![3DHISTECH iSaCS Скачать руководство пользователя страница 39](http://html1.mh-extra.com/html/3dhistech/isacs/isacs_user-manual_442156039.webp)
iSaCS – IHC Stainer, Automated Coverslipper and Scanner
User's Guide
6 Service
6 Service
The following functions are available under the
Service
menu:
Shutdown
– Slide data will be deleted before turning off the device.
Startup
– The four fluid compartments will be calibrated, the system will be flushed out (remaining
materials will be discarded into the
Hazardous Waste
container).
•
Slides left in the Coverslipping, Washing, and Darkfield units are searched for, and you receive
a warning message if the system finds any.
•
If a slide is left in the Washing unit, it will be moved to the bottom position of the incubation
compartment in case it is empty.
Rerun Slide Recognition
– Slide data will be refreshed for all of the slides loaded in.
Barcode Printer
– Barcode data can be printed after editing by clicking
.
September 22, 2016
3DHISTECH Ltd.
39(51)