DIGITAL MULTIMETER

OPERATOR’S MANUAL

1. Overview

The multimeter is characterized at slim size, portable, stable performance and anti-dropping
capacity. Using 6000 counts digit LCD monitor with character 25mm high, they offer clear readings.
With overall circuitry design centering on large-scale IC A/D converters in conjunction and
over-load protection circuit, the meters give excellent performance and exquisite making as a
handy utility instrument.
The meters can be used to measure DC & AC voltage, DC & AC current, resistance, capacitor,
frequency, duty cycle, temperature, transistor hFE, Non Contact AC Voltage (NCV) detection,
positive diode voltage fall and audible continuity.

2. Panel Layout

1

NCV detection area: Non Contact AC Voltage (NCV) detection area.

2

CDS sensor: The CDS sensor can reaction to the ambient brightness range, then
automatically control the LCD backlight to lighten or go out.

3

NCV red light: Non Contact AC Voltage (NCV) detection red light.

4

NCV green light: Non Contact AC Voltage (NCV) detection green light.

5

LCD display: 6000 counts digit, full function symbol display.

6

SELECT key: This key work on the "

" range, press the key to choose resistance,

diode or continuity test, on the voltage or current range, change to DC or AC, on the

range, change to

or

test; If press and hold SELECT key to power on, “Auto

Power Off ” function will be disabled.

7

RANGE Key: Press the “RANGE” key, the meter enters manual range mode, press it more
than 2 seconds again, return to auto mode.

8

REL

Key: Press the “REL

” key, the meter enters relative measuring mode, “REL” is

displayed on the LCD and the present reading becomes the reference value and displayed
on the display. Relative measurement REL

=measurement value-Reference value.

9

Hz/Duty Key: On “ACV/ACA” or “Hz” range, press the “Hz/Duty” key, you can choose the
Frequency or Duty Cycle measurement.

10

HOLD key: Press the “HOLD” key to lock display value, and the “DH” sign will appear on
the display, press it again to exit.

11

MAX/MIN key: Press the “MAX/MIN” key to lock MAXMIN or MAX-MIN value, and the
MAX”, “MIN” or “MAX-MIN” sign will appear on the display, press it over 2 seconds to exit.

12

Rotary Switch: Use this switch to select functions and ranges.

13

μAmAT+: μAmAand Temperature “+” Input Jack

14

20A: 20AInput Jack

15

VΩHz

V/Hz/

Input Jack

16

COMT-: COM and Temperature “-” Input Jack

17

Transistor hFE test Input Jack

18

Crust of meter

19

Protective casing

3. Safety Information
3-1 The meter is designed according to IEC-1010 concerning electronic measuring instruments
with an over-voltage category 1000V (CAT

) and pollution 2.

3-2 Follow all safety and operating instructions to ensure that the meter is used safely and is kept
in good operating condition.
3-3 safety symbols:

Important safety information, refer to the operating manual.

Dangerous voltage may be presence.
Double insulation (protection Class II)

4. Special Cautions for Operation

4-1 The meters can be safe only according to standard procedures when used in conjunctions
with the supplied test leads. To replace damaged test leads with only the same model or same
electric specifications.
4-2 To avid risk of electric shock, do not use the meters before the cover is in place.
4-3 The range switch should be right position for the testing.
4-4 To avoid electric shock and damaging the instruments, the input signals are forbidden to
exceed the specified limits.
4-5 When measuring TV set or switched power, attention should be paid to the possible pulses
that may bring destruction to the circuit.
4-6 Range switch position is forbidden to be changed at random during measurement.
4-7 Take caution against shock in the course of measuring voltage higher than DC 60V & AC 30V.
4-8 Protection fuse should be replaced only with same type and same specification.
4-9 After operation is finished, set function switch at OFF to save battery power.
4-10 If the meter is without usage for long time, take out battery to avoid damage by battery
leakage.

5. GENERAL SPECIFICATIONS

5-1 Max Voltage between input terminal and Earth Ground: CAT

1000V

5-2 Over-range Indication: display “OL” for the significant digit.
5-3 Automatic display of negative polarity “-” .
5-4 Low Battery Indication: “

” displayed.

5-5 Max LCD display: 6000 counts digit.

5-6 Auto range & Manual range control
5-7Auto Power Off: When measurement exceeds 15 minutes without switching mode and

pressing key, the meter will switch to standby mode. Press any key to exit standby mode.
When restart the system, press and hold SELECT key to disable auto power off.

5-8 Auto LCD backlight
5-9 Fuse protection: 800mA/500V Fast Fuse, 20A/500V Fast Fuse
5-10 Power supply: 1.5V×2 “AA” R6P battery
5-11 Operating Temp.: 0

to 40

(relative humidity <85%)

5-12 Storage Temp.: -10

to 50

((relative humidity <85%)

5-13 Guaranteed precision Temp.: 23±5

(relative humidity <70%)

5-14 Dimension: 195x88x40mm
5-15 Weight: approx. 350g (including battery)

6. Testing Specifications

Accuracy is specified for a period of year after calibration and at 18

to 28

(64

to 82

) with relative humidity to 70%.

6-1 DC Voltage

Range

Resolution

Accuracy

60mV

0.01mV

±(0.8% of rdg + 2 digits)

600mV

0.1mV

±(0.5% of rdg + 2 digits)

6V

1mV

60V

10mV

600V

100mV

1000V

1V

±(0.8% of rdg + 2 digits)

-- Impedance: 10MΩ, More than 100MΩ on 60mV/600mV range
-- Overload protection: 1000V DC or 750V AC rms

6-2 AC Voltage (True RMS)

Range

Resolution

Accuracy

60mV

0.01mV

±(1.5% of rdg + 3 digits)

600mV

0.1mV

±(1.0% of rdg + 3 digits)

6V

1mV

60V

10mV

600V

100mV

750V

1V

±(1.5% of rdg + 3 digits)

-- Impedance: 10MΩ, More than 100MΩ on 60mV/600mV range
-- Overload protection: 1000V DC or 750V AC rms
-- Frequency Range: 40 to 400Hz

6-3 DC Current

Range

Resolution

Accuracy

600μA

0.1μA

±(1.2% of rdg + 2 digits)

6000μA

1μA

60mA

10μA

600mA

100μA

6A

1mA

±(2.0% of rdg + 3 digits)

20A

10mA

-- Overload protection: 800mA/500V Fast Fuse

20A/500V Fast Fuse, 20A up to 10 seconds

6-4 AC Current (True RMS)

Range

Resolution

Accuracy

600μA

0.1μA

±(1.5% of rdg + 3 digits)

6000μA

1μA

60mA

10μA

600mA

100μA

6A

1mA

±(2.5% of rdg + 5 digits)

20A

10mA

-- Overload protection: 800mA/500V Fast Fuse

20A/500V Fast Fuse, 20A up to 10 seconds

-- Frequency Range: 40 to 400Hz

6-5 Resistance

Range

Resolution

Accuracy

600Ω

0.1Ω

±(1.0% of rdg + 3 digits)

6kΩ

±(1.0% of rdg + 2 digits)

60kΩ

10Ω

600kΩ

100Ω

6MΩ

1kΩ

60MΩ

10kΩ

±(1.5% of rdg + 3 digits)

-- Overload protection: 500V DC or AC rms

6-6 Capacitance

Range

Accuracy

Resolution

9.999nF

±(3.0% of rdg + 10 digits)

1pF

99.99nF

±(2.5% of rdg + 5 digits)

10pF

999.9nF

100pF

9.999µF

1nF

99.99µF

±(5.0% of rdg + 10 digits)

10nF

999.9µF

±(10.0% of rdg + 20 digits)

100nF

9.999mF

1µF

99.99mF

10µF

-- Overload protection: 500V DC or AC rms

6-7 Frequency

Range

Accuracy

Resolution

9.999Hz

± (0.1% of rdg + 5 digits)

0.001Hz

99.99Hz

0.01Hz

999.9Hz

0.1Hz

9.999kHz

1Hz

99.99kHz

10Hz

999.9kHz

100Hz

9.999MHz

1kHz