4. Technical specifications  

4.1  Analogue inputs  

Voltage inputs (channels A and B):  

Measuring equipment:  2 differential amplifiers  
Measuring range: 

±200mV, ±2V, ±20V 

Surge voltage protection: Max. ±40V 
Connections: 

4-mm safety sockets  

Current input (channel A):  

Measuring range: 

±200mA, ±2A 

Excess current protection: Max. ±2.5A 
Connections: 

4-mm safety sockets  

Sensor inputs (channels A and B):  

Sensor type: 

Analogue  

Sensor identification:  Automatic  
Sensor connections: 

8-pin mini DIN sockets  

Sampling: Continuous 

 

Sampling rate: 

50k samples/s  

Resolution: 

12 bit  

 

 

4.2  Analogue outputs (channels A and B)  

Reference point (earth):  Common  
Measuring range: 

± 5V  

Connections: 4-mm 

safety 

sockets 

and 8-pin mini DIN 
sockets  

Sampling rate: 

10 kSamples/s  

Resolution: 

12 bit  

 

 

4.3  Digital inputs  

Channels: 

4 (separated into 2 
TTL inputs, one of 
which is a high-speed 
time input, plus 2 in-
puts via optocoupler)  

Sampling rate: 

50k samples/s 
100k samples/s (high-
speed time input)  

Connection: 

8-pin mini DIN socket  

 

 

4.4  Digital outputs  

Channels: 6 

 

Signal: TTL 

 

Connection: 

8-pin mini DIN socket  

 

 

4.5  Additional data  

Computer connection:  USB  
Internal data memory:  128k 
Display: 

64 x 122 matrix for 
measurements and 
units  

Power supply: 

4.5 V DC / 300 mA or 
3 batteries (AA, LR6 or 
AM3); owing to their 
longer battery life, use 
of alkaline batteries is 
recommended.  

Dimensions: 

21 x 8 x 4 cm

3

 approx. 

Weight: 

400g approx. (includ-
ing batteries)  

 
 

5. Description  

5.1 Components  

 

 

1 Display 

 

2  Operating controls  
3  Current input for chan-

nel A  

4  Voltage outputs for 

channels A and B  

5  Voltage inputs for 

channels A and B  

 
 
 
 
 
 
 
 

 

 
6  Digital outputs  
7  Digital inputs  
8 Analogue outputs A 

and B 

9  Analogue inputs A and 

 

Содержание 1000539

Страница 1: ... supply unit 4 5V DC 300mA 1 USB cable 1 Installation CD 3 Introduction 3B NETlog is a multimedia data acquisition and evaluation system for current and voltage measurements and for measurements using sensors The system can be used with or with out being connected to a computer With the accompanying 3B NETlab software meas urements can be conducted with arbitrarily chosen parameters as well as in ...

Страница 2: ...Resolution 12 bit 4 3 Digital inputs Channels 4 separated into 2 TTL inputs one of which is a high speed time input plus 2 in puts via optocoupler Sampling rate 50k samples s 100k samples s high speed time input Connection 8 pin mini DIN socket 4 4 Digital outputs Channels 6 Signal TTL Connection 8 pin mini DIN socket 4 5 Additional data Computer connection USB Internal data memory 128k Display 64...

Страница 3: ...ion of measurement parameters for channels A and B Changing between fields while setting the time and date Return to previous menu 5 3 Measurement parameters Name Significance VdcA DC voltage on channel A VacA AC voltage on channel A Idc DC current on channel A Iac AC current on channel A VdcB DC voltage on channel B VacB AC voltage on channel B Bin Binary representation 5 4 Connection of sensors ...

Страница 4: ... hardware manager under Control Panel Hardware and Sound Hard ware Manager Double click USB Controller in the hard ware manager Double click USB Serial Converter Click Drivers Update Driver A small window will open so that you can search for the driver software Click Find driver software on this computer in the window Click the Browse button and select the path for the driver Follow the instructio...

Страница 5: ...te Switch on the 3B NETlog device by pressing the On Off button Press the Date Time three times the date field will appear on the display Press the Store button the input field for SET DATE will appear on the display Specify the desired date in the input field by pressing the Rate or Store button and use the Channel button to skip be tween the year month and day fields Confirm by pressing the Date...

Страница 6: ...n the 3B NETlog memory can be retrieved and saved as text files with the help of the 3B NETdata software In addition the software also provides the capability to set modes of measurement and measurement parameters Note Only one interface 3B NETlog may be connected to the computer when using the software 3B NETdata 9 1 Installation of the driver Before installing the 3B NETlab software it is necess...

Страница 7: ...ntire memory of the instrument Quit exits the program 9 4 Reading and saving measurement data Reading and saving measured data with the 3B NETlog equipment Connect the 3B NETlog equipment to a computer via USB cable Start 3BNETdata and click on the tab Read Data Press Read and wait till the progress bar reaches 100 Select the measurements you wish to be saved from the list on the left Select colum...

Страница 8: ...ce Press Test again and if neces sary refresh by pressing F5 13 Support If you have any queries and or suggestions please feel free to contact our support team Email support 3bnetlog com Internet http www 3bnetlog com 14 The CE marking 3B NETlog conforms to the requirements of these EU specifications EN 61010 1 Protoype tested EN 61326 1 Tested for noise im munity and interference Conformity with ...

Отзывы: