5.4 Test fixture:

Metal plate

of at least 5 square

inches area (38.7 cm

2

) area.

Metal stand

capable of supporting

718, and holding it one inch (2.54
cm) away from the metal plate,
centered.

Connectors

on the plate with which

it can interface with the high voltage
power supply.

Connections

on the metal stand with

which it can be connected to
electrical ground.

5.5 Procedure:
5.5.1 Place the 3M™ 718 Static Sensor on

the metal stand.  Verify that it is
exactly one inch (2.54 cm) away from
the metal plate, and that its position is
centered relative to the plate.

5.5.2 Connect the stand to ground.
5.5.3 Turn on the 718 Static Sensor and

set it to the 2 kV range.

5.5.4 Ground the plate.  Zero the display.
5.5.5 Remove the ground from the plate,

and connect it to the high voltage
power supply.  Apply a 1 kV charge
to the plate.

5.5.6 The instrument should now be

reading 1.000.  If it is not, remove
the battery door and use a
screwdriver to turn the small screw
located inside the battery
compartment. Use a small
screwdriver to turn the small screw,
located on the front right side of the
instrument.  This screw should
adjust the reading on the display.
Once the display has been adjusted
to read 1.000, the low range of the
718 is now calibrated.

5.5.7 Repeat procedures 5.5.2 - 5.5.6  for

the 20 kV range of the meter, using
a test voltage of 5,000 volts.

Содержание 718

Страница 1: ...fying the operation of air ionizers by measuring the neutralization time for a static charge placed on a stationary metal plate Any deviation from this intended use could impair the instrument s effec...

Страница 2: ......

Страница 3: ...lied by a 9V alkaline battery 2 2 To install the battery on either unit 2 2 1 Remove the battery cover located in the lower back of the unit To do this press down on the cover and slide it downward 2...

Страница 4: ...ront of the sensor are lit To shut off the 718 Static Sensor momentarily press down on the membrane switch labeled POWER Determining that the power is off can be verified by seeing that the liquid cry...

Страница 5: ...ight emitting diodes LED s are present on the front face of the instrument These LED s emit two red bullseye targets on the surface of the object being measured As the instrument gets closer to the on...

Страница 6: ...disabled The Automatic Shut Off feature will reset itself the next time the instrument is turned on 4 0 Operation and Use of the 718A Air Ionizer Test Kit It is recommended that the user be familiar...

Страница 7: ...er s operating manual or consult with the ionizer manufacturer to determine what this offset balance should be 5 0 Service Calibration 5 1 Service and Repair in the event that you believe the 718 Stat...

Страница 8: ...he stand to ground 5 5 3 Turn on the 718 Static Sensor and set it to the 2 kV range 5 5 4 Ground the plate Zero the display 5 5 5 Remove the ground from the plate and connect it to the high voltage po...

Страница 9: ...ge Automatic Shut off 20 minutes after last switch activity Operating Time for Battery Greater than 50 hours with new battery 21 C Continuous usage Distance Indicator LED targets Aligned targets indic...

Страница 10: ...ual 1 ea Certificate of Conformance 8 0 Contact Information Customer and Technical Service Within the U S Customer service and technical support can be obtained by calling the 3M Electronic Handling P...

Страница 11: ......

Страница 12: ...less expressly agreed upon in writing by an authorized officer of 3M Warranty Limited Remedy Limited Liability This product will be free from defects in material and manufacture for a period of one ye...

Отзывы: