6

ENGLISH

Step 4:

 Connect the other end of the Ethernet cable to the PC  

(if no Ethernet ports are available in the PC, use an Ethernet-to-USB adapter) .

Step 5:

 Connect the power (mains) cord to the power supply . The other 

end of the power (mains) cable goes into a grounded power outlet 
(not shown in the Figure) . To isolate the pod and thus the scanner from 
supply mains, unplug the power cord from mains .

Step 6:

 Insert a battery fully into the rear of the scanner . (Figure 5, S26) 

or Insert the TRIOS 4 cord into rear of scanner (Figure 5, S27) .

Step 7:

 Place the scanner in the pod .

Scanner Battery Charging (Figure 1, S9 and S10):

Before you start:

 Place the battery charger either on a flat surface hori-

zontally or attach it to the wall mount vertically, with battery containers 
pointing upwards .

Step 1:

 Attach the power cord and plug it to mains power .

Step 2:

 Push the batteries one by one into the charger until they click into 

place . When charging, the LED above each battery will indicate the status:
•  A pulsating green means that charging is ongoing;
•  Constant green means that the charging is completed;
•  Yellow means that an error has occurred, either the battery is not 

inserted correctly, or the charging is outside operating conditions  
for temperature .

CAUTION

If the charger continues to indicate yellow, the battery  
is damaged and should be discarded, unless the battery  
is too hot or too cold (see section 4 .8) .

4.8 Starting and Shutting Down TRIOS

Please follow the steps below to start TRIOS IOS system for the first time:

Step 1:

 Turn the system ON .

S3A: Press the standby button on the top of the backside of the display . 
The TRIOS software starts automatically .

S3P: Start the PC . If the PC has no TRIOS software pre-installed, please 
install the TRIOS software first . In case of difficulties with software instal-
lation, plase contact 3Shape support .

Step 2:

 Log on using a default TRIOS Operator .

Step 3:

 Configure and use your TRIOS IOS system .

For models S3P-8, S3A-10, S3A-18

The wired scanner turns on automatically when the MOVE/MOVE+ cart 
is turned on, or when the smart cord is plugged into the scanner .

For models S3P-2, S3A-14, S3A-22

To activate the wireless scanner, press the button on the top of the 
scanner when the MOVE/MOVE+ cart or computer is turned on .  
The scanner can be turned off manually by removing the battery .

Follow on-screen instructions for connecting to the wireless scanner .

When using the TRIOS IOS system, the user is guided by messages shown 
in the TRIOS software . This includes when calibration of the scanner  
is due . Other messages include a reminder to mount the scanning tip:  
“Please make sure the scanning tip is mounted on the scanner .”

Calibration

A calibration of the handheld TRIOS scanner is required before using TRIOS 
for the first time, or to compensate for thermal variation and aging effects 
due to wear . During regular use the TRIOS software will prompt the user 
when calibration is due . To perform the calibration, attach a cleaned and 
sterilized or high-level disinfected scanner tip on the scanner tube .

Remove the sleeve from the calibration target .

Mount the calibration target on the calibration adaptor  with the grey side, 
marked with the letter B, facing up .

Attach the calibration adaptor to the cleaned and sterilized or high-level 
disinfected scanner tip . Go to the Settings page in the TRIOS software, select 
TRIOS 

 Scanner Management 

 3D calibrate scanner, and then follow the 

on-screen instructions . Remember to save changes when prompted .

For a more detailed description of the 3D calibration, please refer to the 
TRIOS online help .

Color Calibration

A color calibration of the handheld scanner is required  
before each scan in order to get optimal colors .

Attach a cleaned and sterilized or high-level disinfected  
scanner tip on the scanner tube .

Remove the sleeve from the color calibration target .

Mount the color calibration target to the calibration adaptor  
with the color side, marked with the letter A, facing up .

Attach the calibration adaptor to the cleaned and sterilized  
or high-level disinfected scanner tip .

Go to the Settings page in the TRIOS software, select TRIOS 

 Scanner 

Management 

 Color calibrate scanner, and then follow the on-screen 

instructions, which will also include validation of the performed color cali-
bration . Remember to save changes when prompted .

For a more detailed description of the color calibration,  
please refer to the TRIOS online help .

Proper shutdown

Inside the TRIOS software, click on the Power button and select Shutdown 
or press the the standby button on the back of the monitor on the MOVE/
MOVE+ cart . System S3A (MOVE/MOVE+ cart) only: The light under the 
standby button turns off to confirm that the system is shut down .

CAUTION

When switching off the system, allow enough time for it to 
shut down properly before disconnecting the mains cable .

System S3A (MOVE/MOVE+ cart) only

IMPORTANT NOTICE

The standby button can also be used to force TRIOS IOS 
system to shut down in case the screen stops responding . 
To force a shutdown, push and hold the standby button 
pressed for about 5 seconds . Then, restart the system .

