1-CUBE SU 6.2. Скачать руководство пользователя страница 8

 

8

Adjustment of the required temperature or time: 

The required temperature or time in the methods (menu) THERMOSTAT, CALIBRATION and PROFILE are adjusted by 
pressing keys plus +  (to increase the value) or minus – (to decrease the value). First the required  value is changed of 
tenths of degree or  1 seconds, after ten changes (ten presses of key) the required  value is changed of degrees or 10 
seconds, after ten following changes (ten presses of key) the required  value is changed of 10 degrees or 100 seconds 
etc. Once you have adjusted the required data push the key 

OFF 

to save it into the memory, after this step the text 

THERMOSTAT appears again on the display.

 

Method Thermostat:

 

This  algorithm  allows  to  adjust  any  required  temperature  and  the  Water  Bath  maintains  the  temperature  for  the 
required time.  The device has saved the last value  adjusted temperature that was regulated. If you do not want to 
change  the  value  of  the  temperature  start  the  algorithm  itself  by  pressing  the  key 

ON

.  The  text:  REQ.CONTROL: 

appears on the first line of the display, there is the time in minutes 

„min=“

and the real temperature 

„T=“

 in  

0

C

 on the 

second line.  

If the set temperature doesn’t suit you and you need to change it you have to adjust the required temperature before 
you press the key 

ON.  

The required temperature is set with the help of keys plus +  (to increase the value) or minus – 

(to decrease the value) – see the paragraph above: 

Adjustment of the required temperature or time: in the chapter 

Control  algorithms.

  After  setting  the  required  temperature  press  the  key 

OFF 

and

 

in  that  way  the  required 

temperature  is  saved  and  the  text  THERMOSTAT  appears  again  on  the  display.  Now  you  can  start  the temperature 
control  running  by  pressing  the  key 

ON

.  The  algorithm  is  stopped  by  pressing  the  key 

OFF

.    If  there  is  not  enough 

water in the Bath the text 

„add  water“

 appears automatically on the display and it will disappear once the water is 

added. 

Method PROFILE:

 

This  algorithm  allows  to  adjust  any  required  time  behaviour  of  temperatures.

 

It  is  possible  to    adjust  as  many  as  7 

points

 

defined with time and temperature coordinates. After choice of the item PROFILE in the main menu with the 

help of the key 

ROTATION

 one can start the algorithm itself by pressing the key 

ON

.  

After 1 sec the text: START ON: 

                                 OTHER PARAMETERS:  appears on the display. 

If there has been already set the temperature curve before, one can use it again (because it is saved  in the internal 
memory) by holding the key ON at the time when the text: START ON: 

                                                                          OTHER PARAMETERS: appears on the display. 

If there hasn’t been  set the temperature curve yet or it is necesary to set it in a different way, one waits about 1 sec 
till the text SET UP THE CURVE appears on the display. After that the text: time of point 1= appears on the first line and 
one  sets  the  time  in  minutes  on  the  second  line.  Time  is  set  with  the  help  of  keys  plus +  (to increase the value) or 
minus  –  (to  decrease  the  value)  –  see  paragraph:  Adjustment  of  the  required  temperature  or  time:  in  the  chapter 
Control algorithms. After setting the time of point 1 press the key 

OFF 

and

 

in that way time of point 1 is saved in the 

internal memory and after that the text: time of point 2= appears on the first line and  one sets in the same way the 
time of point 2, the time of point 3, the time of point 4, the time of point 5, the time of point 6, the time of point 7.  
After setting the time of point 7 press the key 

OFF 

and 

the

 text: temp. of point 1=  appears on the first line and one 

sets the temperature in 

c

 C.  on the second line. Temperature is set with the help of keys plus +  (to increase the value) 

or minus – (to decrease the value) – see paragraph: Adjustment of the required temperature or time: in the chapter 
Control algorithms. After setting the temperature of point 1 press the key 

OFF 

and

 

in that way temperature of point 1 

is saved in the internal memory and after that the text: temp. of point 2= appears on the first line and  one sets in the 

Содержание SU 6.2.

Страница 1: ...ube com WATER THERMOSTAT TYPE SU 6 2 USER S GUIDE Manufacturer and service 1 CUBEs r o Hamry 3567 580 01 Havl Brod tel 569 433 620 fax 569 422 144 CONTENT 1 Range of the supply 2 2 Installation instructions 2 3 Safety recommendations 3 4 Technical data 4 5 User s guide 5 6 Monitoring RS232 13 7 Maintenance 13 8 Important warnings 13 ...

Страница 2: ...sition for proper operation of the apparatus Plug the Bath with the help of 2 standardized single phase plugs into el socket Connect each bath cold and hot with its own cord Before plugging in the technician has to check the condition of the el circuit that will be used for the Water Bath operation If the result of the el circuit inspection is positive the Water Bath can be plugged in Plug wiring ...

Страница 3: ... condition of the el circuit that will be used for the Water Bath operation If the result of the el circuit inspection is positive the Water Bath can be plugged in There is the Main Switch located on front panel The Main Switch has two positions ON 1 OFF 0 In case of danger switch off the Main Switch and disconnect feed el cord out of the socket Switch off the Main Switch disconnect feed el cord o...

Страница 4: ... 16A circuit breaker Adjustable data type of method THERMOSTAT PROFILE CALIBRATION temperature range from 25 0 C to 90 0 C cold bath from 10 0 C to 30 0 C with the help of membrane keyboard and display time since the begining of test Accuracy accuracy of temperature measurement 0 01 0 C accuracy of temperature regulation 0 1 0 C Data represented on LCD display type of method THERMOSTAT PROFILE ins...

Страница 5: ...ectrically conductive liquid If wou use non electrically conductive liquid then the device will notify all the time that there is not enough water inside the bath even if it isn t true If you use distilled water in the laboratory just add a pinch of salt NaCl and you will get electrically conductive liquid Switch on the cooling on off switch except for the main switch in the cooling bath There is ...

Страница 6: ...r will be controlled switch on switch off with Main switch Once the main switch is switched on the stirrer is switched on automatically too One chip microprocessor is used to control the device The algorithm is saved in the program Program is controlled with the help of 7 keys of the membrane keyboard which is located on the front panel of the equipment The report about the operating condition can...

Страница 7: ...he text THERMOSTAT appears on the display Once you finish working switch off the device with the help of the Main switch Control algorithms The following control algorithms PROFILE THERMOSTAT CALIBRATION can be run with the Water Bath During the process the following information are automatically displayed on the display 1 line The required temperature is displayed 2 line Information about time gi...

Страница 8: ... display and it will disappear once the water is added Method PROFILE This algorithm allows to adjust any required time behaviour of temperatures It is possible to adjust as many as 7 points defined with time and temperature coordinates After choice of the item PROFILE in the main menu with the help of the key ROTATION one can start the algorithm itself by pressing the key ON After 1 sec the text ...

Страница 9: ... point 5 80 time of point 6 105 time of point 7 125 temperature of point 1 30 temperature of point 2 30 temperature of point 3 34 temperature of point 4 34 temperature of point 5 40 temperature of point 6 40 temperature of point 7 25 Method Calibration It is necessary to check the measurement accuracy with the help of etalon once in a six months If the temperature of the bath does not correspond t...

Страница 10: ...e first line now and the value on the second line Do not set the point 2 at this moment therefore press the key OFF 6 There is text point 2 temp on the first line now and the value on the second line Do not set the point 2 at this moment therefore press the key OFF 7 There is text HYSTER1_0 on the first line now and the value on the second line Do not set HYSTER1_0 at this moment therefore press t...

Страница 11: ...n the first line now and the value on the second line Overwrite the value with the help of keys increase or decrease to put the value that you have measured before with the etalon in the water of the water bath Press the key OFF to set the required value 7 There is text HYSTER1_0 on the first line now and the value on the second line Do not set HYSTER1_0 at this moment therefore press the key OFF ...

Страница 12: ...R1_0 as required Press the key OFF to set the required value Hysteresis HYSTER1_0 is preadjusted by the manufacturer as 000 30 8 There is text HYSTER1_100 on the first line now and the value on the second line Overwrite the value with the help of keys increase or decrease to adjust HYSTER1_100 as required Press the key OFF to set the required value Hysteresis HYSTER1_0 is preadjusted by the manufa...

Страница 13: ... help of a hose and clean the vessel 8 IMPORTANT WARNINGS The Cooling Bath must be always located on the nibs and mustn t be tilted or turned upside down in any case If the device is transported at temperature below freezing point it is necessary to leave the instrument for about 2 hours at room temperature before you start to work with it The fuses are located on the back panel of the Water Bath ...

Отзывы: