replace the battery.  Usage of the
3M™ 718A Charger under low-
battery conditions would lead 
to insufficient voltage levels 
being generated.

3.0 Operation of the 718 Static

Sensor

Supplemental notes:
For accurate measurements during usage,
it is recommended that the outside housing
of the 3M™ 718 Static Sensor be
connected to an electrical ground.  This
can be accomplished by having the user
holding the instrument connected to
ground through either a static control wrist
strap, or while wearing static control
footwear.  The enclosure of the 718 is
made of conductive plastic and is,
therefore, electrically connected to
whomever is holding the instrument.  In
addition, ground for the instrument can be
provided through the Voltage Monitor
Output on the front of the case.

The 718 Static Sensor is a precision
electronic instrument.  Improper use or
rough treatment can damage the unit, and
render it incapable of providing accurate
measurements.

3.1 Turning the unit ON and OFF: To

turn on the 718 Static Sensor,
momentarily press down on the
membrane switch labeled POWER.
Determining that the power is on can
be verified by seeing that the liquid-
crystal display (LCD) is on, and that
the red light-emitting diodes in the
front of the sensor are lit.  To shut off
the 718 Static Sensor, momentarily
press down on the membrane switch

labeled POWER.  Determining that
the power is off can be verified by
seeing that the liquid-crystal display
(LCD) is off.

3.2 Making Electrostatic Voltage

Measurements: hold the instrument 1-
inch (2.54 cm) away from the object
being measured.  The display will
update with the voltage measurement
in kilovolts.  If the measured voltage
is greater than the measurement range
of the instrument, a  

–1.

will be

displayed.  At this time, switch to

a greater range.  If over-ranging

occurs even with the 20 kV range
activated, the static charge on the
object cannot be measured with the
718 Static Sensor.

3.3 Measurement Range: all measurements

are in kilovolts (kV) as stated on the
front label of the unit.   The 718 has
two measurement ranges:  0-2 kV and
0-20 kV.  The unit’s current
measurement range mode can be
verified by checking the display.  
Three digits following the decimal
point indicate that the unit is in 0-2 kV
range.  Two digits following the
decimal point indicate that the sensor
is in 0-20 kV range.  To change
between measurement ranges, press 
the RANGE/HOLD button once,
momentarily.

3.4 HOLD Function: in the event that the

user wishes to freeze the current
measurement, the HOLD function of
the 718 Static Sensor may be used.
Simply press the HOLD switch
momentarily and the currently
displayed voltage will be frozen.  

Содержание 718

Страница 1: ...fying the operation of air ionizers by measuring the neutralization time for a static charge placed on a stationary metal plate Any deviation from this intended use could impair the instrument s effec...

Страница 2: ......

Страница 3: ...lied by a 9V alkaline battery 2 2 To install the battery on either unit 2 2 1 Remove the battery cover located in the lower back of the unit To do this press down on the cover and slide it downward 2...

Страница 4: ...ront of the sensor are lit To shut off the 718 Static Sensor momentarily press down on the membrane switch labeled POWER Determining that the power is off can be verified by seeing that the liquid cry...

Страница 5: ...ight emitting diodes LED s are present on the front face of the instrument These LED s emit two red bullseye targets on the surface of the object being measured As the instrument gets closer to the on...

Страница 6: ...disabled The Automatic Shut Off feature will reset itself the next time the instrument is turned on 4 0 Operation and Use of the 718A Air Ionizer Test Kit It is recommended that the user be familiar...

Страница 7: ...er s operating manual or consult with the ionizer manufacturer to determine what this offset balance should be 5 0 Service Calibration 5 1 Service and Repair in the event that you believe the 718 Stat...

Страница 8: ...he stand to ground 5 5 3 Turn on the 718 Static Sensor and set it to the 2 kV range 5 5 4 Ground the plate Zero the display 5 5 5 Remove the ground from the plate and connect it to the high voltage po...

Страница 9: ...ge Automatic Shut off 20 minutes after last switch activity Operating Time for Battery Greater than 50 hours with new battery 21 C Continuous usage Distance Indicator LED targets Aligned targets indic...

Страница 10: ...ual 1 ea Certificate of Conformance 8 0 Contact Information Customer and Technical Service Within the U S Customer service and technical support can be obtained by calling the 3M Electronic Handling P...

Страница 11: ......

Страница 12: ...less expressly agreed upon in writing by an authorized officer of 3M Warranty Limited Remedy Limited Liability This product will be free from defects in material and manufacture for a period of one ye...

Отзывы: