5
5.
U
SER
’
S GUIDE
Picture of the cooling bath - SU6.2:
Control panel
Mixer engine Removable top
Cooling thermostat
Cooling on-off switch
Cooling exchanger
Air inlet
First switch on the Main Switch located on the front panel of the equipment.
NOTICE:
It is forbidden to use the device without filling it with water. It could cause device destruction and staff injury.
The device warns you if there is not enough water inside the bath. It is necessary to use electrically conductive
liquid.
If wou use non electrically conductive
liquid then the device will notify all the time that there is not enough water
inside the bath even if it isn’t true. If you use distilled water in the laboratory just add a pinch of salt (NaCl) and you
will get electrically conductive
liquid.
Switch on the cooling on-off switch except for the main switch in the cooling bath. There is a cooling thermostat just
next to cooling on-off switch and we do reccomend to leave its value as preadjusted „value 5“
Warning! Do not exceed the temp30 degrees Celsius in the cooling bath otherwise the cooling system can
be irreversibly damaged!!!!!