Pannoramic MIDI 1.15 User’s Guide
5 Troubleshooting and Maintenance
Warning!
Crush hazard! Stop the running digitization process before you proceed
with filter module installation or removal.
2.
Pull the pair of the protective cover halves apart or remove the right part completely (see
section
5.3.1 Removing the protective covers
), but do not switch the main unit off.
3.
Select the desired filter in the corresponding software window (see section
4.2.3
Microscope settings/Base Settings
). The filter will be positioned in such a way that the user
can access it without having to turn the filter wheel manually.
4.
Enter the respective parameters in the input fields.
Warning!
Crush hazard! Rotating the filter turret motorically during installation
or removal may lead to injury.
5.
Hold the filter module by its handling pads
(5)
(one on the right, one on the left side of the
module) and insert its upper part in forward tilted position under the upper spring-loaded
clamps
(1)
.
6.
Press the filter module by its handling pads, until it safely connects to the bottom spring-
loaded clamps
(6)
.
Mounting position of the filters: Excitation filter
(3)
; Emission filter
(4)
.
7.
To remove the filter module, first slightly tilt it out of its upper spring-loaded fixtures, then
pull it out of the bottom spring-loaded components and take it off.
8.
Place the protective covers back again (see
5.3.1 Removing the protective covers
).
9.
Turn the HXP120 illuminator on again (consider the cooling time possibly required).
2012. March 13.
3DHISTECH Ltd.
123
(