8.2 Connecting To The Weighing Instrument
The AD-1641 saves the weighing data and environment data at the same time
when receiving the weighing data.
8.2.1 The management of the weighing data
Step 1 Connect the AD-1641 and A&D's weighing instrument using the provided
RS-232C cable.
Step 2 Press the data output key (Ex. PRINT key)
of the weighing instrument to transmit
data. The AD-1641 receives data and
the indicator REC is shown for 1second.
Step 3 Stored data is shown for 5 seconds in the measurement monitor.
8.2.2 The Interval Measurement For The Weighing Instrument
By setting 7.4.Function table "Weighing value interval recording mode
(WEIGHING DATA)" of the AD-1641 to "Interval", it is possible to save the
weighing data sent from the weighing instrument continuously (stream output)
at each interval.
Step 1 Select the "Interval" in 7.4.Function table "Weighing value interval recording
mode" (WEIGHING DATA)
WEIGHING DATA
:
Interval (The factory setting is "Every")
Step 2 Connect the Weighing instrument that has been set in the stream output.
In the case that the AD-1641 has received the weighing data, (WT) is
displayed at lower left corner of the screen of the AD-1641, and weighing
data is displayed
.
In this case, the weighing data is not saved.
Step 3
Pressing the START/STOP key, at the timing set by the interval time, starts
saving weighing data.
Note
■
A recording interval might deviate for a few seconds due to time lags in data
reception.
■
To set the display format of weighing data or to set the stream output of the
weighing instrument, please refer to the "internal setting" "data format" of each
instrument manual
.
■
When the AD-1641 body has received the weighing data, weighing data is
displayed.
Weighing data