34
NexWave
Warnings
Interferential Current (IFC)
Warnings (Continued)
·
Stimulation shall not be applied transcerebrally.
·
Stimulation shall not be applied over swollen, infected, or inflamed areas of skin,
eruptions, e.g. phlebitis, thrombophlebitis, varicose veins, etc.
·
Stimulation shall not be applied over, or in proximity to, cancerous lesions.
·
This device is capable of generating currents and voltages in excess of 10 mA RMS
and 10 V RMS, respectively. A yellow LED indicator is provided next to each lead
wire connector to show that the unit is delivering energy for any non
-
zero value of
stimulation output (1
-
50 mA).
·
Simultaneous connection of a patient to high frequency surgical equipment may
result in burns at the site of the stimulator electrodes and possible damage to the
stimulator.
·
Operations within close proximity (within one meter) of shortwave or microwave
therapy equipment may produce instability in the stimulation output.
Safety References
Zynex Medical, Inc. is only responsible for the safety, reliability and function of this
device when repairs and adjustments have been made by persons authorized by Zynex
Medical, Inc., and this device is used in accordance with the user’s manual. Repairs
and technical safety tests shall only be performed by authorized personnel.
15
Uses electricity from wall socket
instead of 9 V battery
A/C Adapter Input
Channel 2 Intensity Control
Up and Down Button
Channel 1 Intensity Control
Up and Down Button
Lead Wire Connection
Channel 1
Lead Wire Connection
Channel 2
Turns Unit On/Off
On/Off Button
IFC, TENS, NMES
Modality Buttons
Unlock Button Controls
Unlock Button
Treatment Timer
Treatment Timer Button
Display
NexWave
Device Controls
Summary of Contents for NexWave
Page 2: ...2 47...