6.0 Physical Characteristics (All values typical)
(at 1 inch (2.54 cm) distance from sensor to target)
3M
™
Static Sensor 718
Description
Typical Value
Dimensions
0.85" H x 2.4" W x 4.2" L
2.2 cm H x 6.1 cm W x 10.6 cm L
Weight
4.4 oz. (125 g) with battery
Operating Conditions
41º F - 95º F (5º C - 35º C)
Up to 80% RH, non-condensing
Indoor use only
For use at altitudes below 42, 300 ft. (2,000 m)
Pollution Degree II
Class III
Enclosure
Conductive Housing
Power Requirements
One 9-volt alkaline battery
Measurement Ranges
0 – 1,990 kV
0 – 19,900 kV
Voltage Display
Liquid Crystal Display
Display Resolution
1V/inch (0.39V/cm) @ low range
10V/inch (3.9V/cm) @ high range
± 10 counts
5.5.6 The instrument should now be reading
1.000. If it is not, remove the battery
door and use a screwdriver to turn the
small screw located inside the battery
compartment. Use a small screwdriver to
turn the small screw, located on the front
right side of the instrument. This screw
should adjust the reading on the display.
Once the display has been adjusted to read
1.000, the low range of the 3M
™
Static
Sensor 718 is now calibrated.
5.5.7 Repeat procedures 5.5.2 - 5.5.6 for the high of the meter, using a test
voltage of 5,000 volts.