Model 202 Ozone Monitor Manual Rev. J-4
19
4.
ACCESSING THE SERIAL MENU
Measurements and logging tasks can be accessed via the serial port or the USB using
the
2B Technologies Display and Graphing software
(see Appendix A for information
on downloading and working with this free software) or a terminal emulator such as
Tera Term
or HyperTerminal running on an attached computer. Commands can be
sent using the terminal emulator set with the properties listed in the Section 3.5 of this
manual entitled “Collecting Data over the Serial Port in Real Time.” Listed below are
the lower-case letters that are commands for performing certain operations while the
instrument continues to measure:
l
Start logging and write over existing logged data
t
Transmit logged data
e
End logging
h
Output serial data line header
m
Serial menu
If the letter
m
is sent as a command,
menu>
will be displayed in the terminal emulator
window. When the serial menu is accessed, the instrument is no longer making
measurements; it is waiting for the next command to be entered. The following is the
list of menu items accessible from this point:
l
Start logging and write over existing logged data
t
Transmit logged data
e
End logging
h
Output the serial header
b
Ends current file being written to external data logger (avoids losing data that
are still in the buffer)
a
Display list of possible averaging times and the number that must be
entered to change to the desired averaging time
z
Display current zero calibration setting and wait for new setting followed by a
carriage return
s
Display current slope calibration setting and wait for new setting followed by a
carriage return
f
Display current flow calibration factor and wait for new setting followed by a
carriage return
c
Clock menu, display current date and time and wait for
d
or
t
to be entered
d
Asks to enter date in DDMMYY format
t
Asks to enter time in HHMMSS format
p
Perform lamp test with pump on
u
Perform lamp test with pump off
d
Turn the LCD backlight on
g
Turn the LCD backlight off
?
Output the help menu
x
Exit menu and return to measuring
If the instrument is equipped with the cold-weather option, two additional commands
are available:
k
Activates the Lamp Heater connected to J23
j
Deactivates the Lamp Heater connected to J23