CLEANING INSTRUCTIONS

The following instructions are specific to our 

REUSABLE 

laryngoscope blades and handles. These 

instructions are intended for use only by authorized personnel with the required knowledge and training. 

PRODUCTS  MARKED  “        /  DISPOSABLE  /  SINGLE  PATIENT  USE  /  DO  NOT  AUTOCLAVE” 
SHOULD NOT BE REPROCESSED. 

WARNINGS

1.  Follow instructions and warnings as issued by the manufacturer of any decontaminants, disinfectants 

and cleaning agents used. Wherever possible avoid use of mineral acids and harsh, abrasive agents. 

2.  Do not use chemical sterilants with caustic ingredients such as - surgical scrub solution, povidone-

iodine  solution,  bleach,  peroxide  solution,  virox  3  and  sporox.  Do  not  use  garment  or  surface 
disinfectants. 

3.  Do not autoclave with sub standard stainless steel surgical instrument as this may cause a reaction 

and lead to rust or discoloration.

4.  Ultrasonic and plasma cleaning are not permitted, as this will cause damage to the fiber optic light 

guide and lamp.

5.  Do not flash autoclave or dry heat blade or handle, this will invalidate the warranty .
6.  Our  laryngoscopes  are  suitable  for  sterilization  by  Ethylene  Oxide  (ETO),  Sterris  and  Sterrad 

sterilization procedures.

When  reprocessing  medical  devices,  handle  with  care,  wearing  protective  clothing  and  face 
visors or goggles. 

BATTERIES AND LAMPS MUST BE REMOVED FROM THE HANDLE / BLADE 

PRIOR TO CLEANING. 

LIMITATIONS ON REPROCESSING: 

Repeated autoclaving, especially for extended periods will affect 

the fiber optic light guide ends which may lead to a reduction in light output. 

FROM POINT OF USE:

 If possible, soiled blades should be placed in a holding solution (combined 

disinfectant / enzyme solution) immediately after use and prior to cleaning.

PREPARATION FOR DECONTAMINATION:

 Reprocess all blades as soon as it is reasonably practical 

following use. We recommend disassembling the fiber optic light guide from Modular blades for separate 
cleaning.

CLEANING-AUTOMATED:

  Use  only  validated  washer-disinfector  machines  and  cleaning  agents, 

following the manufacturer's instruction for use, warnings and recommended cycles. Load blade and 
handle carefully, ensure the fiber optic light guides are not liable to damage during loading procedure. 
Bulbs should not be cleaned in these machines.

CLEANING-MANUAL: 

Use a sink dedicated for instrument cleaning (not used for hand washing), rinse 

excess soil from blades & handles (water temp <35°C) Keeping the blades submerged, with a brush, 
apply cleaning solution to all surfaces. Pay particular attention to underside of the blade and heel. Always 
brush away from the body. Do not use a wire brush. Take great care not to brush against the Fibre ends in 
the blades to avoid scratching or damage. Rinse blade thoroughly with clean water, so that the water 
reaches all parts of blade, then carefully hand dry or use a drying cabinet. 

DISINFECTION: 

To disinfect Cidex® OPA or 2.4% Glutaraldehyde solution may be used. Always follow 

the manufacturers instructions. After aplication, dry with a clean cloth or filtered air.

LAMP  CLEANING:

  Lamps  should  only  be  cleaned  manually.  Soak  surgical  grade  cotton  swab  in 

decontaminant and apply to wipe off lamp gently. Repeat this procedure twice on every cleaning cycle. 
Carefully hand dry lamp before placing back in position.

 

Replace batteries with the size and type specified in your product's User Manual. Remove all batteries 

from the product at the same time. 

 Charge rechargeable batteries with the instructions and correct charging equipment supplied with 

your product. 

 

Store  batteries  in  a  cool,  dry  place  at  normal  room  temperature,  Remove  batteries  from 
handles that will be stored unused for extended period of time.

 Dispose off batteries according to the local laws and regulations of your region. Some batteries may 

be recycled, and may be accepted for disposal at your local recycling centre. If you are not able to 
identify the applicable rules in your area, please check the instructions of the battery manufacturer. 

 Contact supplier

 

if you have questions about proper handling of batteries. Batteries may present a risk 

of fire, explosion, or chemical burn if mistreated. 

 Do not disassemble, puncture, modify, drop, throw, or cause other unnecessary shocks to batteries. 

 Do not dispose off batteries in a fire or trash incinerator, or leave batteries in hot places such as an 

automobile under direct sunlight. 

 Do not store batteries near an oven, stove, or other heat source. 

 Do not connect batteries directly to an electrical source, such as a building outlet or automobile power-

point. 

 Do not place batteries into a microwave oven, or into any other high-pressure container. 

 Do not immerse batteries in water or otherwise get them wet. 

 Do not short circuit batteries; do not carry loose batteries in a pocket or purse with other metal objects, 

which may inadvertently cause a battery to short circuit. 

 Do not use batteries, or charge rechargeable batteries, that appear to be leaking, discoloured & rusty, 

deformed; emit an odour; or are otherwise abnormal. 

 Do not touch leaking batteries directly; wear protective material to remove the batteries and dispose 

off them properly immediately. 

 Do not mix old/used with new batteries, or install used batteries in other equipment. 

 Do not charge non-rechargeable batteries, or use unspecified charging instructions or equipment for 

rechargeable batteries. 

 Do not continue to charge batteries beyond the specified time. 

 Do not give batteries to young children, or store batteries where children may access them. 

 Do not put batteries in mouth. If swallowed please contact physician or local poison control centre 

immediately.

Warning
Power outputs from some rechargeable batteries can fall rapidly during use, resulting in rapid 
failure of illumination.

Recycling  is  the  best  option  from  both  an  economic  and  environmental  standpoint  for  most 
batteries. Check with companies to make sure that they will take back old batteries when new 
ones are bought. If batteries cannot be recycled (with the exception of alkaline batteries), make 
sure they are tagged for collection by designated personal.

IFU/multiple/2.0/rev1.0/04/2017

1

4

5

2

3

‘AA

‘AA

‘AA

‘AA

1

3

2

‘C’

‘C’

BATTERY SIZE & INSTALLATION

1

3

2

‘AA

‘AA

STERILIZATION: 

To  sterilize,  use  a  validated  vacuum 

autoclave  operating  at  134°-137°C  @  2.25  Bar  for  a 
minimum  holding  time  of  3  minutes.  Always  follow  the 
instructions  of  machine  manufacturer.  When  sterilizing 
multiple  instruments  in  one  autoclave  cycle,  ensure  that 
sterilizer  manufacturers  stated  maximum  load  is  not 

exceeded. Ensure blades are dry before sterilization, if they 
cannot be dried prior to sterilization, then use distilled / de-
ionized water in the final-rinse stage of cleaning.

REASSEMBLY:

  If  the  blade  or  handle  has  been 

disassembled (as may be the case with Modular fiber optic 
blades), reassemble and ensure the device works properly. 
Check the fiber optic light guide fits correctly. Do not apply 

excessive force. Lamps should be placed back in their original housing securely.

INSPECTION AND FUNCTION TESTING: 

Visually inspect and check all lamps and fibre optic light 

guides for damage and wear. To check functionality, connect the laryngoscope blade to the handle and 
pull upward to the ON position. If the unit fails to light or flickers, check the lamp and component fittings 
battery.

PACKAGING:  

All blades and handle should be packed following local protocols.

STORAGE:

 Ensure blades and handles are dry before storage and stored one in dry, clean condition at 

an ambient room temperature. 

ADDITIONAL INFORMATION:

 Other forms of cleaning (alkaline and neutral) and sterilization (Cidex 

OPA, Ethylene oxide Up to 65° C for anesthetic devices) are permitted. However, always follow the 
instructions for use as issued by the manufacturer and always consult with them if in any doubt over the 
suitability of any process used.  

It  is  the  responsibility  of  the  reprocessor  to  ensure  that  the  reprocessing  has  actually  been 
performed using appropriate equipment, materials and personnel and the desired results have 
been achieved. This requires validation and routine monitoring of the process. Likewise any 
deviation  by  reprocessor  from  the  instruction  provided,  must  be  properly  evaluated  for 
effectiveness and potential adverse consequences.

The above listed sterilization guidelines, provided by manufacturer, are intended as suggestions 
only. They are based on procedures compatible with specific materials. Sterilization must be 
performed  to  approved  hospital  protocol.  Manufacturer  cannot  guarantee  that  any  of  the 
recommended methods will guarantee sterility. This must be validated by the hospital and / or 
sterilization equipment manufacturer.

BATTERY INFORMATION :

Remove the screw cap to insert appropriate batteries into handle shaft. For medium handles use two 'C' 
size batteries, for small and stubby handles use two 'AA' size batteries. When inserting batteries the ‘+ve’ 
terminal must be inserted into the tube first on both of the batteries so that the ‘–ve’ terminal is facing the 
open  end.  For  stubby  handles  -  batteries  should  be  inserted  following  the  negative  ‘–ve’  and  ‘+ve’ 
terminals as shown on plastic battery housing provided in all stubby handles. Replace the base cap and 
ensure it is securely tightened. 

Alkaline Batteries:

 are commonly referred to as disposable or non– rechargeable batteries. 

Rechargeable Batteries:

 typically include Nickel – Cadmium (Ni-Cad) or Lithium Ion Batteries. 

DO NOT AUTOCLAVE BATTERIES UNDER ANY CIRCUMSTANCES.

WE RECOMMEND USING ONLY “DURACELL” BATTERIES IN OUR LARYNGOSCOPE HANDLES.

LAMP REMOVAL IN LED HANDLES

Scope Medical Devices Pvt. Ltd.

Reg. Office : 1888 B/10 Baldev Nagar, Ambala City - 134007 INDIA
Factory : 11/10, Jaraut Road, Mandour, Ambala City - 134003 INDIA
Telephone : +91-7082 402 685, 7082 412 685
E-mail : [email protected]   |    Web : www.scopemedical.in   
Buy Online : www.laryngostore.com

2

3

4

5

Reviews: