COMPRESSOR

MODELS:

C3-2, C4-1, C4-2,

C6-2, C7-2 & C11-2

USER MANUAL

Summary of Contents for Aeras Ramvac C11

Page 1: ...COMPRESSOR MODELS C3 2 C4 1 C4 2 C6 2 C7 2 C11 2 USER MANUAL ...

Page 2: ...ents 9 SECTION III INSTALLATION Feet Kit 10 Filter Drain Kit 11 Low voltage Wiring Connections 13 Outlet Plumbing Kit 16 SECTION IV OPERATION How the System Works 18 Getting Started 18 Aeras Control Panel 19 Home Screen 20 Information Screen 22 Technical Screen 24 Settings Screen 44 SECTION V CARE Maintenance 51 SECTION VI USER SERVICE INFO Service Instruction 67 Disposal of Equipment 67 SECTION V...

Page 3: ...ii COMPRESSOR USER MANUAL 866 DTE INFO dentalez com PN 7717 004A NOTES ...

Page 4: ...starting installation procedures review the illustration to become familiar with the components of the compressor FIGURE 1 After the compressor is installed review the features operation procedures and care guidelines with the doctor s staff LEAVE THIS MANUAL IN THE DOCTOR S OFFICE NOTICE Installation by an authorized Ramvac dealer service technician is recommended For any questions about an order...

Page 5: ...nd OFF Dual column desiccant tower provides 100 continuous supply of dry clean air While one column is drying the other is in operation Solenoid valve activates as needed and the chambers switch processes immediately to ensure an uninterrupted supply of air for your practice Design allows for easy desiccant cartridge replacement Stores air at 40 C preventing growth of harmful bacteria Industry lea...

Page 6: ...9 21 29 29 21 29 35 25 30 35 25 30 41 29 32 Dimensions with Sound Cover 33 3 29 34 33 3 29 34 33 3 29 34 35 29 34 35 29 34 41 1 33 36 Peak Airflow Output 80 PSI CFM 6 0 6 6 6 6 9 3 12 0 18 1 Pump Up Cycle Time 85 to 115 PSI Sec 45 45 45 45 40 40 Tank Size Gal 12 12 12 20 20 30 Filtration microns 0 3 Coalescing 0 01 Final Particulate Dryer Duty Cycle 100 Dryer Dew Point PDP 40 F 40 C dBA Levels wit...

Page 7: ...oxide Not intended for use in an oxygen rich environment Mode of Operation Continuous The authorized European representative is Dental Hygienics Decontamination DHD 41 Blackwell Drive Braintree Business Park Braintree Essex CM7 2PU UK Phone 44 01787 877877 ext 200 Caution Warning Biohazard Warning Dangerous Voltage Hot Surface General Mandatory Action Refer to Manual Follow Instructions Refer to A...

Page 8: ...ker Danger of fire or explosion when using flammable substances Never leave children unattended when compressor is in use CAUTION The compressor has several guards designed to protect against accidental contact with dangerous components Do not remove these guards until power has been disconnected turned off from the equipment If service is required with the guard removed and the power on it should...

Page 9: ... immunity of the compressor Portable and mobile radio frequency RF communications equipment can affect medical electrical equipment There may be an instance in low humidity environments that the user of the dental compressed air system may cause the compressor to stop as a result of electrostatic discharge ESD energy To remedy this issue cycle the ON OFF switch on the side of the control An air le...

Page 10: ...s that secure the compressor to the shipping pallet 4 For easier handling four rubber grips have been installed on the frame of the compressor for lifting FIGURE 2 NOTE Lifting the compressor assembly by any other part or location could result in damage to the compressor 5 Verify additional parts shipped with the compressor for use at installation Outlet Plumbing Kit Coalescing Filter Drain Kit Fe...

Page 11: ...ice Install four adjustable rubber feet supplied to ensure unit stands firmly on the floor see SECTION III INSTALLATION FEET KIT Floors should be wood concrete or ceramic tile It is highly recommended that a higher than 30 relative humidity be established and the operator wear a ground strap before physical contact is attempted Compressor is for indoor use only in a dust free climate controlled ro...

