15000-MNL-01 Rev A
32
wait 10 seconds and reconnect the LDI to
the PC. Unplug the unit and wait 10
seconds before reapplying power to the
system. If the error persists, please
contact a technical representative.
PC_APP_WRITE_TIMEOUT_ERR
-40007
LDI UI - communication
timeout with LDI
Usually during a firmware
update
If you were trying to update your
unit’s
firmware, try updating the firmware
again. If the error persists, place the unit
in boot loader mode by holding down the
soft power button until it starts blinking
blue, then try updating the firmware
again. If the error persists, please contact
a technical representative.
PC_APP_WRITE_FLASH_ERR
-40008
LDI UI - error updating
firmware
If you were trying to update your unit’s
firmware, Try updating the firmware
again. If the error persists, place the unit
in boot loader mode by holding down the
soft power button until it starts blinking
blue, then try updating the firmware
again. If the error persists, please contact
a technical representative.
PC_APP_INVALID_RESPONSE
-40009
LDI UI - received invalid
response from LDI
Check USB cable connection to the host
PC. Unplug the unit and wait 10 seconds
before reapplying power to the system. If
the error persists, please contact a
technical representative.