manualshive.com logo in svg
background image

onlinecomponents.com

5

78-8136-2118-8

6.0  Calibration Procedures

The 3M™ Wrist Strap Tester 746 cannot be adjusted, 
but the following steps can be used to determine that the 
tester is operating within its specification.

Equipment needed:

1.   Resistance Substitution Box 

(750 k

Ω

 - 11 M

Ω

, tolerance < 2%)

2.  2 leads (to connect the resistance box)

Wrist strap test circuit check:

Connect the resistance substitution box to the 746 as 
shown in figure 2.
Set the resistance substitution box to each value of 
resistance indicated in the test table and press contact 
plate A.
The LEDs will indicate as shown below if the unit is 
functioning within specifications.

Resistance 

substitution box 

settings

Resistance 

range of 

the 746

LED  

indication

*750 k

Ω

900 k

Ω

750 k

Ω

red

green

9.0 M

Ω

 

11.0 M

Ω

10 M

Ω

green

red

*This example is used in figure 2.

746

wrist strap tester

o.k.

o.k.

A

700 k

900 k

9.0 M

11.0 M

Resistance substitution box

Crocodile clip

figure 2