Page 12: ... electrical information Any means provided to isolate this device from the supply mains shall isolate all poles simultaneously The disconnection means and over current protection are to be provided by the installer qualified electrician in accordance with applicable local and national electric codes This device should always be attached to a dedicated circuit with appropriate wiring and circuit pr...

Page 13: ...of this model Over exertion could result in serious injuries to the installer This procedure may not comply with all safety requirements of federal state or local agencies or safety requirements of the installer s employer If a conflict does exist follow the requirements set forth by the other organizations ASSEMBLY 1 Remove the four nuts 7 16 wrench from the bolts that secure the Aeras compressor...

Page 14: ...position FILTER DRAIN KIT PARTS INCLUDED Drain Hose Elbow Fitting WARNING Prior to starting this assembly read understand and follow the safety information and the assembly instructions in this user manual The bowl on the coalescing filter assembly is under high pressure while the system is running Allow the compressor to stop running and release the pressure prior to removing the bowl Failure to ...

Page 15: ...mply insert either the hose end or the elbow fitting into the fitting on the bottom of the coalescing filter bowl Apply light force until the item stops Gently pull the opposite direction on the hose or fitting to determine if they have properly seated into the connection If the fitting or hose comes out with light force then the hose was not fully seated into the connection and can be re inserted...

Page 16: ...t does exist follow the requirements set forth by the other organizations PARTS INCLUDED Remote Switch FGH Connector Used to control power state of this compressor from a remote location see LOW VOLTAGE REMOTE SWITCHING Auxiliary Connector Used to control power state of a non Aeras vacuum water control valve or another non Aeras compressor see LOW VOLTAGE AUXILIARY CONNECTION Building Maintenance ...

Page 17: ...tch Option Aeras Control Terminals F H Air Switch 24V DC 24V DC DC Common F G H Air Light Common Ramvac Remote Panel or Equivalent with Illuminated 24V DC Switch Aeras Control Terminals F H G 1 2 3 AIR LOW VOLTAGE REMOTE SWITCHING NOTICE Maximum wire length for 18 gauge wire is 500 feet Low voltage remote switching is used to turn the system on and off from a remote location Your Aeras compressor ...

Page 18: ... control valve or another non Aeras compressor to this Aeras compressor The auxiliary power state will match the power state of this Aeras compressor ON STANDBY Auxiliary Wiring Aeras Control Terminals S2 Control Terminals or C2 Control Terminals or Water Control Valve Terminals F H AUXILIARY CONNECTOR AERAS CONTROL PANEL Switch Common Alternative Switching Continued from the previous page OWL Tou...

Page 19: ...compressor is designed for dental procedures and powering dental equipment and can be safely used in the oral cavity for such purposes Under no circumstances should it be supplied or considered as medical breathing air for a patient Installation of this device should be conducted by licensed electrical and plumbing contractors and should always follow the instructions found in the various manuals ...

Page 20: ...TE The final filter has to be installed in the correct flow orientation 4 Thread the facility supply hose into the final filter and tighten Both ends of the hose have swivel fittings to make installation easier 5 Attach the other end of the hose to the facility piping The kit contains an adapter bushing to adapt the 3 8 pipe thread on the end of the hose to a 1 2 pipe thread on the facility piping...

Page 21: ...sing a pressure transducer sensor on the manifold block which runs the motors on demand Pressure is factory set from 85 PSI 586 kPa to 115 PSI 793 kPa GETTING STARTED WARNING Never direct the air stream toward any portion of the body or others in the vicinity Pressurized air streams can be extremely hazardous to your health and can result in death in certain circumstances CAUTION DO NOT attempt to...

Page 22: ... Point 85 PSI Tank Pressure On Alerts By selecting Accept you consent and agree to our Privacy Policy available at https www dentalez com privacy policy and our Terms of Use available at www dentalez com terms Accept CONTROL PANEL POWER SWITCH PRIVACY POLICY ACCEPTANCE SCREEN SHOWN AT INITIAL SYSTEM START UP CONTROL PANEL LCD DISPLAY HOME Tank Pressure Pressure Dew Point Ambient Temperature Equipm...

