Troubleshooting

This section is intended to help solve problems that may be encountered when using the Clean-Trace Luminometer LX25 .

Contact 3M or your local 3M distributor for more information .

Problem

Probable Cause

Remedy

Diagnostic Check Failed

Chamber Door is open

Close the chamber door and retry the Diagnostic Check

Operating Clean-Trace

Luminometer LX25 out of

correct temperature range

Operate the Clean-Trace Luminometer LX25 between 5°C and

40°C . Retry the Diagnostic Check .

ATP Test UXC or H2O left in

ATP test chamber

Remove the ATP Test UXC or H2O from the test chamber and retry

the Diagnostic Check .
If the Diagnostic Check fails after a retry, contact 3M or your

local distributor .

Battery Low

Battery needs to be charged

Connect the power cord cable to the Clean-Trace

Luminometer LX25 and connect the cord to a power outlet to

recharge the battery .

Chamber Door Open

Chamber door was open

while attempting to start the

reading process

Close the chamber door after inserting a Clean-Trace ATP Test UXC

or H2O and before starting the reading process .

No Swab Detected

Insert a Clean-Trace ATP Test UXC or H2O .

Instrument temperature is not

within normal operating range

Instrument temperature is

>40°C or <5°C

 Use the Clean-Trace Luminometer LX25 at ambient temperature

between 5°C and 40°C . If the problem persists, contact 3M or your

local distributor .

Swab Left in Chamber

The Clean-Trace Luminometer LX25 has a sensor to determine if a

Clean-Trace ATP Test UXC or H2O has been left in the chamber for

more than 30 seconds following the completion of a measurement .

If this situation is detected, the instrument will display a “Swab

Left In Chamber” error message on the screen and provide an

audible beep to remind the user to remove the Test following

the measurement .

No Display

Battery fully discharged

Recharge the battery .

Clean-Trace

Luminometer LX25 is in Sleep

Mode or has been turned off .

Press the “Up” button on the cursor control to wake from sleep

mode . If no response, press the “Power” button for 3 seconds to

turn on the Clean-Trace Luminometer LX25 .

Instrument malfunction

Contact 3M or your local distributor .

Results Lower than Expected

Sample / reagent problem

Check to ensure correct sample technique was used and ATP Test

UXC or H2O activation was followed according to Clean-Trace ATP

Test instructions for use . Verify with known positive samples such

as a 3M Positive Control Test .

Instrument malfunction

Contact 3M or your local distributor .

Results Higher than Expected

Direct sunlight or static

discharge can cause ATP Test

to phosphoresce and cause

high readings

After exposure to sunlight or static discharge an immediate repeat

reading is 50% of the first reading, if this is the cause of the higher

than expected results . Subsequent readings become incrementally

lower and finally return to normal .

Avoid exposing ATP Test UXC or H2O to direct sunlight or

static discharge .

Other Errors

1 . Turn off the Clean-Trace Luminometer LX25 and power back

on again .

2 . Retry ATP Test UXC or H2O using the Clean-Trace

Luminometer LX25 .

3 . If the problem persists, contact 3M or your local distributor .

Содержание Clean-Trace Luminometer LX25

Страница 1: ...ETER LX25 7 HOW TO PERFORM A QUICK TEST 7 COMMUNICATING WITH THE 3M QUALITY CONTROL DATA MANGER QCDM 7 REGISTER A CLEAN TRACE LUMINOMETER LX25 TO AN ORGANIZATION ON QCDM 7 ROUTINE SYNCHRONIZATION 8 DE REGISTRATION 8 WI FI SET UP 9 VERIFYING DATA UPLOAD TO QCDM 9 APPLICATION WORKFLOW 9 STARTING THE AUDIT 9 HOME SCREEN 10 1 FLEXIBLE ENDOSCOPE WORKFLOW 10 2 ENVIRONMENTAL SURFACES WORKFLOW 11 3 SURGIC...

Страница 2: ...ided by the product may be impaired Explanation of Signal Word Consequences WARNING Indicates a hazardous situation which if not avoided could result in serious injury or death CAUTION Indicates a hazardous situation which if not avoided could result in minor or moderate injury and or property damage NOTICE Indicates a situation which if not avoided could result in property damage Explanation of S...

Страница 3: ...components with 3M designated replacement parts only To reduce the risks associated with illness or infection Do not use the Clean Trace Luminometer LX25 and 3M Clean Trace ATP Tests in the diagnosis of conditions in humans or animals Do not use the Clean Trace Luminometer LX25 and 3M Clean Trace ATP Tests as an indication of disinfection or sterilization of surfaces or devices CAUTION To reduce t...

Страница 4: ...lly inspect the Clean Trace Luminometer LX25 for damage including cracks or swelling Do not store or use the Clean Trace Luminometer LX25 or power supply in a location exposed to steam high humidity dust or strong vibration Reference the Environmental Operating Conditions table in the Important Safety Information for Clean Trace Luminometer LX25 in this User Manual Use only 3M Clean Trace ATP Test...

Страница 5: ...ntation THIS WARRANTY IS VOID IF A THE HARDWARE HAS BEEN REPAIRED BY PERSONS NOT AUTHORIZED BY 3M OR B THE HARDWARE HAS BEEN ALTERED MODIFIED OR MISUSED OR C THE HARDWARE IS USED WITH PRODUCTS SUPPLIES COMPONENTS OR SOFTWARE NOT SUPPLIED BY 3M FOR USE WITH THE HARDWARE OR D THE HARDWARE OR A COMPONENT IS USED FOR OTHER USES FOR EXAMPLE USE WITH OTHER CIRCUIT BOARDS OR SOFTWARE OR E THE HARDWARE HA...

Страница 6: ...ing less than 10 charge capacity The battery does not need to be fully discharged before recharging When storing a Clean Trace Luminometer LX25 for more than a month the battery level should be at 50 or less charge Do not store a Clean Trace Luminometer LX25 with a fully discharged battery because the battery may lose the ability to be recharged Battery replacement is recommended after two years o...

Страница 7: ...n Trace ATP Test UXC or H2O Instructions for Use to collect the sample activate the ATP Test and insert into the Clean Trace Luminometer LX25 Tap Start to take a measurement Remove the ATP Test UXC or H2O after the measurement is complete All results should be recorded o Only the last 20 Quick Tests are saved on the Clean Trace Luminometer LX25 o The results of a quick test are not saved to the Qu...

Страница 8: ...sing the USB cable 5 After a short period a red spinning circle will appear on the Clean Trace Luminometer LX25 indicating the Clean Trace Luminometer LX25 is connected to the Synchronization Manager 6 On the workstation log into Synchronization Manager Software 7 Select the Start Synchronization button on the workstation Synchronization Manager Software 8 Start the Synchronization on the workstat...

Страница 9: ...d An unsuccessful upload will result in a triangle If using the Synchronization Manager Software to upload results refer to section on Routine Synchronization of Results in this document All results will be marked with a triangle until LX25 results are synchronized and data is uploaded to the QCDM software Application Workflow Before access to the workflow screens is available an organization hier...

Страница 10: ...Workflow 3 Surgical Instrument Workflow 1 Flexible Endoscope Workflow From the Home screen tap Start to begin new test or select a recent failed audit for retest Select device type Select Device This is the serial number or other unique identifier Select Reprocessing Step Review selections o Select technician o Can change device reprocessing step and test plan o Tap Perform Audit Select test point...

Страница 11: ... Select Room Review Selections o Can change Room number test plan o Select Housekeeper and Isolation room option o Tap Perform Audit Choose test point Insert ATP Test UXC or H2O Press Begin Measurement and wait until blue Reading Result screen disappears then press Return Repeat process until all test points have been tested Press Done ...

Страница 12: ...lt screen disappears then press Return Repeat process until all test points have been tested Press Done Performing a Retest To retest a failed test point From the home screen select the audit that requires retesting marked with a red X Select the test point that requires retesting marked with a red X Follow the prompts and select any additional test points if required Complete the retest Select Do...

Страница 13: ... Clean Trace Luminometer LX25 System Diagnostics tap to initiate device diagnostic test Settings menu Settings Menu Functions detail o Wi Fi Settings turn Wi Fi on off set up network o Server settings select or change server connection o Screen brightness increase or decrease screen brightness o Screen timeout choose length of time before screen goes to sleep o Date set current date o Time set cur...

Страница 14: ...an Trace ATP Test UXC or H2O has been left in the chamber for more than 30 seconds following the completion of a measurement If this situation is detected the instrument will display a Swab Left In Chamber error message on the screen and provide an audible beep to remind the user to remove the Test following the measurement No Display Battery fully discharged Recharge the battery Clean Trace Lumin...

Страница 15: ...Clean Trace Luminometer LX25 Make sure the towel used is just damp not dripping 3 When cleaning the Clean Trace Luminometer LX25 avoid the USB cable connection on the side of the Clean Trace Luminometer LX25 4 Rinse disinfectant from Clean Trace Luminometer LX25 by repeating the procedure 2 3 using another disposable towel slightly dampened with distilled or deionized water 5 Allow Clean Trace Lum...

Страница 16: ... St Paul MN 55144 USA www 3M com Medical 3M and Clean Trace are trademarks of 3M Used under license in Canada 2017 3M All rights reserved Issue Date 2017 04 34 8720 2654 6 3M Deutschland GmbH Health Care Business Carl Schurz Str 1 41453 Neuss Germany ...

Отзывы: