A. Determining Active Duct Probe Depth
Step 1. Place the tip of the receiver on the ground directly above the located ADP
position.
Step 2. Maintain the handle orientation perpendicular to the target path.
Step 3. Press
DEPTH
[5].
Step 4.
Press
Sonde Depth
[5] to read ADP Depth
−
The depth to the ADP is displayed in
units as specified in the receiver set up
menu.
−
Five Sonde depth readings can be saved
with the time, and date measured. Press
Mem Select
[SK] after the Sonde depth
displays.
−
Press
Mem Select
[SK] to select a
specific memory location (M1–M5) or
select
Save
[SK].
Save
will place each
entry in sequential order in memory
(M1–M5) until five readings have been
stored. The unit will overwrite saved
entries in excess of five, beginning
with M1.
Step 5. Optional step: Press
Clear All
[SK] to delete all saved depth readings.
Step 6. Press
Mem Select
[SK] to select a specific memory location (M1-M5) to store
the depth readings.
Step 7.
When the preferred location appears on the display, press
Save
[SK]. The
display and memory location will populate with the current information.
Step 8. Each memory location can be reviewed by pressing
Mem Select
[SK].
Step 9. Press
Cable Depth
[5] to switch to the cable depth display, or
Menu/OK
[6] to
return to Locate Mode.
Note: During a depth measurement, the display will exhibit ‘- -’ when the received
signal is too low, too high or erratic.