5. Insert 3M™ Steri‑Gas™ EO Gas Cartridge, Load Chamber, and Close Door
Insert the 3M™ Steri‑Gas™ EO Gas Cartridge into the cartridge bay located inside the GSX Series sterilizer chamber (Figures 73) Push the green lever down over the Steri‑Gas EO Gas
Cartridge to secure in place (Figure 74)
Sterilizer
22.2
˚
C
98.0
kPa
Open door
Cycle: GSX 55C
Aeration (hh:mm): 12:00
Load ID: 120413-1
Cartridge lot number: 2017-02AB
Menu
Menu
Next
Next
Insert 3M
TM
Steri-Gas
TM
EO Gas Cartridge
Insert load
Close door
04/20/2015
05:26:08
6
4
5
2
3
4
1
7
Back
Back
Figure 72
Step Five: Insert Steri‑Gas EO Gas
Cartridge, Insert Load, Close Door
DANGER: To reduce the risks associated with exposure to ethylene oxide,
do not force the 3M™ Steri‑Gas™ EO Gas Cartridges into the cartridge holder as excessive force could damage the cartridge and result in a cartridge leak
CAUTION: To reduce patient risks associated with exposure to potentially non‑sterile devices or inadequate sterilization procedures,
do not force the 3M™ Steri‑Gas™ EO Gas Cartridges into the cartridge holder; excessive force could damage the cartridge and result in a cartridge leak