PANNORAMIC® SCAN RX 1.1
TROUBLESHOOTING AND MAINTENANCE
NOVEMBER 18, 2021
158
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
4.
If the magazine is tilted because of twisted tappets
(1)
, first restore the tappets to their
initial position. They are correctly positioned if vertically aligned exactly one above each
other. Turn the knurled wheel
(4)
by hand, until the tappets are in their initial position
again and the magazine flatly rests on both horizontal guide rails.
Notice:
Any manipulation performed on a mechanical component
(for example transporting
system) by the user must be followed by a complete Pannoramic® SCAN RX initialization
routine (software restart).
9.5.5
Transporting Pannoramic® SCAN RX
The following procedure shows how to transport the Pannoramic® SCAN RX in-house.
For transporting the whole system to other location, see
Appendix
.
Figure 135. Transporting handles
Caution!
Do not use the blue guide rail bar
(1)
to raise, shift or transport the Pannoramic® SCAN RX!