DTM 3000 DIGITAL THERMOMETER FOR THERMOCOUPLE SENSORS OPERATING INSTRUCTIONS 

 

The Series 3000 digital thermometer, controlled by a microcontroller, stands out in terms of high accuracy, low power consumption, simplicity 
of operation, and low weight.  
The unit provides rapid high-precision measurements over a wide temperature range.  Many parameters can be individually 
adjusted via the serial interface.  A large choice of sensors allows its application to a very wide variety of measurement tasks. 

 

 

TECHNICAL DATA 

 

 
  

Measurement range:  
    

Type K  -200..1370°C 
Type S        0..1760°C 
Type T  -200..  400°C                                                 set by the manufacturer 
Type J   -200..1200°C 
Type B    300..1820°C 
Type N –260..1300°C    

Resolution: 

0.1°C/1°C                                                                      can be configured via software 

Measurement accuracy: 

0.5%  FS 

Measurement rate: 

maximum 4/s                                                            can be configured via software 

Battery:   

9V-block 6F22 

Battery life: 

>500 hours at 1 per sec 

Sensor: 

 

 

thermocouples type S,K,T,J,B,N    

Functions: 

Min/Max, Hold 

Interface:   

USB                                                                                cable and software must be  obtained separately 

Cold-junction compensation:                     

intern 

Sensor connection: 

miniature thermocouple socket 

 
 
 

DESCRIPTION OF THE UNIT 

 

 
 

Figure 1 
Symbols in the LCD display 

 
 
 
 
   
 

 
 
 
The unit is switched on by actuating the on/off button.  First of all a segment check is carried out for approx. 2 secs. Here the display must look 
like Figure 1.  If segments are missing or the display appears to be very weak the battery should firstly be checked.  If this is OK, the unit should 
be sent to the manufacturer for inspection.  During this period of time a self-test is carried out in the unit.  If a defective function is detected, an 
error number appears in the display.  If the test has been completed successfully, the current measured value is displayed.  The following error 
messages can occur: 
 
 

 

Err1   upper limit of measurement range exceeded 

 

 

Err2   lower limit of measurement range not reached 

 

 

Err4   battery flat 

 
 

 
 
 
 
 