Pannoramic MIDI 1.15 User’s Guide
5 Troubleshooting and Maintenance
10.
Screw the lateral cover-plate down to the housing.
11.
Connect power plug.
12.
Set hour meter (time counter) to zero.
Caution!
If Fluorescence scanning option is available, create new compensation
images after replacing the bulb in HXP 120. For more information see
section
4.2.2
Image Compensation / Fluorescent Compensation
.
5.3.4 Removing a slide
If a slide got jammed or has suffered breakage during the process of digitization, stop the process
and, if possible, move the slide holder to a position where the slide can be easily removed by clicking
the
Emergency Manual Slide Removal Position
button (see section
4.3.6 Service tab
).
Warning!
Crush hazard! Before you remove a jammed slide, cancel
the running
digitization process, terminate the software session and turn the power
off.
Any manipulation performed on a mechanical component must be followed by a complete device
initialization routine (software restart).
Figure 73 – Removing a slide
2012. March 13.
3DHISTECH Ltd.