• Please direct the arrow on the check valve toward the venturi injector.
• Turn the oxygen generator ON first and then the ozone generator switches (the air pump
stays OFF).
• Turn timer knob clockwise to the amount of time needed for the given application.
• To run continually, turn the timer knob counterclockwise.
• The ozone light and fan should turn on.
• Increase power to the cells by turning the ozone control knob clockwise.
• Prior to shutting off the oxygen generator feed gas to the MP-8000, first turn off the
ozone generator switch and allow the unit to purge with oxygen for about 1 minute.
• Then, turn off the oxygen generator.
NOTE:
Do not turn ON the air compressor switch on the front panel when using oxygen
generator, oxygen bottle or tank.
Using Internal Air Compressor:
This unit comes with a small air compressor built
inside the ozone generator
• Connect the air out of the unit with the
O
2
In using the supplied oxygen tubing, located
on the front panel on the ozone generator side.
• Connect ozone generator (Ozone Out) to the venturi injector.
• Install a check valve 10 inches after ozone outlet tube fitting.
• You may do this by cutting the tube ten inches after ozone outlet fitting.
• Connect both ends of cut tubes to the check valve.
• Please direct the arrow on the check valve toward the venturi injector.
• Turn ON/OFF both the main power and air compressor switch located on the front panel.
• To power down, flip the ozone switch and wait 30 seconds. Then flip the air pump switch.
Using External Air Compressor:
• Install air dryer after the air compressor. Use an oil-less air compressor only.
• Connect air dryer outlet to ozone generator (Oxygen In).
• Set air inflow rate at 1-4 liters per minute based on which model you have.
• Connect ozone generator (Ozone Out) to venturi injector.
• Install a check valve 10 inches after ozone outlet tube fitting. This can be done by
cutting a tube ten inches after ozone outlet fitting.
• Connect both ends of cut tubes to the check valve. Please direct the arrow on the check
valve toward venturi injector.
• Turn ON/OFF main power and air compressor switch at the same time.
• To turn off, flip off the ozone power switch and wait 30 seconds. Then turn off external
air pump.