TROUBLESHOOTING AND MAINTENANCE
NOVEMBER 18, 2021
155
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
4.
Remove slide loader unit.
5.
Press clamping lever
(4)
down. The slide position lock releases.
6.
Retrieve slide
(3)
from the holder.
Caution!
If the slide is broken, make sure to remove all pieces completely from the device. The edge of
the broken area of the slide is sharp and it can lead to injury and contagion! Broken slide can be
removed only by a well-trained person!
7.
Release clamping lever
(4)
and turn down the slide loader
(2)
back into the original
position and pull it inwardly, until the guiding pin connects with the motor block groove
again.
8.
Press the slide loader down to mechanical stop position, in which the locking lever must
engage
(1)
. Use moderate force to press the loader part opposite the motor in clamp
direction
(2)
, until this part mechanically engages, too.
9.
Close the protective cover again (see chapter
9.5.1 Removing the Protective Covers
10.
Launch Pannoramic® Scanner Software for Research control software.
9.5.3
Installing and removing the filter module (fluorescent option)
Figure 133. Installing and removing the filter module