Page 23: ...CD display uses simple touch controls and is easy to navigate The Home screen of the control panel displays the following information FIGURE 9 Main navigational menu links HOME INFO TECHNICAL and SETTINGS Wireless signal strength indicator for detailed wireless status go to TECHNICAL MAINTENANCE MODE NETWORK Time of day to set reset time go to SETTINGS DATE TIME Tank pressure reading normal readin...

Page 24: ...om STANDBY mode to ON from the Home screen press the orange STANDBY switch The switch will turn green and display ON Pressing the green ON switch will return the compressor to STANDBY mode FIGURE 10 07 33 14 PM LTE Settings Technical Info Home 20 hrs Equipment Runtime 70 3 F Ambient Temp 40 0 F Pressure Dew Point 85 PSI Tank Pressure On SYSTEM SWITCH ON 07 33 14 PM LTE Settings Technical Info Home...

Page 25: ...mber of the compressor set at factory DentalEZ Support contact phone number set at factory Service Provider Contact phone number set manually by user SERVICE PROVIDER CONTACT phone number screen link FIGURE 11 INFORMATION SCREEN 07 33 14 PM LTE Settings Technical Info Home Model Brand Hardware Firmware Serial Number DentalEZ Support Contact Service Provider Contact v16 v5 Aeras C4 1 ACR0000000 866...

Page 26: ... return to the previous screen FIGURE 12 ON THE INFO SCREEN PRESS SERVICE PROVIDER CONTACT 07 33 14 PM LTE Settings Technical Info Home Model Brand Hardware Firmware Serial Number DentalEZ Support Contact Service Provider Contact v16 v5 Aeras C4 1 ACR0000000 866 383 4636 000 000 0000 SERVICE PROVIDER CONTACT PHONE NUMBER SERVICE PROVIDER CONTACT SCREEN LINK FIGURE 13 ENTER PHONE NUMBER AND PRESS S...

Page 27: ... Time of day HIGH PRESSURE MODE switch MAINTENANCE MODE switch ALERTS screen link MOTORS screen link SENSORS screen link RUNTIMES screen link FIGURE 14 TECHNICAL SCREEN Runtimes Sensors Motors Alerts High Pressure Mode Maintenance Mode 07 33 14 PM LTE Settings Technical Info Home HOME LINK INFO LINK TECHNICAL LINK ACTIVE SETTINGS LINK WIRELESS SIGNAL STRENGTH TIME OF DAY MAINTENANCE MODE SWITCH OF...

Page 28: ...turn green indicating that the compressor s high pressure mode is ON Pressing the green switch will turn the high pressure mode OFF FIGURE 15 FIGURE 15 TECHNICAL SCREEN SHOWING HIGH PRESSURE MODE SWITCH OFF AND ON Runtimes Sensors Motors Alerts High Pressure Mode Maintenance Mode 07 33 14 PM LTE Settings Technical Info Home HIGH PRESSURE MODE SWITCH DISABLED OFF Runtimes Sensors Motors Alerts High...

Page 29: ...rs Motors Alerts High Pressure Mode Maintenance Mode 07 33 14 PM LTE Settings Technical Info Home ALERTS SCREEN LINK To view the Alerts screen on the Technical screen press ALERTS This will open a new screen for viewing system alerts After viewing alert message press message to dismiss it FIGURE 16 Press TECHNICAL to return to the previous screen PRESS TECHNICAL TO RETURN TO PREVIOUS SCREEN PRESS ...

Page 30: ...s Motors Alerts High Pressure Mode Maintenance Mode 07 33 14 PM LTE Settings Technical Info Home MOTORS SCREEN LINK To view the Motors screen on the Technical screen press MOTORS This will open a new screen for viewing the status of each motor FIGURE 17 Press TECHNICAL to return to the previous screen PRESS TECHNICAL TO RETURN TO PREVIOUS SCREEN ...

Page 31: ...dividually enabled or disabled by pressing the motor icon on the screen The GRAY motor icon indicates that the motor is DISABLED and the GREEN motor icon indicates that the motor is ENABLED SECTION IV OPERATION Technical Motor 1 OFF Motor 2 OFF Standby Motors SYSTEM SWITCH STANDBY MOTORS WILL NOT RUN MOTOR ENABLED GREEN MOTOR DISABLED GRAY Technical Motor 1 OFF Motor 2 ON On Motors MOTOR DISABLED ...

Page 32: ... Alerts High Pressure Mode Maintenance Mode 07 33 14 PM LTE Settings Technical Info Home SENSORS SCREEN LINK Technical Motor 1 Temp Motor 2 Temp Motor 1 Power Motor 2 Power Motor 1 Current Motor 2 Current Tank Pressure Tank Temp 86 2 F 88 0 F 0 Watts 0 Watts 0 0 Amps 0 0 Amps 85 PSI 72 8 F Sensors Technical Ambient Temp Ambient Humidity Line Voltage Regen Time 71 9 F 46 6 243 1 V 45 sec Tank Humid...

Page 33: ...vious screen Runtimes Sensors Motors Alerts High Pressure Mode Maintenance Mode 07 33 14 PM LTE Settings Technical Info Home Technical Motor 1 Motor 2 Desiccant Tower Desiccant Cartridge Air Intake Filter Coalescing Filter Particulate Filter Enabled Runtime 20 hrs 20 hrs 20 hrs 20 hrs 20 hrs 20 hrs 20 hrs 30 hrs Runtimes Technical Total In Use Equipment Runtime Pressure Sensor Ambient Sensor Dew P...

Page 34: ... maintenance mode The Service and Network screen links will also become available FIGURE 21 While in maintenance mode the system can be switched from STANDBY mode to ON mode and back again as needed Runtimes Sensors Motors Alerts High Pressure Mode Maintenance Mode 07 33 14 PM LTE Settings Technical Info Home MAINTENANCE MODE SWITCH OFF FIGURE 21 MAINTENANCE MODE ENABLE SCREENS Service Runtimes Se...

Page 35: ...rs Alerts High Pressure Mode Maintenance Mode 07 33 14 PM Maintenance Mode LTE Settings Technical Info Home Service Network Runtimes Sensors Motors Alerts High Pressure Mode Maintenance Mode 07 33 14 PM Maintenance Mode LTE Settings Technical Info Home Are there any service items to record Yes No Technical Select item for service Motor 1 Motor 2 Filter Pack Desiccant Tower Pressure Sensor Ambient ...

Page 36: ...ous screen FIGURE 23 SERVICE OPTIONS SCREEN Service Runtimes Sensors Motors Alerts High Pressure Mode Maintenance Mode 07 33 14 PM Maintenance Mode LTE Settings Technical Info Home Network SECTION IV OPERATION SERVICE SCREEN LINK Technical Select item for service Motor 1 Motor 2 Filter Pack Desiccant Tower Pressure Sensor Ambient Sensor Dew Point Sensor Service Desiccant Cartridge PRESS TECHNICAL ...

Page 37: ...890 Press MOTOR 1 to return to the previous screen FIGURE 24 Service Current Runtime Current Serial Number Replacement Serial Number WARNING Replacing item will reset Current Runtime to 0 and update the serial number if applicable This action CANNOT BE REVERSED Please confirm item name and serial number above before continuing 20 hrs 1010000000 enter serial number On Replace Motor 1 FIGURE 24 MOTOR...

Page 38: ...rial number above before continuing 20 hrs 1010000000 enter serial number On Replace Motor 1 Service Current Runtime Current Serial Number Replacement Serial Number WARNING Replacing item will reset Current Runtime to 0 and update the serial number if applicable This action CANNOT BE REVERSED Please confirm item name and serial number above before continuing 20 hrs 1010000000 enter serial number Re...

Page 39: ...ime to 0 This action CANNOT BE REVERSED Please confirm and replace ALL filters before continuing 20 hrs 2000 hrs 1980 hrs Replace Filter Pack Service Current Runtime Max Runtime Remaining Runtime Replacing filters will reset Current Runtime to 0 This action CANNOT BE REVERSED Please confirm and replace ALL filters before continuing 20 hrs 2000 hrs 1980 hrs Replace Filter Pack Replace Filter Pack OK Can...

Page 40: ... IV OPERATION Continued on the next page FIGURE 27 DESICCANT TOWER REPLACEMENT SCREENS Service Replace Desiccant Tower Current Runtime Max Runtime Remaining Runtime Current Serial Number Replacement Serial Number WARNING Replacing item will reset Current Runtime to 0 and update the serial number if applicable This action CANNOT BE REVERSED Please confirm item name and serial number above before con...

Page 41: ...SED Please confirm item name and serial number above before continuing 20 hrs 10000 hrs 9980 hrs 1010000000 1234567890 PRESS REPLACE CONFIRM ITEM NAME CONFIRM REPLACEMENT SERIAL NUMBER Service Replace Desiccant Tower Current Runtime Max Runtime Remaining Runtime Current Serial Number Replacement Serial Number WARNING Replacing item will reset Current Runtime to 0 and update the serial number if app...

Page 42: ...E REPLACEMENT SCREENS CANCEL UPDATES ACCEPT UPDATES Service Replace Desiccant Cartridge Current Runtime Max Runtime Remaining Runtime WARNING Replacing item will reset Current Runtime to 0 and update the serial number if applicable This action CANNOT BE REVERSED Please confirm item name and serial number above before continuing 20 hrs 2000 hrs 1980 hrs Technical Select item for service Motor 1 Moto...

Page 43: ...tion CANNOT BE REVERSED Please confirm item name and serial number above before continuing 20 hrs Replace Pressure Sensor Technical Select item for service Motor 1 Motor 2 Filter Pack Desiccant Tower Pressure Sensor Ambient Sensor Dew Point Sensor Service Desiccant Cartridge SELECT SERVICE ITEM PRESS REPLACE Service Replace Pressure Sensor Current Runtime WARNING Replacing item will reset Current R...

Page 44: ...Ambient Sensor OK Cancel CANCEL UPDATES ACCEPT UPDATES Service Current Runtime WARNING Replacing item will reset Current Runtime to 0 and update the serial number if applicable This action CANNOT BE REVERSED Please confirm item name and serial number above before continuing 20 hrs Replace Ambient Sensor Technical Select item for service Motor 1 Motor 2 Filter Pack Desiccant Tower Pressure Sensor Am...

Page 45: ...name and serial number above before continuing 20 hrs Replace Dew Point Sensor Technical Select item for service Motor 1 Motor 2 Filter Pack Desiccant Tower Pressure Sensor Ambient Sensor Dew Point Sensor Service Desiccant Cartridge SELECT SERVICE ITEM PRESS REPLACE Service Replace Dew Point Sensor Current Runtime WARNING Replacing item will reset Current Runtime to 0 and update the serial number ...

Page 46: ...Signal Strength Quality Latency Amount of delay or time it takes to send information from one point to the next milliseconds Press TECHNICAL to return to the previous screen Technical Network info Access type Signal strength Quality Latency LTE M 70 70 572 ms Service Runtimes Sensors Motors Alerts High Pressure Mode Maintenance Mode 07 33 14 PM Maintenance Mode LTE Settings Technical Info Home Net...

Page 47: ...witch METRIC switch SCHEDULE screen link DATE TIME screen link RESET link to restore system defaults this cannot be undone Not Set Schedule Smart Schedule Metric 05 31 12 PM LTE Settings Technical Info Home Restore System Defaults Date Time February 03 2020 at 05 31 12 PM Reset WIRELESS SIGNAL STRENGTH TIME OF DAY HOME LINK INFO LINK TECHNICAL LINK SETTINGS LINK ACTIVE FIGURE 34 SETTINGS SCREEN RE...

Page 48: ...EZ Smart Utility website located at aeras dentalez com The Smart Schedule detects usage and will keep the system running beyond the scheduled standby time for up to one hour Not Set Schedule Smart Schedule Metric 05 31 12 PM LTE Settings Technical Info Home Restore System Defaults Date Time February 03 2020 at 05 31 12 PM Reset Not Set Schedule Smart Schedule Metric 05 31 12 PM LTE Settings Techni...

Page 49: ...Thu Fri Sat Sun 09 00 AM 12 00 PM 01 15 PM 06 15 PM 09 00 AM 12 00 PM 01 15 PM 06 15 PM 09 00 AM 12 00 PM 01 15 PM 06 15 PM 09 00 AM 12 00 PM 01 15 PM 06 15 PM 09 00 AM 12 00 PM 01 15 PM 06 15 PM Not Set Not Set Schedule Settings Mon Tue Wed Thu Fri Sat Sun Not Set Not Set Not Set Not Set Not Set Not Set Not Set Schedule Not Set Schedule Smart Schedule Metric 05 31 12 PM LTE Settings Technical Inf...

Page 50: ...one backward or ZONE to move the time zone forward To turn ON automatic daylight savings time press AUTOMATIC DST switch Switch will display green when ON To add 30 minutes to the time zone if necessary press TZ 0 30 Switch will display green when ON Settings Day Month Year 03 02 2020 DST Zone Automatic DST Date Time Hour Min Sec US Canada only TZ 0 30 07 01 12 AM 05 30 30 MINUTE TIME OFFSET ON MO...

Page 51: ...GHT arrow button on either side of the screen will move the menu option arrows to each available option DAY MONTH YEAR HOUR MINUTE SECOND AM PM or ZONE FIGURE 40 DATE TIME MENU OPTIONS SCREEN Settings Day Month Year 03 02 2020 Zone Automatic DST Date Time Hour Min Sec US Canada only TZ 0 30 05 31 12 AM 06 00 RIGHT ARROW BUTTON LEFT ARROW BUTTON MENU OPTION ARROWS MENU OPTIONS Continued on the next...

Page 52: ...TIONS SCREEN Settings Day Month Year 03 02 2020 Zone Automatic DST Date Time Hour Min Sec US Canada only TZ 0 30 05 31 12 AM 06 00 MENU OPTION ARROWS MENU OPTIONS Settings Day Month Year 03 02 2020 DST Zone Automatic DST Date Time Hour Min Sec US Canada only TZ 0 30 07 01 12 AM 05 30 30 MINUTE TIME OFFSET ON MOVE TIME ZONE BACK MOVE TIME ZONE FORWARD AUTOMATIC DST ON FIGURE 42 DATE TIME MENU OPTIO...

Page 53: ...perature pressure measurements in metric units Celsius kPa On the Settings screen press the METRIC switch The switch will turn green indicating that the compressor is now in metric mode FIGURE 43 Pressing the METRIC switch again will turn OFF metric mode returning measurements to imperial units Not Set Schedule Smart Schedule Metric 07 33 14 PM LTE Settings Technical Info Home Restore System Defau...

Page 54: ...rams parts lists and other required information is available to service personnel by contacting Ramvac KEY POINTS FOR TROUBLE FREE OPERATION Follow the recommended PREVENTIVE MAINTENANCE SCHEDULE Check the coalescing and particulate filters indicator every month see CHECK FILTER ELEMENT SERVICE INDICATOR in this section Keep compressor clean and free of dirt Keep area surrounding compressor free o...

Page 55: ...d indicator does not reach the top MAINTENANCE CONTINUED Test for Air Leaks First Week and Annually thereafter Replace Intake Filter Annually Replace Coalescing Filter Annually or when filter element service indicator moves to completely red while running Replace Particulate Filter Annually or when filter element service indicator moves to completely red while running Inspect Dryer Annually Test S...

Page 56: ... tank pressure If the pressure drop is more than 5 PSI 34kPa in a 5 minute time period leaks must be repaired Repair if needed 6 Open shut off valve MAINTENANCE CONTINUED SECTION V CARE FIGURE 45 CLOSE SHUT OFF VALVE SHUT OFF VALVE CLOSED FIGURE 46 RUN COMPRESSOR TO 115 PSI SHUTOFF 07 33 14 PM LTE Settings Technical Info Home 20 hrs Equipment Runtime 71 3 F Ambient Temp 40 0 F Pressure Dew Point 1...

Page 57: ...FIGURE 49 MAINTENANCE CONTINUED SECTION V CARE FIGURE 48 DISABLE MAINTENANCE MODE Service Runtimes Motors Alerts Technical Info Service Runtimes Motors Alerts Technical Info Are there any service items to record Yes No Runtimes Sensors Motors Alerts High Pressure Mode Maintenance Mode 07 33 14 PM LTE Settings Technical Info Home 07 33 14 PM LTE Settings Technical Info Home 20 hrs Equipment Runtime...

Page 58: ...E 51 3 Open the petcock to drain any air in the tank FIGURE 52 DO NOT pull the pressure relief valve to depressurize the air system MAINTENANCE CONTINUED FIGURE 50 TURN SYSTEM POWER OFF CONTROL PANEL POWER SWITCH TANK PETCOCK FIGURE 52 OPEN TANK PETCOCK TO DRAIN AIR FIGURE 51 CLOSE SHUT OFF VALVE SHUT OFF VALVE CLOSED SECTION V CARE ...

Page 59: ...l screen press SERVICE FIGURE 55 6 On the Service screen press FILTER PACK FIGURE 56 SECTION V CARE MAINTENANCE CONTINUED Service Runtimes Sensors Motors Alerts High Pressure Mode Maintenance Mode 07 33 14 PM Maintenance Mode LTE Settings Technical Info Home Network FIGURE 55 PRESS SERVICE Continued on the next page FIGURE 54 REMOVE INTAKE FILTER CAP INTAKE FILTER CAP SCREWS 2 A INTAKE FILTER B FI...

Page 60: ...Runtime Max Runtime Remaining Runtime Replacing filters will reset Current Runtime to 0 This action CANNOT BE REVERSED Please confirm and replace ALL filters before continuing 20 hrs 2000 hrs 1980 hrs Current Runtime Max Runtime Remaining Runtime Replacing filters will reset Current Runtime to 0 This action CANNOT BE REVERSED Please confirm and replace ALL filters before continuing 20 hrs 2000 hrs 1980 ...

Page 61: ...Tab will snap into place 6 On the Technical screen press SERVICE FIGURE 62 7 On the Service screen press FILTER PACK FIGURE 63 MAINTENANCE CONTINUED SECTION V CARE FIGURE 61 REMOVE FILTER BOWL REPLACE FILTER A B TAB FILTER BOWL 3 MICRON FILTER Continued on the next page FIGURE 60 ENABLE MAINTENANCE MODE Service Runtimes Sensors Motors Alerts High Pressure Mode Maintenance Mode 07 33 14 PM Maintena...

Page 62: ...ing Runtime Replacing filters will reset Current Runtime to 0 This action CANNOT BE REVERSED Please confirm and replace ALL filters before continuing 20 hrs 2000 hrs 1980 hrs Current Runtime Max Runtime Remaining Runtime Replacing filters will reset Current Runtime to 0 This action CANNOT BE REVERSED Please confirm and replace ALL filters before continuing 20 hrs 2000 hrs 1980 hrs Replace Filter Pack OK...

Page 63: ...ight into place Tab will snap into place 7 On the Technical screen press SERVICE FIGURE 69 8 On the Service screen press FILTER PACK FIGURE 70 MAINTENANCE CONTINUED SECTION V CARE Continued on the next page FIGURE 67 ENABLE MAINTENANCE MODE Service Runtimes Sensors Motors Alerts High Pressure Mode Maintenance Mode 07 33 14 PM Maintenance Mode LTE Settings Technical Info Home Network FIGURE 69 PRES...

Page 64: ... Runtime Max Runtime Remaining Runtime Replacing filters will reset Current Runtime to 0 This action CANNOT BE REVERSED Please confirm and replace ALL filters before continuing 20 hrs 2000 hrs 1980 hrs Current Runtime Max Runtime Remaining Runtime Replacing filters will reset Current Runtime to 0 This action CANNOT BE REVERSED Please confirm and replace ALL filters before continuing 20 hrs 2000 hrs 1980...

Page 65: ...e tank when the valve is opened the dryer is working proceed to step 6 5 If water drains from the tank when the valve is opened the dryer is not working contact Ramvac at 866 DTE INFO MAINTENANCE CONTINUED SECTION V CARE TANK PETCOCK FIGURE 76 OPEN TANK PETCOCK FIGURE 74 RUN COMPRESSOR TO 45 PSI 310 KPA 07 33 14 PM LTE Settings Technical Info Home 20 hrs Equipment Runtime 70 3 F Ambient Temp 40 0 ...

Page 66: ...NANCE CONTINUED Continued from the previous page FIGURE 77 DISABLE MAINTENANCE MODE Service Runtimes Motors Alerts Technical Info Service Runtimes Motors Alerts Technical Info Are there any service items to record Yes No Runtimes Sensors Motors Alerts High Pressure Mode Maintenance Mode 07 33 14 PM LTE Settings Technical Info Home FIGURE 78 PRESS STANDBY 07 33 14 PM LTE Settings Technical Info Hom...

Page 67: ... ring on safety valve FIGURE 81 NOTE Safety valve is located below the facility shut off valve A loud burst of escaping air will be heard when air is released from the safety valve 4 If no air comes out of safety valve the valve is defective and must be replaced immediately 5 If air comes out of safety valve the valve is working properly If valve is working properly return compressor to service Co...

Page 68: ... press STANDBY to turn the system ON FIGURE 83 07 33 14 PM LTE Settings Technical Info Home 20 hrs Equipment Runtime 70 3 F Ambient Temp 40 0 F Pressure Dew Point 85 PSI Tank Pressure Standby FIGURE 83 PRESS STANDBY FIGURE 82 DISABLE MAINTENANCE MODE Service Runtimes Motors Alerts Technical Info Service Runtimes Motors Alerts Technical Info Are there any service items to record Yes No Runtimes Sen...

Page 69: ...e 20 contact Ramvac at 866 DTE INFO MAINTENANCE CONTINUED Runtimes Sensors Motors Alerts High Pressure Mode Maintenance Mode 07 33 14 PM LTE Settings Technical Info Home FIGURE 84 PRESS SENSORS Technical Motor 1 Temp Motor 2 Temp Motor 1 Power Motor 2 Power Motor 1 Current Motor 2 Current Tank Pressure Tank Temp 86 2 F 88 0 F 0 Watts 0 Watts 0 0 Amps 0 0 Amps 85 PSI 72 8 F Sensors Technical Ambien...

Page 70: ...le for recycling should be disposed of properly For specific questions regarding material type contact DENTALEZ customer service For decommissioning information on associated equipment from other manufacturers refer to the documentation from the manufacturer If the area of concern is not addressed in this manual contact DENTALEZ customer service at 866 DTE INFO See LIMITED WARRANTY Please have the...

Page 71: ...68 COMPRESSOR USER MANUAL 866 DTE INFO dentalez com PN 7717 004A DENTALEZ PARTS ONLINE To view parts online visit dentalezparts com or scan QR code with a smart phone SECTION VII PARTS ...

Page 72: ...c Environment Guidance Electrostatic Discharge ESD IEC 61000 4 2 8 kV Contact 2 kV 4 kV 8 kV 15 kV Air 8 kV Contact 2 kV 4 kV 8 kV 15 kV Air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical Fast Transient Burst IEC 61000 4 4 2 kV 100 kHz repetition frequency 2 kV 100 kHz repetition frequency Mains po...

Page 73: ...m output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitter as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol Test Freq...

Page 74: ...istance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz d 1 2 P d 1 2 P d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 34 0 34 0 74 1 1 7 1 7 2 3 10 3 7 3 7 7 4 100 11 7 11 7 23 3 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the...

Page 75: ...72 COMPRESSOR USER MANUAL 866 DTE INFO dentalez com PN 7717 004A NOTES ...

Page 76: ...or sterilization is not covered under this warranty nor is normal wear and tear Failure to follow or observe our operating use maintenance and or installation instructions voids this warranty as does installation or other work performed by anyone other than DENTALEZ or its authorized dealers the use of non authorized parts and repairs made by an unauthorized repair facility Liability is limited to...

Page 77: ... DentalEZ Alabama Inc Printed in USA PN 7717 004A March 2020 2500 Highway 31 South Bay Minette Alabama 36507 866 DTE INFO dentalez com ...