Model 306 Ozone Calibration Source Manual Rev. G-1
19
7.3
Running a Calibration of an Ozone Monitor
Once the connections are established between the Model 306 outlet and the inlet of
your ozone monitor, the calibration may be performed. The ozone output of the
Model 306 may be changed manually from the LCD
O3
menu or from the serial menu.
Alternatively, the Model 306 can run a preprogrammed step calibration using a user-
prescribed series of up to 10 different ozone output steps, executed either in the LCD
menu (
Stp
menu) or in the serial menu (
run step
command). See Sections 4.4 and
5.2 for information on these options. The series of steps are defined by the user,
using either the LCD menu (
Cfg
/
Prg
) or the serial menu. See Sections 4.2.2 and
Section 5.2 for information on defining the step program.
The preprogrammed step calibration can be scheduled to run automatically once per
day at a user-defined time. See Section 4.2.3 for information about setting this
scheduler feature from the LCD menu (
Cfg
/
Sch
menu, followed by power cycling the
instrument), and Section 5.2 for setting it remotely from the serial menu (
set hr
command, followed by
hibernate
).