![4tek MIRAGE Скачать руководство пользователя страница 11](http://html.mh-extra.com/html/4tek/mirage/mirage_operating-and-maintenance-manual_445329011.webp)
This manual is the property of
. No part of it may be reproduced in any form.
MIRAGE Line Compressors Instruction Manual
Description of the compressor and its component parts
The equipment is an oil-less compressor designed
to supply compressed air for dental and/or medical
applications.
Exploiting the rotation of the crankshaft, the pistons draw in air from the place where the
compressor is installed via the carter, pressurise it, and transfer it to the air receiver for
storage.
This operation is regulated by a pressure switch, which is typically set at between
6,5 bar and 8.5 bar; the process is ultimately controlled by the safety valve: if the pressure
switch malfunctions and the air receiver pressure exceeds the safety valve setting, the valve
will crack open to vent the overpressure and eliminate the risk of bursting.
Class I appliance in relation to the new medical devices regulation (MDR) 2017/745
The product is not designed to withstand immersion in liquid (IPX0)
Class I: electrical insulation
Oil-less 2-cylinder air compressor (MIRAGE130E), applicable also to all the following
models in the same product family.
2 - 01
MIRAGE130E compressor composed of:
2-cylinder head, equipped with single-phase or three-
phase motor, characterised by an air inlet of 200 l/min, air
delivery at 5 bar 130 l/min
1 air receiver capacity: 25 l
Available also in the version:
40 l vertical air receiver
SMALLMIRAGE70
SMALLMIRAGE100
SMALLMIRAGE130
MIRAGE70
MIRAGE70E
MIRAGE100
MIRAGE100E
MIRAGE130
MOMIR70
MOMIR70E
MOMIR100
MOMIR100E
MOMIR130
MOMIR130E
TOP1.5
TOP1.5E
ECOSIL1
ECOSIL1E
All models are available also without a dryer.
If the device is enclosed in a cabinet, it will be equipped