(D) It is necessary to tighten in a pattern similar to the one shown on this
picture. This will ensure that the end cap is inserted evenly.
•
While the end cap is pulled tight against the housing, place the V-Band
clamps on the ridges of housing and the end caps and tighten the clamps.
•
Loosen the threaded rods of the assembly tool and disconnect the tool
from the housing.
VACUUM INSERTION
•
(E) Inspect and clean the sealing surface of both end caps.
•
Place the L washer around the center nozzle and press down until it
touches the cap nut.
•
Place the center seal (see Assembly Part "d" on page 3) on the center
nozzle. Press it over the threads and down against the
L washer. Place the center seal washer (thinner washer) over the
threads and against the center seal (see Assembly Part "e" on page 3).
Screw the center seal nut onto the nozzle. Once it touches the washer,
loosen it one turn.
•
Bolt blind flanges on both lumen ports and one of the shell ports of the
end caps.
•
Lift the end cap and place center nozzle into center tube of the cartridge.
Slowly slide the end cap in until it touches the outer O-ring (lumenside).
Complete this step for both ends of the contactor.
•
Rotate the end cap to align the marks placed on the end cap and housing
during disassembly.
•
(F) Bolt the vacuum flange onto to remaining shell port.
Connect the vacuum hose to the flange and the vacuum.
D
E
F
L Washer
Center seal
washer
Center seal
Center
seal
nut
10