4

78-8136-2118-8

4.0  Specifications

Base unit: 

158 x 95 x 33 mm (6.25 x 3.75 x 1.25 inches)

Weight: 

200 g (7 ounces) without battery

Battery: 

9V alkaline type, e.g. IEC 6F22

Tolerance: 

750 k

Ω

 + 20%/-0%, 10 M

Ω

 ± 10%

Measurement Voltage:

 

DC (19±1) V (open circuit)

Power Supply 

(not included):

DC 9V, regulated, 75mA (50Hz)

5.0  How to Use

5.1 

Insert 9V battery and make sure the ON/OFF 
switch at the upper left side on the unit is in 
the ON position. The OFF position avoids 
discharge of the battery if the metal contact plate 
is accidentally depressed while the unit is not 
in use. The unit is then ready for operation. The 
backup of the battery can be achieved by using 
an AC Power Supply. In that case, connect the 
AC power supply to the jack at the base of the 
unit. Plug the power supply into an AC 230V 
outlet.

5.2 

Fit the plastic cover onto the wrist strap plug-in 
jack that is not being used. If the "battery" LED 
illuminates, replace it immediately.

Wrist Strap Test

5.3 

Put the wrist band on with ground cord attached, 
and insert the ground cord plug into the wrist 
strap plug-in jack.

5.4 

Depress the metal contact plate and hold it.

5.5 

One of the indicator LEDs will illuminate. 
An illuminated green LED indicates that the 
wrist strap performs within the resistance range 
of 750 k

Ω

 to 10 M

Ω

. If a "red" condition is 

indicated, it is advisable to check the ground cord 
individually for its resistance.

 

For checking the resistance range of the ground 
cord, leave the plug end of the ground cord 
attached to the tester and unsnap the cord from 
the wrist band. Place the metal side of the ground 
cord snap against the metal plate of the tester 
and depress until one of the LEDs is illuminated. 
If the red LED is illuminated, the ground cord 
is outside the resistance range and should be 
replaced. If the green LED illuminates, the 
ground cord can be used, but the wrist band may 
have to be replaced.

Note: In some cases, high contact resistance between 
the person's skin and wrist band will cause the tester 
to show a fail/red LED condition. This resistance may 
be caused by dry skin or hair in the wrist area. If these 
possibilities do not match to the operator, the wrist 
band should be replaced.

5.6 

If the "battery" LED illuminates at any time, 
replace the battery.

Wall Mounting

5.7 

The enclosed 3M™ Dual Lock™ adhesion discs 
shall be screwed on the wall using the attached 
template approximately 1500 mm (5 feet) 
parallel off the floor. Drill three holes of 5 mm 
(0.2 in) diameter at marked location.

5.8 

The enclosed 3M Dual Lock adhesion strips must 
be adhered to the rear of the 3M™ Wrist Strap 
Tester 746 parallel to the bottom edge and below 
the battery compartment.

Note:  Make sure the area on the unit in which 

the Dual Lock adhesion strips are placed is 

free of dust and dirt. Press the tester firmly 

against the discs for fixing on the wall.

5.9 

For removal, lift the tester 746 at the two top 
corners with both hands.