
Step 2 Select the item "REAL TIME" and press the ENTER key to use the
real time transmission mode.
Step 3 Click on a position (cell) on the program (EXCEL etc.) to receive data.
Step 4
Press the data output key (Ex.:PRINT key) of the weighing instrument to
send weighing data. When the main unit of AD-1641 receives data, a
buzzer sounds and weighing data and environmental data are sent to the
computer.
Step 5 Data is displayed for 5 seconds on the measurement monitor.
Note
■
If the timing of the transmission is short, data may not be saved depending
on the PC performance.