4.9 Starting TRIOS Online Help

The TRIOS online help can be started from the TRIOS software by clicking  
on the Help button and choosing User Manual .

Содержание TRIOS S3A-10

Страница 1: ...04 2021 3Shape TRIOS Safety and Setup Guide S3P 2 S3P 8 S3A 10 S3A 14 S3A 18 S3A 22 ...

Страница 2: ...ormity EU 3ShapeTRIOSA SherebydeclaresthattheTRIOSobservestheprovisionsof Medical Device Regulation EU 2017 745 MDR Restriction of Hazardous Substances in Electrical and Electronic Equipment Directive 2011 65 EU RoHS Waste Electrical and Electronic Equipment Directive 2012 19 EU WEEE Registration Evaluation Authorization and Restriction of Chemicals Regulation 1907 2006 EC REACH Radio Equipment Di...

Страница 3: ... Electronic Equipment Directive 2012 19 EU Radio Equipment Directive 2014 53 EU CAUTION EU only CE marking for connected products Further products which are connected to this unit must also bear the CE marking IMPORTANT NOTICE To comply with the CE marking of the integrated wireless module the wireless scanner must only be used indoors Electrical Safety This system has been tested to conform with ...

Страница 4: ...lt in PC 3 scanner batteries and a scanner battery charger You can find the figures in the final section of this guide All system variants comprise Handheld TRIOS intraoral scanner S3 2 S1 and S2 Wired and wireless handheld dental intraoral scanner for scanning the surface of the patient s teeth TRIOS 4 SmartTip TST 01 S3 Re usable scanner tip attached to the handheld TRIOS Scanner The scanner tip...

Страница 5: ...your service provider if there is visible damage or malfunctioning System S3A MOVE MOVE cart only Please follow the steps below for assembling TRIOS IOS system Figure 5 Step 1 Place the MOVE MOVE cart optimally in relation to the patient chair and your work position S24 Step 2 Attach the protection tip to the TRIOS scanner if not already in place S25 Step 3 Insert a battery fully into the rear of ...

Страница 6: ...r is due Other messages include a reminder to mount the scanning tip Please make sure the scanning tip is mounted on the scanner Calibration A calibration of the handheld TRIOS scanner is required before using TRIOS for the first time or to compensate for thermal variation and aging effects due to wear During regular use the TRIOS software will prompt the user when calibration is due To perform th...

Страница 7: ...ependent on local reg ulations and the configuration of the product 5 Safety and Warnings 5 1 Prerequisites CAUTION Read all instructions carefully including all warnings and caution notices You must comply with the warnings in this manual to prevent injury to persons and damage to equipment Proper functionality and safety can only be guaranteed if the safety precautions in this safety guide and o...

Страница 8: ...d it should immediately be calibrated before further use If calibration fails please contact your technical service provider See this guide s instructions on calibrating the handheld scanner WARNING If the handheld scanner is dropped examine the TRIOS IOS system for any mechanical damage If a mechanical damage is observed or suspected treat this as an equip ment failure see this guide s instructio...

Страница 9: ...ower cable network cable and the smart cord must never be subjected to pulling stress Spilled liquids WARNING Do not bring liquids such as beverages near the TRIOS IOS system Do not spill liquids on the TRIOS IOS system Condensation WARNING Variations in temperature or humidity may cause water to condense inside the system which can cause damage Always let the system reach room temperature before ...

Страница 10: ... S3P pod only WARNING Before each patient make sure any PC connected to the system and its peripherals are disinfected It is recom mended to use medical grade peripherals e g keyboards and mice that can be disinfected easily 5 7 Safety when scanning with TRIOS CAUTION The handheld scanner is a highly precise optical instrument and must be handled carefully Do not drop shake or bump the handheld sc...

Страница 11: ... manual for more information CAUTION New tips must be cleaned and sterilized high level disinfected before using them for the first time CAUTION A class B Vacuum autoclave according to EN 13060 is required for sterilization of the scanner tip CAUTION The scanner tip can be re sterilized up to 150 times and must be disposed afterwards as described in section Disposal CAUTION The scanner tip can be ...

Страница 12: ...riod of use IMPORTANT NOTICE All surfaces that the protection tip comes in contact with particularly the scanner tube and its optical window must always be cleaned and disinfected before attaching the protection tip 6 4 Additional cleaning in case of poor scan quality During scanning the tip mirror may become unclean resulting in a loss of scan quality In this situation you can clean the mirror fo...

Страница 13: ...h the correct rating and type 250 V T 3 15 A H must be used To replace the fuses first mount them in the drawer then insert the drawer back into the slot and push all the way in until the tabs click back into place and the drawer is no longer loose in the slot Do not attempt to insert the fuses directly into the slot on the MOVE MOVE cart The mains cable can now be reconnected and the TRIOS IOS sy...