Model 306 Ozone Calibration Source Manual Rev. G-1
5
3.
STARTUP OPERATION
Please read all the following information before attempting to install the Ozone
Calibration Source
™. For assistance, please call 2B Technologies at (303)273-0559
or email us at
NOTE:
Save the shipping carton and packing materials that came with the Ozone
Calibration Source
™. If the Ozone Calibration Source™ must be returned to the
factory, pack it in the original carton. Any repairs as a result of damage
incurred during shipping will be charged.
3.1
Shipping Box Contents
Open the shipping box and verify that it contains all of the items on the shipping list. If
anything is missing or obviously damaged, contact 2B Technologies immediately.
3.2
Startup of the Ozone Calibration Source
™
To operate the Ozone Calibration Source
™, connect it to an external power source
and turn the instrument on by flipping the front panel switch. The instrument requires
a 12 V DC source which can be supplied by: 1) the 110-220 V AC power adapter (2.5
amp or higher), 2) a cigarette lighter adapter plugged into a 12 V DC source such as
found in an automobile or many light aircraft, or 3) a 12 V battery. The source must
be very close to 12 V because of the power requirements of the instrument’s pump.
When using a battery, be certain to attach the positive (red) and negative (black) wires
correctly. A circuit breaker and diode are installed on the circuit board in case of an
electrical short or incorrect battery attachment. If activated, the breaker will reset itself
after a few minutes.
Once turned on, the instrument will briefly display the version number of the software
installed on the microprocessor. The instrument will then display
Warming up …
delT = -xx.xx
where
delT
is the deviation of the temperature of the photolysis chamber from the set
point of 315 K (42°C). As the chamber warms up, the temperature deviation will be
continuously displayed. During the warmup period, the photolysis lamp and the pump
are turned off and the internal 3-way valve is set to divert sample air to the output of
the Model 306.
Warmup will take approximately 15-20 minutes. Once
delT
is within 1
°
C of the set
point, the display will read:
Temperature Set
Press Select