background image

 

 

M601020-210322-Rev009-UM-e

 

Operating Manual

 

Class III Approved Scales 

 
Electronic Bed Scale

 

M601020 

 

Summary of Contents for M601020

Page 1: ...M601020 210322 Rev009 UM e Operating Manual Class III Approved Scales Electronic Bed Scale M601020 ...

Page 2: ...vices 5 Housing and connections 6 Safety 6 Settings 6 Setting date and time 7 Operation 7 Modes of operation 7 Before weighing 7 Weighing 8 Tare function 8 Hold function 8 Determining the Body Mass Index BMI 9 Evaluating Body Mass Index 9 Print out a record 9 Care servicing and recalibration 9 Cleaning 9 Error messages 9 Waste disposal 10 Storage and transport conditions 10 Warranty 10 Recalibrati...

Page 3: ...in dry and level surface before the weighing procedure Scope of delivery 2 Weighing platforms Connection cable between weighing platforms Indicator incl connection cable Mounting Device for the indicator Mains adapter Operating Manual including Certificate of Conformity Assembly Dismantle packaging and foil carefully Weighing platforms are equipped with castors and a handle for easy transport 1 Mo...

Page 4: ...xpose the batteries to temperatures below 10 C 14 F or above 45 C 113 F Rechargeable batteries lose their charging capacity over the time and require a longer charging period the older they are If the battery performance drops considerably then it is time to replace the battery New batteries are supplied in partial charged condition Batteries achieve their full capacity after several complete char...

Page 5: ... 0 9 Clear Enter and Navigation functions Key 6 menu direction to the right Key 4 menu direction to the left Key 8 menu direction down Key 2 menu direction up For better understanding see Pic 1 1 the blue arrows Used to navigate through menus to select numerical values and to increase or lower the value Special function 0 key In weight mode you can change between gross and net weight using figure ...

Page 6: ...at the mains cable does not pose any potential trip hazard Do not drop the weighing bridges shoes or subject them to violent shocks Do not operate the unit in hazardous or unstable conditions Do not dip or submerge the unit into water or other liquids Have all repairs and calibration carried out by professional personnel authorised by ADE Set up the unit as far away as possible from equipment that...

Page 7: ...ime and consecutive numbers for print commands are being set in this mode To start setting mode hold the GROSS NET key pressed until TIME appears on the display Configuration mode set up mode The values for different parameters including the calibration could be set in the configuration mode This func tion is for authorized service personnel only Test mode There are different diagnostic functions ...

Page 8: ...alues The user is able to insert manually the gross weight which is for example mentioned at the identification plate of the bed Enter the appropriate tare value just by using the key pad under normal weighing mode The value will be stored after use of the ENTER key or TARE The ADE display is now ready for weighing and changes to net mode NET status indicator 5 is on Enter the tare button to leave...

Page 9: ...e Printing is not possible if the net weight is 0 The print function only produces one print out A reset is required to print out another copy Care servicing and recalibration Cleaning Remove the mains plug before cleaning the unit Do not use any aggressive cleaning materials for cleaning only use a damp cloth or conventional disinfectant Ensure that no liquids or any fluid gets into the scale Err...

Page 10: ...aterial or workmanship presentation of invoice is necessary All moveable parts bat teries cables mains units rechargeable batteries etc are excluded Warranty does not include natural wear damages caused by incorrect or negligent handling overuse or unsuitable or improper storage The cost of transport will be borne by the customer should the equipment be located anywhere other then the custom er s ...

Page 11: ...rt Full duplex RS 232 38400 19200 9600 4800 2400 1200 bps Printer Port Output only RS 232 or active 20 mA current loop 9600 4800 2400 1200 bps Both Ports 7 or 8 data bits even odd or no parity Operator Interface Display 6 digit LCD display 7 segment 0 815 in 21 2 mm digits LCD annunciator net center of zero standstill kg lb g oz units Battery Set points Hold HI OK LO Keyboard 24 key flat membrane ...

Page 12: ...facturer This product has been manufactured in accordance with the harmonised European standards They con form to the regulations of the EC directives listed below NAWI Directive 2014 31 EC Medical Devices Directive 93 42 EEC EMC Directive 2014 30 EC regarding electromagnetic compatibility Low Voltage Directive 2014 35 EC RoHS Directive 2011 65 EC in the currently valid versions This declaration l...

Page 13: ...M601020 210322 Rev009 UM e 13 ...

Page 14: ...14 M601020 210322 Rev009 UM e ...

Page 15: ...M601020 210322 Rev009 UM e 15 ...

Page 16: ...16 M601020 210322 Rev009 UM e Dealer ...

Reviews: