background image

Barcode Scanner CLV62x

Online Help

O N L I N E H E L P

Standard Line

CLV62x
Bar Code Scanner

Summary of Contents for CLV62x

Page 1: ...Barcode Scanner CLV62x Online Help O N L I N E H E L P Standard Line CLV62x Bar Code Scanner ...

Page 2: ...11 79276 Reute Germany Trademark Windows 2000TM XPTM VistaTM and Internet ExplorerTM are registered trademarks or trade marks of the Microsoft Corporation in the USA and other countries AcrobatTM ReaderTM is a trademark of Adobe Systems Incorporated Software Tool Function Version Device description CLV62x Device specific software module for configuration software SOPAS ET V 2 0 SOPAS ET Configurat...

Page 3: ...3 1 Codabar 13 2 2 3 2 Code 39 14 2 2 3 3 UPC EAN 15 2 2 3 4 2 5 Interleaved 16 2 2 3 5 Code 93 17 2 2 3 6 Code 128 family 18 2 2 3 7 Pharmacode 19 2 2 4 Data processing 20 2 2 4 1 Output control 21 2 2 4 2 Evaluation conditions 23 2 2 4 3 Match code teach in 27 2 2 4 4 Filter Sorter for output formatting 28 2 2 4 5 Output format 29 2 2 5 Network Interface IOs 31 2 2 5 1 Serial 32 2 2 5 2 Ethernet...

Page 4: ...Online Help Barcode Scanner CLV62x 4 SICK AG Division Auto Ident Germany All rights reserved 8011967 S345 2008 04 23 Contents ...

Page 5: ... configuration basic configuration of the barcode scanner is set for use as a stand alone device Information on mounting installation maintenance and troubleshooting are listed in the operating instructions of the barcode scanner document no 8011965 Important Further information on the design of the barcode scanner as well as the barcode technology is available from SICK AG Auto Ident division On ...

Page 6: ...Chapter 1 Online Help Barcode Scanner CLV62x 6 SICK AG Division Auto Ident Germany All rights reserved 8011967 S345 2008 04 23 Notes on this document ...

Page 7: ...ercent evaluation The percentage related to the actual number of scans on the code is displayed Codabar Activate deactivate decoding of code type CODABAR see also chapter 2 2 3 Code configuration page 12 Code 39 Activate deactivate decoding of code type CODE 39 see also chapter 2 2 3 Code configuration page 12 UPC EAN Activate deactivate decoding of code type UPC EAN see also chapter 2 2 3 Code co...

Page 8: ...of the barcode Also see the frequency diagram in the operating instructions Quietzone ratio Select ratio between quiet zone white area before and after the barcode and the code s spaces Barcode quality Select the average barcode quality Only deviates from the STANDARD setting if the barcode has imperfections or is poorly printed Codelabel distance Only for PHARMACODE code types Enter the distance ...

Page 9: ...igger From Enter the CAN address of the device from which the start trigger is derived For this device CAN must be selected for the TRIGGER DISTRIBU TION Stop delay Enter delay between start trigger and the closing of reading gate Select possible delay unit Select source operation of stop trigger OR Select further source operation of stop trigger optional This source is linked to the first source ...

Page 10: ...group Dyn reading config assignment table The number and order of configurations for the dynamic reading configuration is deter mined via the parameters of the Dyn reading config assignment table group Parameter Function Distribute on Select trigger receiver If the trigger is distributed via CAN CAN must be indicated as the source of the trigger at the trigger receiver Parameter Function Laser tim...

Page 11: ...number of impulses multiplied by the resolu tion of the external incremental encoder per time Alternatively a fixed speed can be selected Increment The increment source and the resolution speed are configurated via the parameters of the Increment group Parameter Function Increment source Select increment source Fixed speed Enter speed of the conveyor distance If applicable select unit System incre...

Page 12: ...9 Activate deactivate decoding of code type CODE 39 Configuration see chapter 2 2 3 2 Code 39 page 14 UPC EAN Activate deactivate decoding of code type UPC EAN Configuration see chapter 2 2 3 3 UPC EAN page 15 2 5 Interleaved Activate deactivate decoding of code type 2 5 INTERLEAVED Configuration see chapter 2 2 3 4 2 5 Interleaved page 16 Code 93 Activate deactivate decoding of code type CODE 93 ...

Page 13: ... of the Length group Parameter Function Multireads Enter the minimum number of required successful readings that must be achieved to accept a barcode as valid Start Stop identically Set whether only barcodes with identical start and stop charac ters are to be read Transmit start stop Activate deactivate output of the start and stop character together with code content Check digit test Select the t...

Page 14: ... the General group Length A length test of the barcode can be configured via the parameters of the Length group Parameter Function Multireads Enter the minimum number of required successful readings that must be achieved to accept a barcode as valid Transmit start stop Activate deactivate output of the start and stop character together with code content Full ASCII Activate deactivate output of the...

Page 15: ...ame bar sequences as the UPC A code This type of code is decoded as a 12 digit UPC A code in the evaluation routing of the EAN 13 and UPC A The scanner only decodes this 12 digit codes as UPC A if UPC A or UPC A and EAN 13 are activated The code is only interpreted as EAN 13 with a leading zero if EAN 13 is activated General Reading conditions for code types UPC and EAN are configured via the para...

Page 16: ...ccept a barcode as valid Check digit test 1 Select the type of check digit test for the first code length Check digit test 2 Select the type of check digit test for the second code length Check digit test 3 Select the type of check digit test for the third code length Check digit test 4 Select the type of check digit test for the fourth code length Check digit test 5 Select the type of check digit...

Page 17: ... 93 General Reading conditions for code type CODE 93 are configured via the parameters of the General group Length A length test of the barcode can be configured via the parameters of the Length group Parameter Function Multireads Enter the minimum number of required successful readings that must be achieved to accept a barcode as valid Parameter Function Code length mode Select the type of code l...

Page 18: ...ODE 128 is multiple valued valency 4 Fig 2 8 Example of Code 128 General The reading conditions for the code family CODE 128 are configured via the parameters of the General group Length A length test of the barcode can be configured via the parameters of the Length group Parameter Function Code 128 Activate deactivate scanning of code type CODE 128 EAN 128 Activate deactivate scanning of code typ...

Page 19: ...E Length Use the parameters of the group Length to configure testing of the barcode length Parameter Function FC1 value at first position Enter FC1 values that have to be positioned at the beginning of codes that are to be read FC1 value within the code Enter FC1 values that have to be positioned in the middle of codes that are to be read Hint If the code type PHARMACODE shall be read the READING ...

Page 20: ...act of the Timeout parameters on the reading gate Code summarisation The parameters of the Code summarisation group are used to configure whether several similar codes are output as one code or as several codes Parameter Function Time out Enter maximum time after which a read code will be output after the end of the reading gate 50 ms Start Sensor 1 Reading gate Parameter Function Separate codes b...

Page 21: ...me on the output time of the code contents Parameter Function Control Select type of control Output time Select output time of the read code contents Delay Enter delay between end of reading gate and output of the code contents If applicable select unit Output condition Select required condition for the output of the code contents This condition can be configurated see chapter 2 2 4 2 Evalua tion ...

Page 22: ... 2008 04 23 CLV62x Fig 2 12 Impact of the parameter value As soon as possible Output time on the output time of the code contents Fig 2 13 Impact of the parameter values End of label Output time on the output time of the code contents Output time Reading gate Code Output condition 100 ms Reading gate Code Output time ...

Page 23: ... codes Activate deactivate test of minimum number of valid codes per reading gate Minimum Enter minimum number of valid codes which must be scanned to be transmitted Check maximum number of valid codes Activate deactivate test of maximum number of valid codes per reading gate Maximum Enter maximum number of valid codes which must be scanned to be transmitted Parameter Function Conditions Edit cond...

Page 24: ...t wildcards to describe the desired code content Regular expression Select regular expression to describe the desired code content Test Tests whether the desired code matches the condition Min Enter minimum code length Max Enter maximum code length Don t care Activate deactivate the code length check Code type Select desired code type Don t care Activate deactivate the code type check Code validit...

Page 25: ...is true then the result is true other wise the result is false NOT If the following part condition is true the result is false Add bracket around part conditions Add bracket around part conditions Remove row Remove part condition or link between two part conditions Rows can also be shifted by drag drag Parameter Function Add condition Select a previously defined condition for linking with other co...

Page 26: ... type Code 128 that contain the character sequence 45 Condition is met output of MATCHCODE and the code content in the output string Condition is not met output of NOMATCH in the output string No valid code could be read output of NoRead in the output string In order to achieve such a result the following CONDITION and an appropriate OUTPUT FORMAT must be created see chapter 2 2 4 5 Output format ...

Page 27: ...luation condition that will be teached in Invert condition Activate deactivate whether evaluation condition will be logically inverted Teach in triggered from Select trigger source for teach in of evaluation condition Teach in triggered by buttons Activate deactivate button on bar code scanner as trigger source Teach in code content Activate deactivate teach in of code content Teach in code ID typ...

Page 28: ...ect trigger is necessary as well chapter 2 2 1 1 Object trigger control page 9 Don t change code config The settings of the Code configuration group will not be changed Limit code config to last teached in code With activated function only the teached in codes are set to active see chapter 2 2 3 Code configuration page 12 Expand code config by last teached in code The settings of the Code configur...

Page 29: ...by definable data interfaces For this two different output formats telegrams can be defined The format can also depend on condi tions The first format of the reading results is defined via the parameters of the Output format 1 group Parameter Function Filter Filter code contents according to various criteria Define filter criteria Sorter Sort code contents according to various criteria Define sort...

Page 30: ...tbeat format group Parameter Function Output format 2 Enter output format of the reading results Open input field Close input field Select condition Right mouse click Insert new condition data field or special char acter Click on the bottom row of the data field Select attribute charac teristics Maximized Activate deactivate view with open input fields Hint The created output format can be marked ...

Page 31: ...nnection of the scanner in the SICK CAN sensor network is defined via the parameters of the Master Slave group Master Slave means that a group of scanners read the barcodes of the same object The common reading result is transmitted to the host Viewed from the host side the master slave group works like a single scanner Parameter Function Network assistant Call up dialog in which the most importan...

Page 32: ...al host interface is configured with the parameters of the Serial host interface group The serial host interface provides the reading result for further processing by the host com puter Parameter Function Assign to Select connection for monitoring Monitored devices Enter the device numbers of the scanners which are to be moni tored by this scanner Maximum starting time of moni tored devices Enter ...

Page 33: ...t Heartbeat interval Enter time between two heartbeats Restart interval on sending Restart interval after the output of a reading result Gateway Activate PROFIBUS DP interface or DeviceNet respectively Parameter Function Use device ID as profibus address Activate deactivate the scanner s device ID as profibus address Device ID The device ID of the scanner is displayed Operating mode Select the pro...

Page 34: ...ormat Select the format for the output of the reading result via the Eth ernet host interface Multiplexer output Activate deactivate data output via multiplexer Server Client Select whether the scanner works as a server or a client IP port Enter the IP port for the exchange of network data Server address Enter server address Activate heartbeats Activate the output of a regular heartbeat Heartbeat ...

Page 35: ... 1 group Parameter Function Mode Select the type of CAN network Use device ID as knot ID Activate deactivate the use of the scanner device ID as the ID of the knot in the CAN network For distribution of the device ID see chapter 2 1 Quickstart page 7 Device ID Scanner device ID in the CAN network is indicated Baudrate Select network speed Output format Select output format via which the reading re...

Page 36: ...canner with a cable connector standard version and for the Ethernet version via the CDB620 connection module in combination with the parameter memory module CMC600 Parameter Funktion Control Select the type of control Sensitivity Select sensitivity of sensor Logic Select logic of the external input 2 Debouncing Enter detection time for signal at the external input 2 If applicable select unit Param...

Page 37: ...CDB620 connection module in connection with the CMC600 parameter memory module or of the connection module Beeper The beeper is configured with the parameters of the Beeper group Parameter Function Output 2 Select condition that is to trigger a signal at output 2 Logic Select logic level of the selected output 2 Control Select type of control Duration Enter signal duration If applicable select uni...

Page 38: ...ormat Select the format for the output of the reading result via the serial fieldbus interface Usage output Bit0 Select the function of output bit 0 Usage output Bit1 Select the function of output bit 1 Usage input Bit0 Select the function of input bit 0 Usage input Bit1 Select the function of input bit 1 Parameter Function Manufacturer The manufacturer of the scanner is indicated Order number The...

Page 39: ...e Time of latest occurrence is indicated Number Number of message is indicated Description Text of message is indicated Info Further information concerning the message is indicated Status Status of message is indicated Counter The number of message occurrences is indicated Symbol Function Select signal that has to be monitored Open a previously recorded monitoring diagram Save the current monitori...

Page 40: ...LQIR VLFN HV France 3KRQH 0DLO LQIR VLFN IU Great Britain 3KRQH 0DLO LQIR VLFN FR XN India 3KRQH 0DLO LQIR VLFN LQGLD FRP Italia 3KRQH 0DLO LQIR VLFN LW Japan 3KRQH 0DLO VXSSRUW VLFN MS Nederlands 3KRQH 0DLO LQIR VLFN QO Norge 3KRQH 0DLO DXVWHIMRUG VLFN QR Österreich 3KRQH 0DLO RIILFH VLFN DW Polska 3KRQH 0DLO LQIR VLFN SO Republic of Korea 3KRQH 0DLO NDQJ VLFNNRUHD QHW Republika Slowenija 3KRQH 0...

Reviews: