iSaCS – IHC Stainer, Automated Coverslipper and Scanner
User's Guide
Notes Regarding Operational Safety
Filling up containers
•
DI Water
: Distilled water, final 0.04% Tween20 non-ionic detergent solution (min. 0.5l)
•
Wash Buffer
: Diluted (1x), final 0.04% Tween20 non-ionic detergent solution (min. 0.5l)
If using a different type of container for Wash Buffer, you must ensure that the container
does not contain detergent, and must be completed with Tween20.
Coverslip
•
Coverglass measurements: 22x50 mm, 170μm thickness, clean and non-adhering
•
Mounting medium vial (
DAKO Faramount
, 1000μl), cap and insert must be removed before
insertion
•
Fill 1 ml distilled water into the pipetting tip holder
Reagent chain
•
All reagents should contain final 0.2% Tween20, reagent dead volume approx. 600μl
•
Reagents should not contain any air bubbles at the time of use, all fluid must form a single
volume unit, no droplets are allowed anywhere at the reagent container area.
Slide
•
Use only
SuperFrost Ultra Plus
slides after baking, deparaffinization and after at least 20 min.
of over 90 C HIER.
•
Slide dimensions: 25.2x75.5x1 mm, label area: 19 mm, white label coating
•
Check tissue sample placement on slides: leave 2.5 mm margin on both sides and 5 mm
margin from bottom and top
•
Tissue should adhere firmly to the glass surface on the entire area
Pipette tip
•
Use
Robbins AUTOTIP, 200μl
pipette tips only (Manuf: Molecular BioProducts Inc., Product
Code: 1043-04-5). Other types of pipette tips cannot be used.