TH60N(E) Multi-function Infrared Ear Thermometer 

Functions 

Room Temperature

 

Suitable  ambient  temperature  is  important  for  the  baby  and  patient,  the  thermometer  always  help  you  recognize  the 
room temperature. 

 When in power off, room temperature will always be shown on the screen with 

  icon. 

 Please see the “Use of the thermometer” section to learn how to get the correct room temperature.   

 How to turn off Room Temperature: When in power on, press and hold the “ON/MEM” button for 5 sec. until   

OFF”

 will be shown on the screen to release the button. 

 

Ear Temperature 

The  thermometer  has  been  designed  for  practical  use.  It’s  not  meant  to  replace  a  visit  to  the  doctor.  Please  also 
remember  to  compare  the  measurement  result  to  your  regular  body  temperature. 

Please  consult  with  doctor  if  you  have 

health concerns.

 

 Please see the “Use of the thermometer” section to learn how to measure the body temperature. 

 

Surface Mode 

The Surface mode shows the actual, unadjusted surface temperatures, which is different from the body temperature. It 
can help you to monitor if the object temperature is suitable for the baby or patient, for example the baby’s milk. 

 Please see the “Use of the thermometer” section to learn how to measure the object temperature. 

 

Fever Alarm

 

If  thermometer  detects  a  body  temperature

 

37.5°C  under  ear  mode,  there  will  be  a  long  beep  sound  followed  by 

three short beep sound to warn the user for potential fever. 

 

Last Reading 

When you get a new temperature reading in ear mode, the last reading will be shown on the screen (at top right corner) 
with the 

  icon. 

 

Memory Locations

 

There are total 25 sets memories for ear measurements.   

 When in power on, Press the “ON/MEM” button to see the temperature stored with 

  icon.   

 

°C / °F Switch

 

  In  “power  off”  mode  (The  LCD  only  shows  the  room  temperature).  Press  and  hold  the  “SCAN”  button,  then 

immediately press and release the “ON/MEM” button. After 3 seconds, icon “°C” will be switched to icon “°F”. (You may 
use the same way to switch back to “°C”) 

 

 

 

 

Use of the thermometer

 

1. Always make sure the probe is clean, and without damage.   
2. Power on:    Press “ON/MEM” button.(see figure 1)   

3. Mode selection:   

3.1

 

Ear mode

: Press the “ON/MEM” button, the default mode of thermometer is ear mode. The thermometer is ready   

for use after you see the    icon on the screen and hear two beep sounds (see figure 1). In   
this mode, you can get the body temperature by ear measurement.   

3.2 Surface Mode

After power on, press and hold the “ON/MEM” button, and press the “SCAN” button   

              one time for “Infrared thermometer” mode

 to see

 

  icon on your LCD display (see figure2).   

              In this mode, you can get the target surface temperature. 

 

4. Temperature taking:   

4.1 Ear temperature measuring, points for attention:   

             

Remarks:    a. It is recommended that you measure 3 times with the same ear. If the 3 measurements are different, select the highest temperature.

 

b. Before the measurement, please stay in a stable environment for 5mins and avoid the exercise, bath for 30mins   

     

          c. To avoid the risk of cross contamination, please clean the probe according to “

Care and cleaning

” section after each use.

 

d. The different temperature varies in healthy persons between different parts of the body can be between 0.2~1°C 

e. Clinical repeatability: 0.23°C (<1 year old),0.22°C (1~5 years old),0.21°C (>5 years old) 

** When you get a new temperature reading in ear mode, the last reading will be shown on the screen (at top right corner) with the 

  icon.** 

Specifications 

 

Temperature measurement range:   

Ear mode: 34~42.2°C (93.2~108°F), Surface mode: -22~80°C (-7.6~176°F) 

 

Operating temperature range: 10~40°C (50~104°F) 

 

Storage temperature range: It should be stored at room temperature between -20~+50°C, 
RH

85% 

Transportation temperature shall be less than 70°C, RH

95% 

 

Comply with ASTM E1965-98, EN12470-5:2003 Clinical thermometers-Part   
5:Performance of infra-red ear thermometers(with maximum device), 
IEC/EN60601-1-2(EMC), IEC/EN60601-1(Safety) standards. 

 

Accuracy: 

    For ear mode: +/-0.2°C (0.4°F) within 35.5~42°C (95.9~107.6°F), 

+/-0.3°C (0.5°F) for other range.   

For surface mode: +/-0.3°C (0.5°F) within 22~42.2°C (71.6~108°F),   

    /-4% or +/-2°C (4°F) whichever is greater. 

 

Fever alarm & Temperature records under “Ear mode” 

 

°C / °F Switch mode 

 

Battery: one lithium cell battery (CR2032 *1pcs)

 

 

This thermometer converts the ear temperature to display its “oral equivalent.”

 

 

E:LED Back Light (optional) 

LED Back Light will automatically turned on after measurement, and automatically turned 
off after 5 seconds. 

 

There is no gender and age limitation for using infrared thermometer.   

  This is not an AP

 

or APG product. 

 

 

 

   

   

Figure 1          Figure2               

                                                           

4.1.1 Gently pull the ear back to straighten the ear canal (Fig.4.1.1) and snugly position the probe into the ear canal, aiming towards the 

membrane of the eardrum to obtain an accurate reading (Fig.4.1.2).   

4.1.2  Press  and  hold  the  “SCAN”  button  until  you  hear  a  beep  sound.  Remove  the  probe  from  ear  and  read  the  temperature 

measurement on the LCD.

 

(Rear Cover) 

Figure. 4.1.1 

 

Figure 4.1.2 