background image

2.0 Components

Figure 1 shows the SAAFShield Detecting Unit with the internal 

components labeled.
The SAAFShield Detecting Unit Real-Time Reactivity Monitor 

includes the following components:
•  (1) Oscillator Printed Circuit Board (PCB)
•  (3) Connector slots for Quartz Crystal Microbalance (QCM)  

 PCBs
•  (1) Black plastic thermoplastic case with airflow zones and  

  protective cage
•  (1) Connector slot for interfacing with the SAAFShield  

  Reading Unit
•  (3) QCM PCBs (packaged separately)  

IMPORTANT: 

To open the QCM PCB package prior to  

installation, please carefully follow the instructions supplied  

below. Mishandling the QCM PCB can cause irreparable 

damage.  

 

3.0 Installing QCM

Tools Required:
•  Phillips head screwdriver
•  Scissors, exacto knife, or razor blade

Instructions:

1)  Ensure that the Electrostatic Discharge (ESD) protected bag 

contains the appropriate QCM (Gold, Silver, Copper) that you 

would like to install. See Figure 2.

2)  Carefully cut open the ESD bag and remove any protective 

packaging around the QCM PCB.

3)  The QCM-PCB has a protective black foam backing—do not 

remove this foam.

4)  You will need to remove the plastic protective banding from 

the QCM to expose the reactive elements on the QCM.  

Carefully cut one end of the plastic protective banding as 

shown in Figure 3 with a razor blade or exacto knife.  

CAUTION:

 In some instances, the plastic band material  

may be tightly adhered to the metal cap. This is done to  

provide a tight seal on the cap and ensure complete  

protection of the QCM reactive element during shipment. 

Please do not force open the metal cap. Doing so may  

cause irreparable damage to the QCM’s reactive element  

within, making the device non-operational. It is advised that  

the metal cap be gently removed from the QCM and  

handled carefully at all times.

2

Figure 1 – SAAFShield

 Detecting Unit with internal components labeled

Figure 2 

Figure 3 

QCM PCBs

Connector slots 

(connects to  

QCM-PCB)

EVA foam  

spacers

Outside  

casing

Oscillator PCB

Cutouts  

for airflow

Ribbon cable connector slot  

(connects to Reading Unit)

Отзывы: