With a processed 3M™ Attest™ Rapid Readout Biological Indicator, a positive (+) result

indicates a sterilization process failure. Act on any positive test results immediately.

Determine the cause of the positive 3M™ Attest™ Rapid Readout Biological Indicator and

retest the sterilizer following current facility policies and procedures.

Optional Visual pH Color Change Result

3M™ Attest™ Rapid Readout Biological Indicators 1295 for Vaporized Hydrogen Peroxide

Sterilization may be further incubated for a visual pH color change.  Due to the high

sensitivity of the fluorescent result, there is no advantage to continuing to incubate 3M™

Attest™ Rapid Readout Biological Indicators 1295 after the fluorescent result has been

determined and documented. This optional further incubation is not intended for routine

use but may be used for special studies.
In the case of a positive control BI, a yellow color change of the growth media will appear

within 24 hours. Any observation of a yellow color within the vial indicates a positive

result. The pH color change result can be recorded and the BI discarded as soon as the

yellow color is observed and the result recorded.
For processed 3M™ Attest™ Rapid Readout Biological Indicators 1295, the use of a

humidified incubator is recommended to avoid media dry out. If a visual pH color change

result is desired, transfer the processed 1295 BI to a humidified incubator after the

fluorescent result has been determined in the 3M™ Attest™ Auto

reader 490H. A media

color change from purple to yellow indicates a sterilization process failure. A negative pH

color change result, i.e., media remains purple, can be assessed at 7 days.

INSTRUCTIONS FOR USE

 DISPOSAL

Instrument contains a lithium battery. Dispose of instrument in accordance with federal,

state, and local requirements.

FEATURES OF THE 3M

ATTEST

 AUTO

READER 490H

Audible Alarm

An audible alarm will sound, the LCD display backlight will flash, and a (+) symbol will

be displayed on the LCD panel directly under the incubating BI when a positive result is

detected. An alarm will also sound if a 3M™ Attest™ Rapid Readout Biological Indicator

is removed from the 3M™ Attest™ Auto

reader 490H well before a final result has been

detected. Data will be lost if the 3M™ Attest™ Rapid Readout Biological Indicator is not

replaced in the incubation well within 10 seconds.

Alarm On/Off

The audible alarm for positive results can be turned off for a particular 3M™ Attest™

Rapid Readout Biological Indicator. This is accomplished by pressing the

Buzzer Alarm

OFF

 button after the alarm has sounded. This will turn off the alarm for that BI only. Any

positive 3M™ Attest™ Rapid Readout Biological Indicator results subsequently detected

in other wells will result in the alarm sounding. Removing the 3M™ Attest™ Rapid

Readout Biological Indicator will reactivate the alarm for that well. Pressing the

Buzzer

Alarm OFF

 button will not silence the alarm for error or caution code alarm warnings. To

disable the alarm, see configuration section.

Stackable

To save counter space, 3M™ Attest™ Auto

readers 490Hs may be stacked two high.

Information Button

When the Information button is pressed once the 3M™ Attest™ Auto

reader 490H

displays two lines of text in the LCD display.
The top row indicates (left to right):
• 3M™ Attest™ Auto

reader 490H serial number

• 3M™ Attest™ Auto

reader 490H embedded software version (V X.Y.Z)

• Current date and time.
The bottom row indicates (left to right):
• The BI type configuration for wells 1

5 (i.e., 1295)

• The BI type configuration for wells 6

10 (i.e., 1295)

Press the information button twice and the 3M™ Attest™ Auto

reader 490H LCD display

will present two additional lines of text.
The top row indicates:
• IP address (if unit is connected to an Ethernet jack).
The bottom row indicates (left to right):
• Date (YYYY

MM

DD) and time that power was supplied to the 3M™ Attest™

Auto

reader 490H.

Press the information button three times and the 3M™ Attest™ Auto

reader 490H LCD

display will present the unique Media Access Control (MAC) Address for the device.

Temperature Monitoring

The 3M™ Attest™ Auto

reader 490H contains an internal temperature self

diagnostic

program. Should the temperature fall outside the specified range (60ºC +/

2ºC) an error

code will be displayed. The heater block temperature may be checked by depressing the

Temperature button once.

Easy Configuration

The 3M™ Attest™ Auto

reader 490H may be configured using either the buttons located

on the front of the unit or using the optional 3M™ Attest™ Auto

reader 490H Web App.

Optional 3M

Attest

 Auto

reader 490H Web App

If desired, the 3M™ Attest™ Auto

reader 490H may be connected to a PC via an Ethernet

connection for additional capabilities:
• Remotely view current status of BIs in the 3M™ Attest™ Auto-reader 490H via

networked PC or mobile hand held devices. For full functionality, the use of an HTML5

compliant browser with JavaScript enabled is required.

• Remotely view and configure 3M™ Attest™ Auto

reader 490H settings.

• View the last 100 BI records.

• Print out documentation using a local or networked printer and/or save documentation

as an electronic file.*

Note:

To access the 3M™ Attest™ Auto

reader 490H Web App within a secure network,

the PC or mobile device must have access to the secure network; options include

connecting via a network Ethernet cable, signing in to the facility

s wireless network, or

connecting via a virtual private network (VPN).
* Refer to your facility

s record keeping policy and procedure.

Содержание Attest Auto-reader 490H

Страница 1: ...io 5Manual del operador 6Bedieningshandleiding 7Användarhandbok 8Brugsvejledning 9Brukerhåndbok 0Käyttöopas Manual do Utilizador Manual do Operador Εγχειρίδιο Χειριστή HInstrukcja obsługi FPříručka pro obsluhu SPriročnik za uporabo MИнструкция по эксплуатации Upute za korištenje dOperatör Kılavuzu I操作指南 jคู มือปฏิบัติงาน T작업자 매뉴얼 ...

Страница 2: ......

Страница 3: ...2 Product Description 2 Warranty 2 Safety and Precautions 2 Health and Safety Information 3 Explanation of Product and Packaging Label Symbols 3 Specifications 4 Instructions for Use 4 Features 7 Care and Cleaning 8 Troubleshooting Guide 8 Optional 3M Attest Auto reader 490H WebApp 9 Troubleshooting the 3M Attest Auto reader 490H WebApp 12 Repair and Replacement 12 ...

Страница 4: ...fluorescence reading or a positive visual pH color change result indicate a sterilization process failure However due to the high sensitivity of the fluorescent result there is no advantage to continuing to incubate processed 3M Attest Rapid Readout Biological Indicators 1295 once the fluorescent result has been documented Other biological indicators BIs are not compatible with this device and can...

Страница 5: ...H complies with the CE mark related to the Low Voltage Directive LVD 2006 95 EC as confirmed in the Declaration of Conformity The 3M Attest Auto reader 490H complies with the RoHS Directive Directive 2011 65 EU of the European Parliament and of the Council of 8 June 2011 on the restriction of the use of certain hazardous substances in electrical and electronic equipment The 3M Attest Auto reader 4...

Страница 6: ...sed to sunlight or strong incandescent light Do not place instrument close to any device that emits a strong electro magnetic field Do not use on a vibrating surface Select Install Correct Power Plug The 3M Attest Auto reader 490H power supply comes with multiple blade configurations Install the blade configuration appropriate for your country into the power supply and plug into appropriate AC gro...

Страница 7: ...etting Setting the 3M Attest Auto reader 490H Clock The 3M Attest Auto reader 490H date and time are displayed in this format YYYY MM DD HH MM with the time displayed using a 24 hour clock To change the year if necessary push the up or down buttons to set desired year then press to advance to the month adjust as necessary and press to advance to the day Use the button to move to the time and set t...

Страница 8: ...ce the result is determined it is continuously displayed on the LCD panel until 10 seconds after the BI has been removed from the incubation well Note After the first minute of incubation if 3M Attest Rapid Readout Biological Indicator is inadvertently removed before the incubation period is complete an alarm will sound and a C2 Caution code will be displayed on the LCD panel under that well You m...

Страница 9: ... positive 3M Attest Rapid Readout Biological Indicator results subsequently detected in other wells will result in the alarm sounding Removing the 3M Attest Rapid Readout Biological Indicator will reactivate the alarm for that well Pressing the Buzzer Alarm OFF button will not silence the alarm for error or caution code alarm warnings To disable the alarm see configuration section Stackable To sav...

Страница 10: ...unit and plug it back in to clear error If error doesn t clear call for repair or replacement E3 Memory Error 3M Attest Auto reader 490H unable to save read data from memory Unplug the unit and plug it back in to clear error If error doesn t clear call for repair or replacement E7 Communications Error This error will occur if internal communications have failed Unplug the unit and plug it back in ...

Страница 11: ...ddress URL Note To access the 3M Attest Auto reader 490H Web App within a secure network the PC or mobile device must have access to the secure network options include connecting via a network Ethernet cable signing in to the facility s wireless network or connecting via a virtual private network VPN A Serial number URL 1 Connect the supplied Ethernet cable from the 3M Attest Auto reader 490H to a...

Страница 12: ...er 490H Web App Status History 3M Attest Auto reader 490H ATTEST 1295 ATTEST 1295 Auto reader 490H 1 2 3 4 5 24 6 7 8 9 10 SN 200299 2017 05 11 17 13 1 Open the Results Detail page After inserting a new 3M Attest Rapid Readout BI in the 3M Attest Auto reader 490H it is a best practice to open the 3M Attest Auto reader 490H Web App Results Detail page for that BI by clicking on the appropriate well...

Страница 13: ...k on the desired radio buttons to change settings Language Selected setting indicates language in which text on 3M Attest Auto reader 490H Web App pages will appear on the PC mobile device screen Pushbutton Configuration Selecting Enable allows the user to configure the 3M Attest Auto reader 490H from the buttons on the unit itself Selecting Disable prevents the user from being able to configure t...

Страница 14: ...cility virtual private network VPN on the mobile device Data entered by user on Results Detail page disappears User entered data is written to temporary web browser storage Browser may have cleared data Set web browser to accept allow or enable cookies As a best practice print or save Results Detail record as soon as BI result is final BI result displayed on 3M Attest Auto reader 490H LCD display ...

Страница 15: ......

Страница 16: ...t Paul MN 55144 1 800 228 3957 3M com Medical 3M Deutschland GmbH Health Care Business Carl Schurz Str 1 41453 Neuss Germany 2017 3M All rights reserved 3M and Attest are trademarks of 3M Used under license in Canada Issue Date 2017 03 34 8720 4946 4 ...

Отзывы: