Section 7.
Performance Verification

The 963/963E Benchtop Air Ionizers
are factory adjusted to provide
optimum performance.  Further
adjustment in the field is not possible.
However, the following instructions
can be followed to determine whether
the Ionizer is performing to
specification.  The testing follows the
procedure outlined in the standard for
Ionization, ANSI/ESD S3.1.  Please
refer to this standard for more
complete information.

7.1 Equipment Needed

3M 711 Charge Analyzer or
equivalent charge plate monitor
(CPM).  If an alternate CPM is
used, please refer to its Operating
Manual for details on how to
perform the following instructions.

7.2 Static Discharge Time

The 963/963E Benchtop Air
Ionizer will reduce the charge on
the 6" x 6" square isolated metal
plate on the CPM from ± 1000
volts to ± 100 volts in less than 1
second (High fan speed).  The
metal plate for the CPM must be
located at a distance of one foot
(30cm) from the ionizer and
centered in the air stream.

7.21 Attach the flat plate electrode to

the 711.  Refer to the owner's
manual for the 711 for complete
operating instructions.

7.22 Place the 711 on its side allowing

for viewing of the display.
Position the 711 so that the plate
is parallel to the ionizer at a
distance of one foot.  The plate of
the 711 should be centered (up &
down, left & right) in the air
stream. It may be necessary to

raise the ionizer from the surface
to allow for centering on the
711charge plate.  If so, use a
block of appropriate height to
elevate the 711.  Please note that
the plate must be kept totally
isolated from ground and that the
edge of the plate should be raised
up a minimum distance of 3
inches from the work surface.
This is illustrated in Fig. 11 of
ANSI/ESD-S3.1.  

7.23 Turn on the ionizer at high speed

and allow it to run for five
minutes. 

7.24 Charge the plate positive as

described in the 711 operating
instructions for "Static Decay
Time" mode.  Observe the
discharge time indicated on the
711.  Repeat this step for
negative polarity.

7.3 Ion Balance

The ionized air blower will stay
within an offset voltage of +/-15
volts (max. deviation from zero)
at a distance of one foot, when
measured using the following
procedure.  

7.31 Position the ionizer and CPM as

stated above in Section 7.22.

7.32 Turn on the ionizer at high speed

and allow it to run for five
minutes.

7.33 Use a ground wire to ground the

charge plate of the 711.  This will
remove any/all residual charge
present on the charge plate.  If
the CPM does not zero, adjust
the zero control.

7.34 Remove the ground wire and

observe the display on the 711.
The voltage (either ±) observed
during this time is the "offset
voltage" and is a measure of
instantaneous ion imbalances
produced by the ionizer.

Содержание 963

Страница 1: ...3 963 963E Benchtop Air Ionizer Instructions ...

Страница 2: ...age Safety Information 4 1 Description 5 2 Performance 5 3 Power Requirements 5 4 Installation 5 5 Operation 6 6 Maintenance 6 7 Performance Verification 7 8 Physical Characteristics 8 9 Customer and Technical Service 9 ...

Страница 3: ...ts When cleaning the 963 963E housing or emitter points verify that the unit is off and disconnected from electrical power Collected dirt on this unit should be removed regularly to prevent accumulation that may result in fire Class 1 equipment Explanation of Symbols Caution refer to user instruction manual High voltage present inside unit Read and understand all safety information before installi...

Страница 4: ...ican style three prong plug The 963E requires DC 24V power which is supplied through a Mini DIN connector on the back of the unit The 963E is packaged with a universal power supply capable of converting AC 100V 240V 50 60 Hz into DC 24V The universal power supply uses a 3 0 9 m cord to connect to the ionizer and has an IEC 320 input socket for incoming power Please note that due to the multitude o...

Страница 5: ...of the ON positions and that the ionizer is now in use Section 6 Maintenance Occasional cleaning of the case and of the emitter points are the only routine maintenance procedures required 6 1 Cleaning the case wipe the case with a soft cloth moistened with water If a stronger cleaning solution is required mild detergent or alcohol may be used Do not use solvents that will attack the plastic case 6...

Страница 6: ...e foot The plate of the 711 should be centered up down left right in the air stream It may be necessary to raise the ionizer from the surface to allow for centering on the 711charge plate If so use a block of appropriate height to elevate the 711 Please note that the plate must be kept totally isolated from ground and that the edge of the plate should be raised up a minimum distance of 3 inches fr...

Страница 7: ...ni DIN plug Power Outlet Cord 6 1 8m cord with IEC 320 Not Included and NEMA 5 15 plugs Dimensions 7 0 W x 9 H x 4 D 7 0 W x 9 H x 4 D w mounting base 18 cm W x 23 cm H x 10 cm D 18 cm W x 23 cm H x 10 cm D Weight 2 5 lb 1 1 kg 2 5 lb 1 1 kg Air Volume low speed 68 SCFM 1 92 m3 min 76 SCFM 2 15 m3 min high speed 105 SCFM 2 97 m3 min 112 SCFM 3 17 m3 min Static discharge time 1 ft 30 cm 1 second 1 ...

Страница 8: ... can be obtained by calling the 3M Electronic Handling Protection Division Customer Service 800 328 1368 Technical Support 512 984 3200 Outside of the U S For customer service and technical support please contact your local representative of the 3M Electronic Handling Protection Division ...

Страница 9: ...10 ...

Страница 10: ... Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES EXPRESS AND IMPLIED INCLUDING FITNESS AND MERCHANT ABILITY The express warranty will not apply to defects of damage due to accidents neglect misuse alterations operator error or failure to properly maintain clean or repair products Limit of Liability In no event will 3M or Seller be responsible or liable for...

Страница 11: ...Litho in the USA 3M 2000 34 7049 7753 6 Recycled paper 40 pre consumer 10 post consumer 3 Electronic Handling Protection Division 6801 River Place Blvd Austin TX 78726 9000 800 328 1368 www 3m com ehpd ...

Отзывы: